OneTouch Ultra Blue 022-896Recalls & reported problems
LifeScan U.S. LLC · OneTouch Ultra Test Strips Blue for blood glucose testing with: OneTouch Ultra 2, OneTouch UltraMini, OneTouch Ultra, OneTouch UltraSmart, OneTouch Ping and OneTouch UltraLink Meters · Class 2
All models of OneTouch Ultra Blue (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00353885009720), received between Jan 7, 2017 and Jun 17, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Bent1
Patient problems
- Blurred Vision1
- Chest Pain1
- Headache1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
DEVICE EVALUATION: THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS FAILED TESTING; THE REPORTED ISSUE WAS CONFIRMED. THE TEST STRIPS WERE FOUND TO HAVE SCRAP TOP TAPE OVER THE ELECTRODES. ON (B)(6) 2019, THE REPORTER CONTACTED LIFESCAN USA ALLEGING THAT ¿SOME¿ OF THE PATIENT¿S ONETOUCH ULTRA TEST STRIPS DIDN'T ¿HAVE BLACK AND WHITE STRIPS ON THE STRIP, ONLY BLUE LABEL¿. THIS COMPLAINT WAS INITIALLY RULED OUR BECAUSE INFORMATION OBTAINED DURING THE INITIAL CALL DID NOT REASONABLY SUGGEST A MALFUNCTION HAD OCCURRED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT AS A RESULT OF THE REPORTED ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8702771 ↗InjuryJan 7, 2017Bent; Bent
ON (B)(6) 2016, THE LAY USER/PATIENT CONTACTED LIFESCAN (LFS) TO REPORT THAT HIS ONETOUCH ULTRA TEST STRIPS WERE BENT AND PEELING. THE COMPLAINT WAS CLASSIFIED BASED ON CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE PATIENT REPORTED THAT THE ISSUE OCCURRED ¿2 TO 3 WEEKS¿ PRIOR TO CONTACTING LFS. THE PATIENT MANAGES HIS DIABETES BY SELF-ADJUSTING HIS INSULIN DOSES AND INCREASED THE DOSE OF HIS HUMALOG INSULIN BY 1-2 UNITS DURING THE DAY AND HIS LANTUS INSULIN BY 2 UNITS AT NIGHT IN RESPONSE TO THE ALLEGED ISSUE. THE PATIENT REPORTED THAT ¿30 MINUTES¿ AFTER THE ALLEGED ISSUE OCCURRED HE DEVELOPED SYMPTOMS OF ¿HEADACHES, BLURRED VISION, BOTHER WITH HIS LEGS AND CHEST PAINS¿ AND STATED THAT ¿A COUPLE OF WEEKS¿ PRIOR TO CONTACTING LFS HE VISITED HIS DOCTOR, WHO ¿REFERRED HIM TO ANOTHER DOCTOR¿. DURING THE CALL THE CCA NOTED THAT MORE THAN ONE STRIP IN THE VIAL WAS AFFECTED AND THEY WERE BENT AND PEELING PRIOR TO INSERTING THE STRIP INTO THE METER. THE PRODUCT WAS REPLACED AND REQUESTED BACK FOR EVALUATION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT REPORTEDLY DEVELOPED SIGNS AND/OR SYMPTOMS THAT MEET LFS¿ CRITERIA FOR A SERIOUS INJURY REPORTABLE ADVERSE EVENT AFTER THE ALLEGE
Patient problems: Blurred Vision; Chest Pain; Headache
MAUDE report 6231579 ↗
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Cholestech LDX™ 97991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page