Skip to content

IonicRF™ RFG-IONICRecalls & reported problems

St. Jude Medical, Inc. · Generator · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude IonicRF radiofrequency generator is the power unit used in radiofrequency ablation procedures. Its 104 reports, 47 of them injuries, often describe grounding pad warnings, probes not being recognized, overheating and procedures being abandoned, and one report lists a third-degree burn. There were no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
105
Deaths reported
0
Injuries reported
48
Recalls naming it
0

Summary

The FDA's MAUDE database contains 105 adverse event reports that list this exact device (DI 05415067029317), received between Nov 17, 2020 and Sep 29, 2026 — classified as 48 injuries, 57 malfunctions.

The most frequently reported specific device problem is “Output Problem” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 4 reportsDec 2025: 1 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Output Problem27
  • Insufficient Heating24
  • Connection Problem10
  • Grounding Malfunction8
  • Communication or Transmission Problem7
  • Temperature Problem6
  • Adverse Event Without Identified Device or Use Problem5
  • Unexpected Shutdown4

Patient problems

  • No Clinical Signs, Symptoms or Conditions98
  • Muscle Weakness/Atrophy2
  • Pain2
  • Speech Disorder2
  • Burning Sensation1
  • Electric Shock1
  • Full thickness (Third Degree) Burn1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 105 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Device Remains Activated

    It was reported the patient displayed muscle spasms without the device being turned on. No adverse patient consequences were reported, but the procedure was abandoned after the probe had been inserted in the patient.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 50933628 ↗
  • InjurySep 24, 2026Unintended Electrical Shock

    It was reported that during ablation procedure, patient experienced unintended shocks. As result, the procedure was abandoned to address the issue.

    Patient problems: Pain

    MAUDE report 50889540 ↗
  • MalfunctionAug 28, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50503222 ↗
  • MalfunctionAug 11, 2026Communication or Transmission Problem

    It was reported the device was not recognizing the probes and overheating issue. The procedure was aborted as a result and there were no patient consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50212510 ↗
  • InjuryAug 7, 2026Grounding Malfunction

    It was reported that during an RFA procedure there was a grounding pad malfunction that was unable to be cleared, and the procedure was aborted. The patient experienced no consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50182318 ↗
  • InjuryMar 30, 2026Temperature Problem

    IT WAS REPORTED THAT DURING THE RF ABLATION THE PROBE TEMPERATURE DROPPED THEN SHOT UP AND THE MACHINE REGISTERED OVER TEMPERATURE ON CHANEL [X]. GROUNDING PAD DETACHED MESSAGE WAS ALSO REPORTED. TROUBLESHOOTING RESOLVED THE GROUNDING PAD MESSAGE, BUT THE TEMPERATURE MESSAGE WAS NOT RESOLVED. AS A RESULT, THE PROCEDURE WAS ABANDONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24720979 ↗
  • MalfunctionMar 13, 2026Insufficient Heating

    IT WAS REPORTED THAT DURING AN RFA PROCEDURE, A MESSAGE WAS SEEN DURING A PROCEDURE. TROUBLESHOOTING WAS DONE BUT THE ISSUE PERSISTED. THE PROCEDURE WAS ABANDONED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24587655 ↗
  • InjuryFeb 6, 2026Grounding Malfunction

    IT WAS REPORTED THAT THERE WAS A GROUNDING PAD MALFUNCTION DURING THE RFA PROCEDURE AND AS A RESULT, THE PROCEDURE WAS ABORTED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283425 ↗
  • InjuryJan 13, 2026Impedance Problem

    IT WAS REPORTED THAT DURING A PROCEDURE THE GENERATOR EXPERIENCED IMPEDANCE ERRORS. AS A RESULT, THE PROCEDURE WAS ABANDONED WITH NO CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24059728 ↗
  • InjuryJan 13, 2026Unexpected Shutdown

    IT WAS REPORTED THAT DURING AN RFA PROCEDURE, THE GENERATOR UNEXPECTEDLY SHUT DOWN, AND THE PROCEDURE WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24058958 ↗
  • InjuryJan 5, 2026Grounding Malfunction

    IT WAS REPORTED THAT THERE WAS A GROUNDING PAD MALFUNCTION DURING THE RFA PROCEDURE AND AS A RESULT, THE PROCEDURE WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23962906 ↗
  • MalfunctionDec 22, 2025Insufficient Heating; Insufficient Heating

