AMX NavigateRecalls & reported problems
GE Medical Systems, L.L.C. · AMX Navigate Designated Medical Device Assembly · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00195278003348), received between Nov 7, 2025 and May 5, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Sharp Edges” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K210982); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Sharp Edges2
Patient problems
- Abrasion1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2026Sharp Edges
ON 06-APR-2026, A GE HEALTHCARE FIELD ENGINEER (FE) WAS ONSITE AT (B)(6) HOSPITAL (USA) PERFORMING SERVICE ON AN AMX NAVIGATE MOBILE X-RAY SYSTEM. DURING THE SERVICE ACTIVITY, THE FE OBSERVED THAT A FASTENER WAS MISSING FROM THE X-RAY TUBE COVER. THE MISSING FASTENER WAS CONFIRMED TO BE THE SAME COMPONENT ASSOCIATED WITH A PREVIOUSLY IDENTIFIED SENTINEL EVENT. NO PATIENT OR USER INJURY WAS REPORTED AT THE TIME OF DISCOVERY. BASED ON THE KNOWN ASSOCIATION WITH A PRIOR SENTINEL EVENT, THIS OCCURRENCE REPRESENTS A REPORTABLE MALFUNCTION, AS RECURRENCE OF THIS MALFUNCTION COULD POTENTIALLY LEAD TO A SERIOUS INJURY IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25080887 ↗InjuryNov 7, 2025Sharp Edges; Sharp Edges
ON (B)(6) 2025, A TECHNOLOGIST AT (B)(6) (USA) REPORTED AN INCIDENT INVOLVING THE AMX NAVIGATE MOBILE X-RAY SYSTEM. WHILE POSITIONING THE X-RAY TUBE FOR A PATIENT EXAMINATION, THE TECHNOLOGIST RIGHT MIDDLE FINGER WAS LACERATED BY A SHARP EDGE ON THE COLLIMATOR FACEPLATE. THE INJURY RESULTED IN A 1.5-INCH CUT, WHICH REQUIRED SIX STITCHES. BASED ON THE TREATMENT ADMINISTERED, THE INJURY IS CLASSIFIED AS SERIOUS.
Patient problems: Abrasion
MAUDE report 23498115 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K210982. One submission often covers many models, so check the product and lot details in each record.
- Z-2590-2021 · Jun 24, 2021 · TerminatedAMX Navigate Mobile X-Ray System
Other System, x-ray, mobile devices
- Philips MobileDiagnost wDR 2.2Sedecal · System, X-Ray, Mobile3 recalls7 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
- Optima XR240amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- Optima XR220amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile1 recall16 reports
- SEDECAL Soltus 500Sedecal · System, X-Ray, Mobile1 recall4 reports
- MobileDaRt Evolution 503-79999-01Shimadzu Corporation · System, X-Ray, Mobile1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.