Skip to content

NovoFine® Plus 32G 185650Recalls & reported problems

Novo Nordisk Inc. · NovoFine® Plus 32G needles are disposable, single use, sterile needles and intended for use with pen injector devices for the subcutaneous injection. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00301691856502), received between Dec 18, 2020 and Dec 18, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 18, 2020Break; Break

    CASE DESCRIPTION: CONCERNED AN ELDERLY MALE PATIENT (AGE NOT REPORTED). NAME: NOVOFINE PLUS 32G X 4MM, BATCH NUMBER: 20B14N. MICROSCOPIC EXAMINATION PERFORMED. A VISUAL EXAMINATION OF THE BACK NEEDLE OF THE RETURNED PRODUCT WAS PERFORMED. THE NEEDLES WERE TESTED FOR FLOW WITH MEASURE THREADS. NEEDLES TESTED FOR RESISTANCE TO BREAKAGE. THE PRODUCT WAS FOUND TO BE NORMAL. THE RESULTS WERE FOUND TO COMPLY WITH SPECIFICATIONS. THE BATCH DOCUMENTATION WAS REVIEWED. NO ABNORMALITIES RELATING TO THE OBSERVED PROBLEM WERE FOUND. DURING TEST IT HAS NOT BEEN POSSIBLE TO DETECT ANY IRREGULARITIES RELATED TO THE COMPLAINT ON THE UNUSED NEEDLES. THE BATCH DOCUMENTATION HAS BEEN REVIEWED AND FOUND TO BE NORMAL. FINAL MANUFACTURER'S COMMENT: 01-FEB-2021: FIVE SEALED NEEDLES WERE RETURNED TO NOVO NORDISK FOR EVALUATION. SINCE NO FAULTS WERE FOUND ON THE RETURNED NOVOFINE NEEDLES AND ONLY VERY LIMITED INFORMATION REGARDING THE PATIENTS HANDLING OF THE SUSPECTED DEVICE IS REPORTED IN THE CASE, IT IS NOT POSSIBLE TO ELUCIDATE A CLEAR ROOT CAUSE FOR THE EXPERIENCED ADVERSE EVENT AND THUS NOT POSSIBLE TO FIND SIMILAR INCIDENTS TO THE ONE REPORTED IN ARGUS CASE (B)(4). CONSIDERING THE NATURE OF THE EVEN

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 11039558 ↗

Other Needle, Hypodermic, Single Lumen devices