Skip to content

Alcon, TurboSonics, Flared ABS 8065740806Recalls & reported problems

Alcon Laboratories, Inc. · 30° Round, 1.1mm TurboSonics® Flared ABS® Tip · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00380657408061), received between Jun 30, 2023 and Oct 17, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2023Class IIOpen, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2410-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break5
  • Material Twisted/Bent1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 17, 2025Break; Break

    A HA SYSTEM REPORTED THAT SLEEVE OF THE PHACOEMULSIFICATION TIP RUPTURED DURING CATARACT WITHOUT INTRAOCULAR LENS IMPLANT SURGERY. THE PROCEDURE WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23321308 ↗
  • MalfunctionApr 21, 2025Break; Break

    A CUSTOMER REPORTED VIA HEALTH AUTHORITY THAT THE PHACOEMULSIFICATION TIP HAD BROKEN DURING CATARACT SURGERY (WITH INTRAOCULAR LENS IMPLANT (IOL)). THE SURGERY WAS SCHEDULED FOR RIGHT EYE OF THE PATIENT. THE SURGERY WAS COMPLETED AFTER REPLACEMENT OF TIP WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21875030 ↗
  • MalfunctionMay 2, 2024Break; Break

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT THE TIP WAS SUDDENLY BROKEN AND THE SURGERY COULD NOT BE CONTINUED DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANT (IOL). AFTER REPLACING THE HANDPIECE AND PHACO TIP, THE SURGERY COMPLETED SMOOTHLY, THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19234154 ↗
  • MalfunctionApr 5, 2024Break; Material Twisted/Bent; Break; Material Twisted/Bent

    A HEALTH CARE PROFESSIONAL REPORTED THAT PHACO TIP WAS BENT AND WAS BROKEN DURING CATARACT SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19051626 ↗
  • MalfunctionJul 24, 2023Suction Failure; Suction Failure

    A CUSTOMER REPORTED THAT ASPIRATION WAS ABLE TO PERFORM NORMALLY AT UNKNOWN TIMING OF CATARACT SURGERY. SURGERY COMPLETED WITH THE REPLACEMENT OF THE TIP. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17377344 ↗
  • MalfunctionJun 30, 2023Break; Break

    A CUSTOMER REPORTED THAT BROKEN NIPPLE FOUND DURING THE CATARACT SURGERY. ANOTHER SET OF NIPPLE HANDLES REPLACED, AND THE SURGERY WAS COMPLETED NORMALLY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17235650 ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page