Endo Stitch 173016Recalls & reported problems
Covidien LP · Suturing Device · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien Endo Stitch suturing device is an endoscopic tool used to close tissue. Many of the 1,331 reports describe parts breaking or detaching, such as needles breaking during surgery, and some describe needle fragments remaining in patients. The reports include 1 death, 300 injuries and 1,030 malfunctions. Reports fell from 72 in the prior 12 months to 37 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,332
- Deaths reported
- 1
- Injuries reported
- 297
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,332 adverse event reports that list this exact device (DI 10884521100008), received between Jun 27, 2016 and Sep 24, 2026 — classified as 1 death, 297 injuries, 1,034 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (484 reports).
2 other recalls cover devices cleared under the same FDA submission (K934738); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component484
- Human-Device Interface Problem462
- Break432
- Detachment Of Device Component343
- Component or Accessory Incompatibility142
- Difficult to Open or Close70
- Adverse Event Without Identified Device or Use Problem52
- Mechanical Jam27
Patient problems
- No Known Impact Or Consequence To Patient485
- No Clinical Signs, Symptoms or Conditions315
- No Consequences Or Impact To Patient187
- Foreign Body In Patient142
- Unintended Radiation Exposure122
- Insufficient Information92
- No Code Available67
- Radiation Exposure, Unintended59
Recent adverse event reports
The 25 most recent of 1,332 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Break
It was reported that during a laparoscopic inguinal hernia repair procedure, the needle broke. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50877113 ↗MalfunctionSep 23, 2026Component or Accessory Incompatibility; Detachment of Device or Device Component
It was reported that during a laparoscopic hysterectomy, the device was used to close the vaginal cuff. One reload was used successfully; however, a second reload could not be loaded properly. A subsequent reload was loaded, and during use, a portion of the needle broke off and fell into the patient¿s abdomen. X-ray imaging performed on (b)(6) 2026, visualized the needle fragment in the abdomen. The physician re-entered the abdomen and attempted to locate the fragment but was unable to find it. The fragment was intentionally left in the patient because the physician determined that conversion to an open procedure to attempt retrieval was not preferable. Surgical time was extended by 30 minutes or more.
Patient problems: Foreign Body In Patient; Unintended Radiation Exposure
MAUDE report 50863576 ↗MalfunctionSep 22, 2026Detachment of Device or Device Component; Human-Device Interface Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862819 ↗MalfunctionSep 15, 2026Detachment of Device or Device Component
It was reported that during laparoscopic hysterectomy for vaginal cuff closure, when the physician opened the device after taking the first bite of tissue, the suture and part of the needle disengaged. The needle and suture fell into the patient cavity and were retrieved by the physician. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50766479 ↗MalfunctionSep 4, 2026Detachment of Device or Device Component
It was reported that during a laparoscopic hysterectomy, the needle disengaged from the device. A second handle was used to resolve the issue. There was no patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50600962 ↗MalfunctionAug 12, 2026Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50226851 ↗MalfunctionAug 4, 2026Component or Accessory Incompatibility; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50129974 ↗InjuryJul 28, 2026Detachment of Device or Device Component
According to the reporter, during a laparoscopically assisted vaginal hysterectomy, when the endo stitch was used to close the vaginal cuff, the needle broke off. An x-ray was done and the needle was retrieved. The surgeon used a traditional needle driver to complete the case.
Patient problems: Unintended Radiation Exposure
MAUDE report 50061146 ↗MalfunctionJul 1, 2026Break
IT WAS REPORTED THAT DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY WITH VAGINAL CUFF CLOSURE, THE THREE NEEDLES WAS REPORTED TO HAVE BROKEN. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25682373 ↗InjuryJun 24, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC NISSEN FUNDOPLICATION, ESOPHAGOGASTRODUODENOSCOPY, THE DEVICE WAS DROPPED FROM SUTURING DEVICE DUE TO MALFUNCTION WHILE SUTURING TISSUE. THERE WAS MORE PATIENT TRAUMA TRYING TO REMOVE THE NEEDLE THAN TO LEAVE IT BEHIND. THE NEEDLE WAS SEEN AND WAS NO RETRIEVED ON POST-OP X-RAY.
Patient problems: Unintended Radiation Exposure
MAUDE report 25610879 ↗InjuryJun 19, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC DUODENAL SWITCH, WHILE APPROXIMATING THE SMALL BOWEL, THE SURGEON HAD COMPLETED THE USE OF THE SUTURE, THE DEVICE WORKED COMPLETELY FINE. ONCE THE SURGEON HAD CUT THE SUTURE TO TAKE THE ENDOSTITCH DEVICE OUT OF THE ABDOMEN, THE NEEDLE HAD FALLEN LOSE AND IT WAS LOST IN THE OPERATING ROOM. THE C-ARM WAS BROUGHT INTO THE ROOM TO SEE IF THE NEEDLE COULD BE FOUND INSIDE THE PATIENT¿S ABDOMEN, BUT THE RESULT CAME BACK NEGATIVE. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES. THE ENDOSTITCH DEVICE WAS USED AGAIN WITH NO ISSUE, BUT THE NEEDLE WAS UNABLE TO BE FOUND.
