Bigfoot Unity Diabetes Management System ASM-10046Recalls & reported problems
Bigfoot Biomedical, Inc. · Bigfoot Unity Cap for Novolog-Aspart · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00850003506173), received between Jul 23, 2025 and Jul 25, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (12 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 21, 2025Class IIOpen, Classified
Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not have all information necessary to make a timely treatment decision.
Z-1997-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch12
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22615983 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22615914 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22613161 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22612227 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22610115 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606795 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606677 ↗MalfunctionJul 25, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22606378 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22600739 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22594165 ↗MalfunctionJul 24, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22591332 ↗MalfunctionJul 23, 2025Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
A CUSTOMER REPORTED RECEIVING DEVICE ERROR FROM A BIGFOOT UNITY DEVICE. THIS IS A KNOWN ISSUE FOR THE SERIAL NUMBER ASSOCIATED WITH THIS COMPLAINT, ADDRESSED BY THE FIELD ACTION FA-001 (FDA RECALL 3016525500/05162025/C). THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22583834 ↗
Other Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems devices
- FreeStyle Libre 2 71992Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 death23,063 reports
- FreeStyle Libre 3 72081Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls16,326 reports
- FreeStyle Libre 2 72000Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems453 reports
- FreeStyle Libre 3 72080Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall274 reports
- FreeStyle Libre 2 71951Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls72 reports
- FreeStyle Libre 3 App 72075-01Abbott Diabetes Care Inc · Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 recall23 reports
- Bigfoot Unity Diabetes Management System FG-300173Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall0 reports
- Bigfoot Unity Diabetes Management System FG-300175Bigfoot Biomedical, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated, Not For Use With Automated Insulin Delivery Systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page