Accu-Chek Guide 07398794001Recalls & reported problems

Roche Diagnostics Operations, Inc. · ACCU-CHEK GUIDE MG/DL METER 923 (US) · Class 2

All models of ACCU-CHEK GUIDE (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This is the Accu-Chek Guide blood glucose meter, a home testing device that measures sugar in a drop of blood. The 61 reports mostly describe the screen showing the wrong unit of measurement, mmol/L instead of mg/dL, along with meters that would not power up and hard-to-read or flashing displays. Seven reports describe injuries and there are no deaths. No recalls name this device, and reports have dropped sharply, from 8 in the prior 12 months to none in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
61
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 61 adverse event reports that list this exact device (DI 00365702609105), received between Feb 24, 2020 and Jul 29, 2025 — classified as 7 injuries, 54 malfunctions.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (49 reports).

2 other recalls cover devices cleared under the same FDA submission (K160944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 23 reportsAug 2024: 18 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem49
  • Adverse Event Without Identified Device or Use Problem4
  • Display Difficult to Read2
  • Failure to Power Up2
  • Flare or Flash1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Insufficient Device Problem Information1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions53
  • Hypoglycemia4
  • Hyperglycemia2
  • Elevated Ketones/ Diabetic Ketoacidosis1
  • Headache1
  • Heart Failure/Congestive Heart Failure1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 61 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE UNIT OF MEASURE ON THE BLOOD GLUCOSE MONITOR WAS DISPLAYED IN MMOL/L (FOR EXAMPLE, 8.3). THE UNIT OF MEASURE ON THE BACK OF THE METER IS IN MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22640940 ↗
  • MalfunctionFeb 19, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS LABELED WITH AN INCORRECT UNIT OF MEASURE. THE METER WAS SUPPOSEDLY READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21408229 ↗
  • InjuryJan 3, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS UNAVAILABLE FOR USE DUE TO POWER CONCERNS. THE POWER ISSUE ON THE METER STARTED THE DAY PRIOR TO THE EVENT, THEREFORE THE USER WAS UNABLE TO TEST. IT WAS REPORTED THAT THE CUSTOMER TOOK ALL OF HER MEDICATIONS AS NORMAL, EXCEPT HER NOVOLOG INSULIN, PRIOR TO THE EVENT. THE USER REPORTED THAT HER NOVOLOG INSULIN IS ON A SLIDING SCALE BASED ON HER BLOOD GLUCOSE RESULT, AND SINCE SHE WAS UNABLE TO TEST, SHE WITHHELD HER NOVOLOG. ON THE MORNING OF THE INCIDENT AT APPROXIMATELY 6:00AM, THE CUSTOMER EXPERIENCED A SEVERE HEADACHE. AT 7:00AM SHE ATTEMPTED TO TEST ON HER GUIDE METER AND IT STILL DID NOT HAVE POWER. THE USER'S DAUGHTER CALLED THE EMT'S. UPON ARRIVAL, THE EMT'S RECEIVED A BLOOD GLUCOSE RESULT OF 427 MG/DL ON THEIR PROFESSIONAL METER. THE USER WAS TREATED WITH OXYGEN AND 18 UNITS OF NOVOLOG. THE EMT'S STAYED WITH THE CUSTOMER FOR 30-45 MINUTES, UNTIL SHE RECOVERED. THE USER WAS ONLY TREATED AT HOME AND WAS NOT TRANSPORTED TO THE HOSPITAL.

    Patient problems: Headache

    MAUDE report 21069082 ↗
  • MalfunctionDec 12, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR DISPLAYED THE INCORRECT UNIT OF MEASURE FOR WHICH IT WAS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20918467 ↗
  • MalfunctionDec 12, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR DISPLAYED THE INCORRECT UNIT OF MEASURE FOR WHICH IT WAS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20915473 ↗
  • MalfunctionNov 27, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE CUSTOMER COULD NOT BE REACHED FOR ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20798565 ↗
  • MalfunctionNov 14, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THE PATIENT RECEIVED THE BLOOD GLUCOSE MONITOR IN THE WRONG UNIT OF MEASURE. THE CONFIGURED UNIT OF THE METER IS MMOL/L BUT THE UNIT ON THE LABEL IS MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20683233 ↗
  • MalfunctionNov 5, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR DISPLAYED THE INCORRECT UNIT OF MEASURE FOR WHICH IT WAS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20608421 ↗
  • MalfunctionOct 15, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20457630 ↗
  • MalfunctionSep 16, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR DISPLAYED THE INCORRECT UNIT OF MEASURE FOR WHICH IT WAS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20226522 ↗
  • MalfunctionSep 11, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20189229 ↗
  • MalfunctionAug 28, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR DISPLAYED A RESULT IN MMOL/L INSTEAD OF MG/DL. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20086694 ↗
  • MalfunctionAug 23, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE USER REPORTED THE SERIAL NUMBER OF HER DEVICE WAS LISTED IN THE RECALL LETTER FOR THE BLOOD GLUCOSE MONITOR READING IN THE INCORRECT UNIT OF MEASURE. THE USER DISCARDED THE DEVICE AND COULD NOT PROVIDE ANY INFORMATION OFF OF THE BACK LABEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20046814 ↗
  • MalfunctionAug 22, 2024Display Difficult to Read; Display Difficult to Read

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20039504 ↗
  • MalfunctionAug 15, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE UNIT OF MEASURE WAS INCORRECT ON THE BLOOD GLUCOSE MONITOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19995330 ↗
  • MalfunctionAug 15, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19995182 ↗
  • MalfunctionAug 9, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19952803 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED, THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE. FOR WHICH, IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19919350 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917265 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917260 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L. BASED ON THE SERIAL NUMBER, THE BLOOD GLUCOSE MONITOR SHOULD BE READING IN MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917259 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917244 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917235 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917228 ↗
  • MalfunctionAug 6, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE BLOOD GLUCOSE MONITOR WAS READING IN THE INCORRECT UNIT OF MEASURE FOR WHICH IT IS LABELED. THE METER WAS READING IN MMOL/L INSTEAD OF MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917216 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160944. One submission often covers many models, so check the product and lot details in each record.

  • Z-2174-2024 · May 13, 2024 · Open, Classified
    Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectivenes
  • Z-0747-2020 · Sep 30, 2019 · Terminated
    Blood Glucose Monitoring System, Model Number 07400870001 ACCU-CHEK GUIDE SAMPLE KIT MG/DL - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quant

Other System, Test, Blood Glucose, Over The Counter devices

All System, Test, Blood Glucose, Over The Counter devices →