Patient problems: Unintended Radiation Exposure
MAUDE report 25559904 ↗MalfunctionJun 5, 2026
"MALFUNCTION OF ENDO STITCH SUTURING DEVICE, REFERENCE NUMBER 173016. LOT J5E3790EY, UDI: (B)(4). DURING PROCEDURE, 0 SILK ES-9 ENDO-NEEDLE WAS DROPPED FROM SUTURING DEVICE DUE TO MALFUNCTION WHILE SUTURING TISSUE. PER DR. [REDACTED], IT WOULD CAUSE MORE PATIENT TRAUMA TRYING TO REMOVE THE NEEDLE THAN TO LEAVE IT BEHIND. EVENT WAS DOCUMENTED IN SURGICAL COUNTS AND NEEDLE WAS DOCUMENTED AS A RETAINED PACKED ITEM IN FLOWSHEETS/AVATAR. NEEDLE WAS SEEN AND VERIFIED BY SURGEON ON POST-OP XRAY. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25408140 ↗MalfunctionMay 3, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF TOTAL LAPAROSCOPIC HYSTERECTOMY, WHILE THE SCRUB TECHNICIAN WAS LOADING A NEW RELOAD DURING VAGINAL CUFF CLOSURE, THE BLACK RELEASE BUTTON ON THE HANDLE BROKE OFF. THE DEVICE WAS NOT USED IN THE PATIENT. ANOTHER HANDLE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25064298 ↗InjuryApr 30, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE NEEDLE WAS USED AND THE NEEDLE BROKE OFF. IT WAS REPORTED THAT THE NEEDLE FRAGMENT REMAINED IN THE PATIENT¿S TISSUE AND WAS UNRETRIEVABLE.
Patient problems: Foreign Body In Patient
MAUDE report 25050287 ↗InjuryApr 14, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A GYNECOLOGICAL PROCEDURE, THE NEEDLE CAME OFF FROM THE DEVICE AND COULD NOT BE SAFELY RETRIEVED FROM THE PATIENT WITHOUT RISKING FURTHER HARM. THE PATIENT RECEIVED POSTOPERATIVE 40 MG SUBCUTANEOUS INJECTION AS PART OF STANDARD CARE.
Patient problems: Foreign Body In Patient
MAUDE report 24877316 ↗MalfunctionApr 9, 2026Break
IT WAS REPORTED THAT DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, WHILE BEING PASSED THROUGH THE VAGINAL CUFF FOR VAGINAL CUFF CLOSURE, THE NEEDLE BROKE. UPON EXITING THE CUFF, THE SURGEON NOTICED ABOUT 4MM OF THE NEEDLE WAS MISSING. A NEW PRODUCT WAS OPENED TO COMPLETE THE VAGINAL CUFF CLOSURE. THE MISSING PORTION COULD NOT BE LOCATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829718 ↗MalfunctionMar 25, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE TOTAL LAPAROSCOPIC HYSTERECTOMY WHILE CLOSING THE VAGINAL CUFF, THERE WERE NO ISSUES WITH LOADING OR TOGGLING OF THE DEVICE, HOWEVER, WHEN THE SURGEON ATTEMPTED TO TAKE A BITE OF TISSUE, THE NEEDLE FELL OUT OF THE DEVICE. THE ISSUE OCCURRED ON THREE SUTURES. THE NEEDLE COMPONENTS FELL INTO THE CAVITY OF THE PATIENT AND WAS RETRIEVED WITH A GRASPER. A NEW INSTRUMENT AND SUTURE WAS OPENED TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24693046 ↗InjuryMar 19, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, A LAPAROSCOPIC HYSTERECTOMY WAS UNCOMPLICATED UNTIL CLOSURE. THE VAGINAL CUFF WAS CLOSED USING THE DEVICE. WHEN THE BACK SUTURE WAS PERFORMED TO LOCK THE SUTURE IN PLACE, THE NEEDLE WAS TOGGLED, BECAME EMBEDDED IN TISSUE, AND DISLODGED FROM THE DEVICE. UPON REMOVAL, THE SUTURE BROKE OFF FROM THE NEEDLE, WHICH REMAINED RETAINED IN THE VAGINAL CUFF. MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE NEEDLE VAGINALLY AND LAPAROSCOPICALLY. C-ARM FLUOROSCOPY DEMONSTRATED THE NEEDLE IN SITU NEAR THE LEFT PUBIC RAMUS, RAISING CONCERN THAT THE NEEDLE WAS LOCATED IN THE VAGINAL CUFF. THE NEEDLE WAS UNABLE TO BE PALPATED OR LOCATED IN THE RESECTED TISSUE. A MAGNET WAS USED IN ATTEMPTS TO LOCATE THE NEEDLE LAPAROSCOPICALLY AND VAGINALLY WITHOUT SUCCESS. TISSUE WAS RESECTED; HOWEVER, DUE TO PROXIMITY TO THE BLADDER EDGE AND NEAR A VASCULAR PEDICLE, THE DECISION WAS MADE TO ABANDON FURTHER ATTEMPTS AT REMOVAL AND TO CLOSE THE CUFF.
