Skip to content

FreeStyle Libre 2 71992Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Libre 2 Flash Glucose Monitoring System Sensor · Class 2

All models of FreeStyle Libre 2 (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This FreeStyle Libre 2 sensor has 17,204 adverse event reports in the FDA database, including 1 death and 4,754 injuries. Reports mostly describe incorrect or imprecise glucose readings, sensors that cannot give readings, wrong on-screen messages, and sensors that detach. Reports in the last 12 months (3,284) are far higher than in the 12 months before (905). No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23,063
Deaths reported
1
Injuries reported
4,759
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23,063 adverse event reports that list this exact device (DI 00357599800000), received between Sep 18, 2020 and Sep 30, 2026 — classified as 1 death, 4,759 injuries, 18,303 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (16,855 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 293 reportsDec 2023: 243 reportsJan 2024: 249 reportsFeb 2024: 237 reportsMar 2024: 208 reportsApr 2024: 281 reportsMay 2024: 238 reportsJun 2024: 240 reportsJul 2024: 220 reportsAug 2024: 236 reportsSep 2024: 233 reportsOct 2024: 215 reportsNov 2024: 135 reportsDec 2024: 69 reportsJan 2025: 132 reportsFeb 2025: 78 reportsMar 2025: 78 reportsApr 2025: 73 reportsMay 2025: 83 reportsJun 2025: 80 reportsJul 2025: 82 reportsAug 2025: 52 reportsSep 2025: 19 reportsOct 2025: 24 reportsNov 2025: 6 reportsDec 2025: 4 reportsJan 2026: 10 reportsFeb 2026: 15 reportsMar 2026: 25 reportsApr 2026: 48 reportsMay 2026: 33 reportsJun 2026: 36 reportsJul 2026: 609 reportsAug 2026: 2,504 reportsSep 2026: 5,853 reportsOct 2026: 0 reports2024202520265,853
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings16,855
  • Device Displays Incorrect Message2,545
  • Unable to Obtain Readings2,516
  • Detachment of Device or Device Component1,231
  • Device Alarm System834
  • Low Readings714
  • High Readings392
  • Adverse Event Without Identified Device or Use Problem240

Patient problems

  • No Clinical Signs, Symptoms or Conditions17,639
  • Loss of consciousness2,262
  • Hypoglycemia2,222
  • Convulsion/Seizure768
  • Hyperglycemia763
  • Device Embedded In Tissue or Plaque622
  • Dizziness485
  • Insufficient Information473

Recent adverse event reports

The 25 most recent of 23,063 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Detachment of Device or Device Component

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 50955132 ↗
  • InjurySep 30, 2026Unable to Obtain Readings; Device Displays Incorrect Message

    Patient problems: Hypoglycemia

    MAUDE report 50953813 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 200 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 4 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946446 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 64.00 mg/dL compared to readings of 172.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943968 ↗
  • MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50929337 ↗
  • MalfunctionSep 28, 2026Detachment of Device or Device Component

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 50926223 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926045 ↗
  • MalfunctionSep 28, 2026Low Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925952 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925951 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925292 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of "[130mg/dL]" compared to glucose values of "[60mg/dL]" respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the "[C]" Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914871 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 52 mg/dL compared to glucose values of 240 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912315 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of "[100 mg/dL]" compared to glucose values of "[55 mg/dL]" respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the "[C]" Zone, showing the difference in values to be clinically significant. The length of sensor wear was "[6 days]". There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911951 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL compared to readings of 188 mg/dL respectively obtained on a Competitor Brand Sensor and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50911560 ↗
  • MalfunctionSep 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906384 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906322 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906320 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906274 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906212 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906191 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906091 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50906025 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904236 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904235 ↗
  • MalfunctionSep 25, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903699 ↗

Other Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems devices

All Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page