    IT WAS REPORTED THE GENERATOR CANNOT REACH TARGET TEMPERATURE DURING RFA PROCEDURE, AND EVENTUALLY DISCONNECTED. AS A RESULT, THE PROCEDURE WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23873365 ↗
  • InjuryNov 26, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE GROUNDING PAD TEST FAILED REPEATEDLY. AS A RESULT, THE PROCEDURE WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23660483 ↗
  • MalfunctionNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY LESIONING PROCEDURE PATIENT EXPERIENCED A SUPERFICIAL SKIN BURN, SUBSEQUENTLY LEADING TO LOCALIZED SOFT TISSUE NECROSIS. AS A RESULT, EXTENDED WOUND CARE WAS DONE TO ADDRESS THE ISSUE. PATIENT IS STABLE.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 23584229 ↗
  • MalfunctionNov 12, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, AFTER 50 NERVE PULSES, THE DEVICE STOPPED WORKING WITH AN ERROR MESSAGE FAILING TO RECOGNIZE THE ELECTRODE. THE ELECTRODE WAS REPLACED BUT ISSUE PERSISTED. THE INSERTION PORTS WERE THEN SWITCHED HOWEVER HIGH IMPEDANCES WERE STILL PRESENT. THE PROCEDURE WAS THEN ABANDONED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23535988 ↗
  • MalfunctionNov 11, 2025Insufficient Heating; Insufficient Heating

    IT WAS REPORTED AN RFA PROCEDURE WAS ABANDONED DUE TO GENERATOR WOULD NOT REACH A SUFFICIENT TEMPERATURE AND WOULD NOT GENERATE A LESION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23521914 ↗
  • InjurySep 17, 2025Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT THE PATIENT'S PROCEDURE WAS ABANDONED DUE TO AN UNEXPECTED SHUTDOWN WITH THE GENERATOR. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23075874 ↗
  • InjuryAug 6, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS ABANDONED DUE TO CONNECTION ISSUES WITH THE GROUNDING PAD. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22734514 ↗
  • MalfunctionJul 21, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY ABLATION PROCEDURE ON (B)(6) 2025, THE GENERATOR WILL NOT RECOGNIZE THE GROUNDING PAD AFTER MULTIPLE ATTEMPTS. THE GENERATOR EXCEEDED MAXIMUM TEMPERATURE AND THE PROCEDURE WAS ABANDONED. THERE WERE NO CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22561457 ↗
  • MalfunctionJul 17, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT DURING AN RFA PROCEDURE, AN ERROR WAS DISPLAYED DUE TO GROUNDING PAD NOT ATTACHING. TROUBLESHOOTING WAS DONE BUT THE ISSUE PERSISTED. THE PROCEDURE WAS ABANDONED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22532417 ↗
  • InjuryJul 15, 2025Output Problem; Insufficient Device Problem Information; Output Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT AN UNKNOWN ERROR WITH THE IONIC GENERATOR CAUSED THE PROCEDURE TO BE ABANDONED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22511893 ↗
  • MalfunctionJun 26, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT DURING AN RFA PROCEDURE, MULTIPLE PROBES WERE NOT REACHING TEMPERATURE. TROUBLESHOOTING WAS DONE BUT THE ISSUE PERSISTED. THE PROCEDURE WAS ABANDONED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342816 ↗
  • MalfunctionJun 26, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT DURING AN RFA PROCEDURE, MULTIPLE PROBES WERE NOT REACHING TEMPERATURE. TROUBLESHOOTING WAS DONE BUT THE ISSUE PERSISTED. THE PROCEDURE WAS ABANDONED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342814 ↗
  • InjuryJun 11, 2025Insufficient Heating; Insufficient Heating

    IT WAS REPORTED THAT THE GENERATOR WAS UNABLE TO REACH THE CORRECT TEMPERATURE. THE PROCEDURE WAS ABANDONED AS A RESULT. NO ADVERSE CONSEQUENCES WERE REPORTED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22191393 ↗
  • InjuryJun 8, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT DURING A PROCEDURE. IT WAS ALSO REPORTED THAT THE GENERATOR HAD HIGH IMPEDANCES. NO OTHER ADVERSE EFFECTS WERE NOTED.

    Patient problems: Pain

    MAUDE report 22164475 ↗

Other Generator, lesion, radiofrequency devices

All Generator, lesion, radiofrequency devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page