Bigfoot Unity Diabetes Management System ASM-10046Recalls & reported problems

Bigfoot Biomedical, Inc. · Bigfoot Unity Cap for Novolog-Aspart · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00850003506173), received between Jul 23, 2025 and Jul 25, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (12 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 12 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2025Class IIOpen, Classified

    Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision.

    Z-1997-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Freezes or Fails to Launch12

Patient problems

  • No Clinical Signs, Symptoms or Conditions12

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22615983 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22615914 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22613161 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612227 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22610115 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22606795 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22606677 ↗
  • MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22606378 ↗
  • MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22600739 ↗
  • MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22594165 ↗
  • MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22591332 ↗
  • MalfunctionJul 23, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22583834 ↗

Other Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems devices

All Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems devices →