Boston Scientific Neuromodulation Corporation · Nitinol TC Electrode, 15cm, 3m Cable · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00813250017519), received between Sep 19, 2017 and Jul 16, 2020 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Defective Device” (5 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Apr 29, 2026Class IIOpen, Classified

    Reusable electrodes may not meet expected performance levels.

    Z-2403-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 30, 2019Class IITerminated

    The firm initiated a voluntary correction of certain Cosman TCN Electrodes due to an incorrect revision of the Instructions for Use (IFU) being packaged with these products.

    Z-1697-2020 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Device5
  • Contamination /Decontamination Problem1
  • Material Discolored1

Patient problems

  • No Patient Involvement5
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2020Defective Device; Defective Device

    IT WAS REPORTED THAT HUB FELL OFF OF THE ELECTRODE. ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THE DEVICE HAS BEEN RETURNED AND EVALUATED. THE ALLEGATION OF "NOT READING TEMPERATURE" WAS CONFIRMED. THE SHAFT OF THE ELECTRODE WAS SEVERED AT THE HUB AND NOT RETURNED FOR ANALYSIS. THE PROBABLE CAUSE SELECTED IS "UNINTENDED USE ERROR CAUSED OR CONTRIBUTED TO EVENT". THE SEVERED SHAFT WAS LIKELY DUE TO UNINTENDED EXCESSIVE EXTERNAL MECHANICAL FORCE.

    Patient problems: No Patient Involvement

    MAUDE report 10283447 ↗
  • MalfunctionOct 15, 2019Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT AN RF ELECTRODE BECAME DETACHED FROM THE HUB. THE ELECTRODE WILL BE SENT BACK FOR ANALYSIS. ADDITIONAL INFORMATION WAS RECEIVED THAT THE DEVICE WAS NOT RETURNED FOR ANALYSIS.

    Patient problems: No Patient Involvement

    MAUDE report 9192863 ↗
  • MalfunctionOct 8, 2019Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT AN ELECTRODE SNAPPED AT THE BASE OF THE HUB AND FELL OUT.

    Patient problems: No Patient Involvement

    MAUDE report 9169807 ↗
  • MalfunctionOct 1, 2019Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT THE ELECTRODE FELL OUT OF THE HUB.

    Patient problems: No Patient Involvement

    MAUDE report 9142380 ↗
  • MalfunctionSep 10, 2019Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT AN ELECTRODE BECAME SEPARATED FROM THE HUB.

    Patient problems: No Patient Involvement

    MAUDE report 8981178 ↗
  • MalfunctionSep 19, 2017Material Discolored; Contamination /Decontamination Problem; Material Discolored; Contamination /Decontamination Problem

    DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODE SHOWED THAT THE EPOXY WAS DISCOLORED. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES, THE EPOXY BECOMES BRITTLE AND DISCOLORED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6877586 ↗

FDA clearance and classification

The FDA regulates product code GXD (generator, lesion, radiofrequency) under 21 CFR 882.4400 as a Class II device. 21 CFR 882.4400 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the NA TCN-15-3M

Has the NA TCN-15-3M been recalled?

Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-2403-2026 (Class II), began on Apr 29, 2026: Reusable electrodes may not meet expected performance levels. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NA TCN-15-3M?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 19, 2017 and Jul 16, 2020. In those reports, the most common specific device problems are “Defective Device” (5), “Contamination /Decontamination Problem” (1) and “Material Discolored” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA TCN-15-3M?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the NA TCN-15-3M, and is it still sold?

The FDA's device database lists Boston Scientific Neuromodulation Corporation as the company responsible for it (brand NA, model TCN-15-3M). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.