Patient problems: Foreign Body In Patient; Unintended Radiation Exposure
MAUDE report 24638102 ↗InjuryMar 13, 2026Break; Human-Device Interface Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY BILATERAL SALPINGO-OOPHORECTOMY, WHILE USING THE SUTURING INSTRUMENT TO CLOSE THE VAGINAL CUFF AFTER REMOVAL OF THE CERVIX AND UTERUS, THE DEVICE WAS DIFFICULT TO UNLOAD AND THE STRING FELL OUT AFTER THE FIRST BITE, AND UPON REMOVAL OF THE SUTURING DEVICE, OPERATING ROOM STAFF NOTICED A PIECE OF THE NEEDLE WAS MISSING AND IMMEDIATELY INFORMED THE PHYSICIAN. AN INTRAOPERATIVE X-RAY WAS PERFORMED TO LOCATE THE RETAINED NEEDLE FRAGMENT. ATTEMPTS TO LOCATE THE DEVICE LED TO INCREASED BLEEDING DURING SURGERY, PROLONGED ANESTHESIA TIME DUE TO EXTENDED SURGICAL TIME OF MORE THAN 30 MINUTES, AND THE INSERTION OF AN ADDITIONAL TROCAR, RESULTING IN ANOTHER SCAR. THE CUFF WAS CLOSED VAGINALLY, HEMOSTASIS WAS ACHIEVED, AND AFTER LOCATING THE NEEDLE FRAGMENT ON X-RAY BUT BEING UNABLE TO RETRIEVE IT DUE TO PATIENT BLEEDING, THE SURGICAL TEAM DECIDED NOT TO INTERVENE FURTHER.
Patient problems: Foreign Body In Patient; Hemorrhage/Blood Loss/Bleeding; Laceration(s); Unintended Radiation Exposure
MAUDE report 24599470 ↗InjuryMar 10, 2026Break
ACCORDING TO THE REPORTER, DURING SUTURING OF A VAGINAL CUFF, THE NEEDLE BROKE IN THE TISSUE DURING THE PROCEDURE. IMAGING CONFIRMED THAT A NEEDLE REMNANT WAS RETAINED IN THE TISSUE, AND ATTEMPTS WERE MADE TO RETRIEVE THE RETAINED FRAGMENT BOTH LAPAROSCOPICALLY AND VAGINALLY HOWEVER, THE FRAGMENT COULD NOT BE RECOVERED, AND THE DECISION WAS MADE TO CLOSE THE PROCEDURE.
