Skip to content

Jelco 402310Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Jelco (100)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00351688066520), received between Dec 5, 2024 and Mar 25, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Disconnection” (4 reports).

4 other recalls cover devices cleared under the same FDA submission (K041399); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection4
  • Break1
  • Leak/Splash1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Break; Retraction Problem; Break; Retraction Problem

    IT WAS REPORTED THAT WHEN THE NEEDLE WAS RETRACTED A CLICK WAS HEARD BUT A PORTION OF THE NEEDLE REMAINED EXPOSED. IT WAS ALSO INFORMED THAT LIST INFORMING THAT IT HAS EXACT SAME FAILURE TO FULLY ENGAGE SAFELY AND BOTH RESULTED IN CLINICIAN'S NEEDLE STICKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21688890 ↗
  • MalfunctionMar 25, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT PEDIATRICS TEAM HAD BEEN EXPERIENCING NEEDLES FALLING OFF PREFILLED SYRINGES ¿ SPECIFICALLY FOR FLU SHOT SYRINGES. NEEDLES SEEMINGLY UNSCREWED THEMSELVES AND FELL OFF. THE EVENT OCCURRED DURING REMOVAL/ AT THE END OF THERAPY. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687820 ↗
  • MalfunctionMar 20, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT PEDIATRICS TEAM HAD BEEN EXPERIENCING NEEDLES FALLING OFF PREFILLED SYRINGES ¿ SPECIFICALLY FOR FLU SHOT SYRINGES. NEEDLES SEEMINGLY UNSCREWED THEMSELVES AND FELL OFF. THE EVENT OCCURRED DURING REMOVAL/ AT THE END OF THERAPY. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21659841 ↗
  • MalfunctionMar 20, 2025Disconnection; Disconnection

    IT WAS STATED THAT PEDIATRICS TEAM HAD BEEN EXPERIENCING NEEDLES FALLING OFF PREFILLED SYRINGES ¿ SPECIFICALLY FOR FLU SHOT SYRINGES. NEEDLES SEEMINGLY UNSCREWED THEMSELVES AND FELL OFF. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS NO PATIENT HARM OR ADVERSE EVENTS REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21657683 ↗
  • MalfunctionDec 5, 2024Disconnection; Leak/Splash; Disconnection; Leak/Splash

    IT WAS REPORTED THAT THE NEEDLES FELL OFF WHEN THEY WERE GIVEN ON PATIENTS. THIS HAPPENED AFTER THE ADMINISTRATION OF THE DEVICES AND THERE WERE TWO NEEDLES REPORTED HAVING THIS ISSUE. THE PATIENTS WERE REPORTED NOT TO BE IMPACTED. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20859330 ↗

FDA clearance and classification

Cleared through Abbreviated 510(k) K041399 on Jul 28, 2004 (decision: Substantially Equivalent). Submitted by Smiths Medical Asd, Inc. as “Hypodermic Needle-Pro Edge Needle Protection Device”. The FDA review took 63 days. FDA record ↗

The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041399. One submission often covers many models, so check the product and lot details in each record.

  • Z-1733-2019 · Feb 11, 2019 · Terminated
    Smith Medical JELCO Hypodermic Needle-Pro EDGE Safety Device Safety Device Rx Only Sterile Product Usage: Is a disposable, single use device intended for injection or aspiration o
  • Z-0897-2015 · Nov 21, 2014 · Terminated
    Hypodermic Needle Pro Edge" Safety Device 20G x1; Reorder Number: 402010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of flu
  • Z-0899-2015 · Nov 21, 2014 · Terminated
    Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended
  • Z-0898-2015 · Nov 21, 2014 · Terminated
    Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 10mL Luer lock syringe; Reorder Number: 4102010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intend

Common questions about the Jelco 402310

Has the Jelco 402310 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00351688066520), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Jelco 402310?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Dec 5, 2024 and Mar 25, 2025. In those reports, the most common specific device problems are “Disconnection” (4), “Break” (1) and “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Jelco 402310?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Jelco 402310 FDA approved?

It was cleared through Abbreviated 510(k) K041399 on Jul 28, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Jelco 402310, and is it still sold?

The FDA's device database lists Icu Medical, Inc. as the company responsible for it (brand Jelco, model 402310). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page