N/A ECS25ARecalls & reported problems

Ethicon Endo-Surgery, LLC · Endoscopic Curved Intraluminal Stapler (ILS) 20 Titanium Adjustable Height Staples · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon curved intraluminal stapler is an endoscopic stapler used to join tissue. The 235 reports mostly describe staples that did not form or fire properly, and some describe staples formed incorrectly, with surgeons finishing with a new stapler or sutures. The reports include 32 injuries and no deaths, and reports rose to 36 in the last 12 months from 17 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
233
Deaths reported
0
Injuries reported
32
Recalls naming it
0

Summary

The FDA's MAUDE database contains 233 adverse event reports that list this exact device (DI 10705036003496), received between Dec 1, 2017 and Sep 25, 2026 — classified as 32 injuries, 201 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (82 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 10 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 6 reportsMay 2024: 3 reportsJun 2024: 6 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 4 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 7 reportsFeb 2026: 6 reportsMar 2026: 4 reportsApr 2026: 4 reportsMay 2026: 3 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple82
  • Failure to Fire72
  • Ejection Problem23
  • Adverse Event Without Identified Device or Use Problem15
  • No Apparent Adverse Event13
  • Failure to Cut11
  • Appropriate Device Problem Term/Code Not Available8
  • Firing Problem7

Patient problems

  • No Clinical Signs, Symptoms or Conditions157
  • No Consequences Or Impact To Patient45
  • No Code Available20
  • Failure to Anastomose16
  • Hemorrhage/Blood Loss/Bleeding9
  • Blood Loss3
  • Post Operative Wound Infection2
  • Tissue Damage2

Recent adverse event reports

The 25 most recent of 233 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896462 ↗
  • MalfunctionSep 25, 2026Failure to Fire

    It was reported that during an Unknown surgery, could not fire. Changed the new one to complete surgery. There was no patient consequence reported. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50896249 ↗
  • MalfunctionSep 16, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50788054 ↗
  • MalfunctionAug 13, 2026Failure to Form Staple

    It was reported that during an Unknown surgery, after fired, noted the staples malformed. Changed the new one to complete surgery. No additional information can be provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50235996 ↗
  • MalfunctionJun 29, 2026Failure to Form Staple; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING THE PROCTRECTOMY SURGERY, AFTER FIRED, NOTED SOME STAPLES MALFORMED AND OOZING OCCURRED. SUTURE WAS USED TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25652969 ↗
  • MalfunctionJun 18, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE RECTUM PROCEDURE SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25553372 ↗
  • MalfunctionMay 26, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25281089 ↗
  • MalfunctionMay 23, 2026Ejection Problem

    IT WAS REPORTED THAT BEFORE THE PROCTOCOLECTOMY SURGERY, NOTE THE SPRING PLATE OF THE ANVIL WAS LOOSE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25264698 ↗
  • MalfunctionMay 7, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25113519 ↗
  • MalfunctionApr 22, 2026Failure to Form Staple; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING THE UNK SURGERY. AFTER FIRED, NOTED THAT A FEW STAPLES FORMED WITH IRREGULAR SHAPES AND ANASTOMOTIC SITE WAS OOZING. TO STOP OOZING WITH COMPRESSION HEMOSTASIS. THERE WAS NO PATIENT CONSEQUENCE REPORTED, NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956678 ↗
  • MalfunctionApr 22, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24956591 ↗
  • MalfunctionApr 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE ESOPHAGECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831343 ↗
  • MalfunctionApr 2, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769862 ↗
  • MalfunctionMar 31, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24747829 ↗
  • MalfunctionMar 31, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24747793 ↗
  • MalfunctionMar 30, 2026Failure to Fire

    IT WAS REPORTED THAT DURING LAP NEPHRECTOMY PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24731419 ↗
  • MalfunctionMar 30, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24718605 ↗
  • MalfunctionFeb 24, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24433701 ↗
  • MalfunctionFeb 20, 2026Ejection Problem

    IT WAS REPORTED THAT BEFORE THE UNK SURGERY, OPEN THE PACKING AND NOTED COULD NOT DETACH THE ANVIL. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405540 ↗
  • MalfunctionFeb 20, 2026Failure to Form Staple; No Apparent Adverse Event

    IT WAS REPORTED THAT DURING THE LAP GASTRECTOMY SURGERY, AFTER FIRED, NOTED SOME STAPLES MALFORMED AND OOZING OCCURRED. SUTURE WAS USED TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401700 ↗
  • MalfunctionFeb 18, 2026Failure to Cut

    IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THAT THE TISSUE WAS NOT CUT COMPLETELY. THEN CHANGED THE NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24381921 ↗
  • MalfunctionFeb 18, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE LAP GASTRECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24381647 ↗
  • MalfunctionFeb 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24294660 ↗
  • MalfunctionJan 8, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE RIGHT HEMICOLECTOMY SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005720 ↗
  • MalfunctionJan 5, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23966974 ↗

Other Endoscopic tissue approximation device devices

All Endoscopic tissue approximation device devices →