Patient problems: Foreign Body In Patient; Unintended Radiation Exposure
MAUDE report 24554541 ↗MalfunctionMar 4, 2026
LAPAROSCOPIC HYSTERECTOMY UNCOMPLICATED UNTIL CLOSURE- THE VAGINAL CUFF WAS CLOSED WITH ENDOSTITCH DEVICE AND WHEN BACK SUTURE TO LOCK THE SUTURE IN PLACE OCCURRED THE NEEDLE WHEN TOGGLED WAS EMBEDDED IN TISSUE AND DISLODGED FROM THE ENDOSTITCH DEVICE- UPON REMOVAL SUTURE BROKE OFF OF THE NEEDLE WHICH REMAINED RETAINED IN CUFF- MULTIPLE ATTEMPTS COMPLETED TO REMOVE NEEDLE VAGINALLY AND LAPAROSCOPICALLY- C- ARM FLUOROSCOPY PERFORMED SHOWING THE NEEDLE IN SITU LEFT BY PUBIC RAMI WHERE CONCERN NEEDLE IN VAGINAL CUFF NOTED. UNABLE TO BE PALPATED OR LOCATED IN RESECTED TISSUE. MAGNET USED TO TRY AND LOCATE IN TISSUE LAPAROSCOPICALLY AND VAGINALLY AND UNABLE TO BE FOUND. TISSUE RESECTED BUT GIVEN CLOSE TO BLADDER EDGE AND NEARING VASCULAR PEDICLE DECISION WAS MADE TO ABANDON REMOVAL AND CUFF WAS CLOSED VAGINALLY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 24505589 ↗MalfunctionMar 3, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BYPASS, DURING THE CLOSING OF THE MESENTERY, THE NEEDLE DISENGAGED AND FELL INTO THE CAVITY. THE RELOAD WAS REPLACED, BUT THE SAME ISSUE OCCURRED. THE NEEDLE WAS RETRIEVED BY USING GRASPERS. A NEW ENDOSTITCH INSTRUMENT AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24499805 ↗MalfunctionFeb 25, 2026Difficult to Open or Close; Human-Device Interface Problem
IT WAS REPORTED THAT DURING A THORACIC LAPAROSCOPIC PROCEDURE, THE JAWS OF THE DEVICE WOULD NOT OPEN AND WERE STUCK TOGETHER, AND THE NEEDLE REMAINED IN THE JAWS OF THE DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24449196 ↗InjuryFeb 17, 2026Detachment of Device or Device Component; Human-Device Interface Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC INSERTION OF A MEDIPORT, AFTER TWO SUCCESSFUL DEVICE USES WHEN SUTURING THE SURROUNDING TISSUE OF THE CATHETER TO THE ABDOMINAL WALL, THE DEVICE WAS RELOADED BY THE SCRUB NURSE AND HANDED OFF TO THE SURGEON. THE DEVICE WAS CYCLED, THEN THE SURGEON INSERTED IT IN THE CAVITY OF THE PATIENT, PLACED THE TISSUE BETWEEN THE JAWS OF THE INSTRUMENT, CLOSED THE JAWS AND CYCLED THE INSTRUMENT, HOWEVER THE NEEDLE DID NOT GO THROUGH THE TISSUE AND WAS LODGED IN THE ABDOMEN. ATTEMPTS TO PULL THE SUTURE OUT WITH THE NEEDLE IN THE PATIENT'S ABDOMEN WERE UNSUCCESSFUL AND THEN THE NEEDLE BROKE OFF THE SUTURE. THE NEEDLE WAS LEFT IN THE ABDOMEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24361633 ↗MalfunctionFeb 16, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, A NEEDLE FELL OFF.
Patient problems: Insufficient Information
MAUDE report 24359248 ↗
FDA clearance and classification
Cleared through 510(k) K934738 on Mar 16, 1994 (decision: Substantially Equivalent). Submitted by United States Surgical, A Division of Tyco Healthc as “Autosuture Endoscopic Suturing Device”. The FDA review took 166 days. FDA record ↗
The FDA regulates product code OCW (endoscopic tissue approximation device) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K934738. One submission often covers many models, so check the product and lot details in each record.
- Z-2281-2026 · Apr 15, 2026 · Open, ClassifiedCovidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;
- Z-2543-2017 · May 3, 2017 · TerminatedSutures. ENDO STITCH: ENDO STITCH POLYSORB 0 48" U/D DLU SU, Item Number: 170056; ENDO STITCH POLYSORB 0 48" VIO DLU SU, Item Number: 170052; ENDO STITCH POLYSORB 2/0 48" U/D DLU
Common questions about the Endo Stitch 173016
Has the Endo Stitch 173016 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521100008), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Endo Stitch 173016?
The FDA's MAUDE database lists 1,332 adverse event reports for this device, received between Jun 27, 2016 and Sep 24, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (484), “Human-Device Interface Problem” (462) and “Break” (432); the most common patient problems are “Foreign Body In Patient” (142), “Unintended Radiation Exposure” (122) and “Radiation Exposure, Unintended” (59). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Endo Stitch 173016?
1 report is classified as a death, 297 as injuries and 1,034 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Endo Stitch 173016 FDA approved?
It was cleared through 510(k) K934738 on Mar 16, 1994. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Endo Stitch 173016, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Endo Stitch, model 173016). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic tissue approximation device devices
- ProTack 174006Covidien LP · Endoscopic tissue approximation device705 reports
- N/A ECS29AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device1 death517 reports
- N/A ECS25AEthicon Endo-Surgery, LLC · Endoscopic tissue approximation device233 reports
- OverStitch™ M00505530Boston Scientific Corporation · Endoscopic tissue approximation device1 recall75 reports
- Polysorb 170053Covidien LP · Endoscopic tissue approximation device1 recall63 reports
- Surgidac 173024Covidien LP · Endoscopic tissue approximation device70 reports
- NA M00505550Boston Scientific Corporation · Endoscopic tissue approximation device1 recall12 reports
- Polysorb 170052Covidien LP · Endoscopic tissue approximation device57 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page