COOLIEF* Cooled RF Advanced Generator CRG-ADVANCEDRecalls & reported problems
Avanos Medical, Inc. · COOLIEF* Cooled RF Advanced Generator · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00193493473014), received between Feb 13, 2025 and Sep 23, 2026 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Impedance Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K192491); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Impedance Problem3
- Break1
- Difficult to Remove1
- Patient-Device Incompatibility1
- Sparking1
Patient problems
- Arrhythmia3
- Abdominal Pain1
- Asystole1
- Inflammation1
- No Clinical Signs, Symptoms or Conditions1
- Peritonitis1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 23, 2026Patient-Device Incompatibility
It was reported that during a cervical radiofrequency ablation (RFA) procedure, a patient with a pacemaker/implantable cardioverter-defibrillator experienced asystole approximately 17 seconds after radiofrequency treatment was initiated. Cardiopulmonary resuscitation (CPR) was performed and the patient was successfully resuscitated. The procedure was stopped.Information notes:Patient with Pacemaker implantable cardioverter-defibrillator (ICD). Boston Scientific Pacemaker ICD Representative was contacted prior to procedure. Magnet was given to staff from BostonScientific representative the day before the procedure and was placed on patient pacemaker ICD prior to starting procedure. Grounding Pad placed on Right Flank/Lower Back. The Disposable (or Diros) Trident Radiofrequency Hybrid insulated cannula DTRH length unknown. No issue during Sensory Testing. No issue during Motor Testing.Additional information received 04-Sep-2026 the generator that was used at another location. Magnet placement was verified prior to initiating treatments. At the time of asystole the RFA temperature setting was 80 Celsius.
Patient problems: Asystole
MAUDE report 50872969 ↗InjuryAug 6, 2026Impedance Problem
It was reported during a radio frequency ablation (RFA) case "the patient experienced what appeared to be an episode of near asystole. This event occurred during motor testing, prior to treatment delivery, of the genicular nerves; the patient almost went "Asystolic/Flatlined. Anesthetist intervened. Treatment continued, no issues afterwards. Patient was healthy, had no pacemaker."
Patient problems: Arrhythmia
MAUDE report 50158914 ↗InjuryAug 6, 2026Impedance Problem
It was reported during a radio frequency ablation (RFA) case "the patient experienced what appeared to be an episode of near asystole." This occurred mid-treatment. In this case, treatment waspaused until the patient recovered, after which the procedure was completed without further issue.
Patient problems: Arrhythmia
MAUDE report 50158906 ↗MalfunctionAug 5, 2026Impedance Problem
Patient problems: Arrhythmia
MAUDE report 50141188 ↗MalfunctionMay 28, 2025Difficult to Remove; Sparking; Difficult to Remove; Sparking
IT WAS REPORTED, DIFFICULTY REMOVING THERAPY CABLE FROM GENERATOR THAT RESULTED IN "SPARK" FROM PORT WHERE THERAPY CABLE CONNECTS. THE CUSTOMER WAS USING STANDARD BLACK CABLE FOR CASE. NO PUMP ON UNIT. THE PATIENT¿S PROCEDURE WAS RESCHEDULED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22100036 ↗MalfunctionFeb 13, 2025Break; Break
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 9611594-2025-00018 FOR THE FIRST REPORT. IT WAS REPORTED, "PATIENT REFERRED BY POLE ST HELIER (PSH) FOR RADIOLOGICAL GASTROTOMY ON (B)(6) 2024 IN A PATIENT WITH A NEUROLOGICAL HISTORY (FOLLOWING STROKE). NO IMMEDIATE COMPLICATIONS AFTER GPR ON THE SAME DAY, FOUND ON THE GROUND AT 10:30PM, OVER THE RIGHT BARRIER. FOUND FLAT ON HIS STOMACH. SAID TO HAVE PAIN ABOVE RIGHT EYEBROW. NO REDNESS, HEMATOMA, WOUND OR EDEMA. CALL TO DOCTOR ON CALL: POST-FALL MONITORING: NO PAIN, NO HEMATOMA, NO EDEMA, NO VOMITING. PUPILS SYMMETRICAL AND REACTIVE. CLEAR FOR THE REST OF THE NIGHT. ON (B)(6)2024 MORNING: COMPLAINS OF SHOULDER G-PAIN AT GPR LEVEL ADDITION OF ACUPAN THROUGH CATHETER AT 10AM. AT 10:30 A.M., 2 OUT OF 3 FIXATIONS HAD ALREADY FALLEN OUT. NO HEMATOMA, NO OOZING. RETURN TO POLE ST HELIER. PATIENT'S CURRENT CONDITION: CALL FROM REFERRING PHYSICIAN ON (B)(6) 2024 AS PATIENT PRESENTS WITH ABDOMINAL PAIN WITH GUARDING, BIOLOGICAL INFLAMMATORY SYNDROME. PRESCRIPTION FOR AN EMERGENCY ABDOMINAL
Patient problems: Abdominal Pain; Inflammation; Peritonitis
MAUDE report 21378825 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K192491. One submission often covers many models, so check the product and lot details in each record.
- Z-2622-2023 · Aug 1, 2023 · CompletedCOOLIEF RF Generator is an 80-watt Radio-Frequency (RF) Generator with four independent RF channels, allowing all RF channels to be activated individually or altogether. The system
Other Generator, lesion, radiofrequency devices
- NT2000iX™ RFG-NT-2000St. Jude Medical, Inc. · GENERATOR, LESION, RADIOFREQUENCY2 recalls310 reports
- NA TCN-10Boston Scientific Neuromodulation Corporation · GENERATOR, LESION, RADIOFREQUENCY2 recalls54 reports
- NT1100™ RFG-NT-1100St. Jude Medical, Inc. · GENERATOR, LESION, RADIOFREQUENCY107 reports
- NA TCN-15-3MBoston Scientific Neuromodulation Corporation · GENERATOR, LESION, RADIOFREQUENCY2 recalls6 reports
- IonicRF™ RFG-IONICSt. Jude Medical, Inc. · Generator, lesion, radiofrequency105 reports
- NA TCN-5Boston Scientific Neuromodulation Corporation · GENERATOR, LESION, RADIOFREQUENCY2 recalls1 report
- NA CSK-TC10-3MBoston Scientific Neuromodulation Corporation · GENERATOR, LESION, RADIOFREQUENCY1 recall34 reports
- NA CSK-TC15Boston Scientific Neuromodulation Corporation · GENERATOR, LESION, RADIOFREQUENCY1 recall22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page