Skip to content

Foreign Body Retrieval Hood Protector 92810Recalls & reported problems

Avanos Medical, Inc. · Foreign Body Retrieval Hood Protector · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00350770928104), received between Sep 20, 2019 and Nov 5, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED "WHEN USING THE PROTECTIVE CAP ON THE ENDOSCOPE, IT BROKE OFF... THE PROTECTIVE CAP BROKE INSIDE THE PATIENT." ADDITIONAL INFORMATION RECEIVED 13-OCT-2025 STATING "PATIENT UNINJURED. NO FURTHER MEDICAL INTERVENTION REQUIRED. PATIENT DISCHARGED WITHOUT COMPLICATIONS."

    Patient problems: Foreign Body In Patient

    MAUDE report 23480448 ↗
  • MalfunctionSep 20, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED "THE FOREIGN BODY HOOD RETRIEVER CAME OFF OF THE SCOPE & LODGED IN THE [PATIENT'S] CRICOPHARYNGEUS". THERE WAS NO REPORTED INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9098095 ↗

Other endoscopic access overtube, gastroenterology-urology devices

All endoscopic access overtube, gastroenterology-urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page