Skip to content

Alcon, TurboSonics, MicroTip 30KTSRecalls & reported problems

Alcon Laboratories, Inc. · 30° Kelman®, 0.9 mm TurboSonics® MicroTip™ Tip · Class 2

All models of Alcon, TurboSonics, MicroTip (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380657403042), received between Nov 7, 2021 and Nov 7, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 28, 2023Class IIOpen, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2410-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 7, 2021Break; Suction Failure; Break; Suction Failure

    A PHYSICIAN REPORTED THE CATARACT MATERIALS WERE NOT EMULSIFYING AND THE FRAGMENTS WERE NOT GETTING ASPIRATED DURING A CATARACT PROCEDURE. WHEN THE SURGEON CHECKED IT, HE FOUND THE PHACOEMULSIFICATION (PHACO) TIP HAD AN UNEVEN SHAPE AND IT WAS CRACKED AT THE BENT AREA. ANOTHER PHACO TIP WAS OBTAINED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12763853 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alcon, TurboSonics, MicroTip 30KTS

Has the Alcon, TurboSonics, MicroTip 30KTS been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2410-2023 (Class II), began on Jun 28, 2023: Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Alcon, TurboSonics, MicroTip 30KTS?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 7, 2021. That report lists device problems “Break” (1) and “Suction Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alcon, TurboSonics, MicroTip 30KTS?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alcon, TurboSonics, MicroTip 30KTS, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Alcon, TurboSonics, MicroTip, model 30KTS). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page