FreeStyle Libre 3 72081Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Libre 3 Sensor Kit US · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Abbott FreeStyle Libre 3 sensor has 12,122 adverse event reports, with no deaths and 1,059 injuries. Many describe incorrect or low glucose readings, and some mention loss of consciousness or seizures. There are 2 open recalls: one from November 2025 for carbon build-up that may cause incorrect low readings, and one from July 2024 for erroneously high readings that could lead to larger insulin corrections. Reports rose from 971 in the prior 12 months to 9,011 in the latest 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 16,326
- Deaths reported
- 0
- Injuries reported
- 1,056
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 16,326 adverse event reports that list this exact device (DI 00357599818005), received between Oct 4, 2022 and Sep 30, 2026 — classified as 1,056 injuries, 15,268 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (10,299 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 24, 2025Open, Classified
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
Z-1020-2026 · Root cause: Process control · FDA record ↗
- Jul 24, 2024Open, Classified
Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.
Z-2692-2024 · Root cause: Process design · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings10,299
- Low Readings4,499
- Device Alarm System446
- Device Displays Incorrect Message380
- Unable to Obtain Readings364
- Detachment of Device or Device Component321
- Adverse Event Without Identified Device or Use Problem126
- Delivered as Unsterile Product117
Patient problems
- No Clinical Signs, Symptoms or Conditions15,139
- Hypoglycemia391
- Loss of consciousness375
- Device Embedded In Tissue or Plaque172
- Hyperglycemia156
- Convulsion/Seizure151
- Dizziness121
- Insufficient Information113
Recent adverse event reports
The 25 most recent of 16,326 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 54 mg/dL compared to glucose values of 277 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50951475 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 64 mg/dL compared to readings of 190 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947820 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947160 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 336.00 mg/dL compared to readings of 160.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946699 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 125.00 mg/dL compared to readings of 74.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 2 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946600 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 68.00 mg/dL compared to readings of 190.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944283 ↗MalfunctionSep 30, 2026Peeled/Delaminated
It was reported that the black plastic on the inside of the trays is delaminating.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944282 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 2.80 mmol/L compared to readings of 15.80 mmol/L respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944143 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 39 mg/dL compared to readings of 138 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943511 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 350.00 mg/dL compared to readings of 110.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 12 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931471 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 60.00 mg/dL compared to readings of 210.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931082 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 367 mg/dL compared to readings of 139 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929616 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 72.00 mg/dL compared to readings of 192.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929515 ↗MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of X mg/dL compared to readings of X mg/dL respectively obtained on a X and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was X days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918156 ↗MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915834 ↗MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL compared to readings of 166 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was unknown. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913196 ↗MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 70.00 mg/dL compared to readings of 180.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50911561 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906110 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50906109 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905307 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50.00 mg/dL compared to readings of 160.00 mg/dL respectively obtained on a Competitor Brand Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was unknown days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899146 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
The customer reported receiving erratic readings on their Abbott Diabetes Care device. It is unknown whether the customer noted a trend arrow on the device. The customer reported receiving readings of 300mg/dL, 50mg/dL and 200mg/dL within 10 minutes. The results when plotted on a Parkes Error Grid fell into the "C" zone showing the difference in values to be clinically significant. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899090 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 100 mg/dL compared to glucose values of 30 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898585 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 2.50 mmol/L, 2.90 mmol/L compared to readings of 8.30 mmol/L, 8.70 mmol/L respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898174 ↗MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 221 mg/dL compared to readings of 110 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50897018 ↗
Other Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems devices
- FreeStyle Libre 2 71992Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 death23,063 reports
- FreeStyle Libre 2 72000Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems453 reports
- FreeStyle Libre 3 72080Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall274 reports
- FreeStyle Libre 2 71951Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls72 reports
- FreeStyle Libre 3 App 72075-01Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall23 reports
- Bigfoot Unity Diabetes Management System ASM-10046Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall12 reports
- Bigfoot Unity Diabetes Management System ASM-10053Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall0 reports
- Bigfoot Unity Diabetes Management System FG-300171Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.