TurboSonics, Microsmooth, MicroTip, Alcon 8065750159Recalls & reported problems
Alcon Laboratories, Inc. · TurboSonics Microsmooth MicroTip Parts Kit · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380657501595), received between Jan 9, 2025 and Dec 23, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Contamination” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination3
- Product Quality Problem1
Patient problems
- Endophthalmitis3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 23, 2025Contamination; Contamination
A PHYSICIAN REPORTED THAT THE PRODUCT WAS CONTAMINATED, AND IT CAUSED ENDOPHTHALMITIS IN THREE PATIENTS AFTER PHAKIC INTRAOCULAR LENS IMPLANTATION SURGERY. THE PROCEDURE WAS COMPLETED WITHOUT REPLACEMENT SINCE THERE WAS NO PRODUCT MALFUNCTION. THIS PATIENT WAS TREATED WITH ONE INTRAVITREAL INJECTION. THE CURRENT PATIENT CONDITION IS UNKNOWN. THIS REPORT PERTAINS TO THE SECOND OF THREE PATIENTS INVOLVED IN THESE REPORTED EVENTS. FURTHER INFORMATION HAS BEEN REQUESTED. NONE RECEIVED TILL DATE.
Patient problems: Endophthalmitis
MAUDE report 23874726 ↗InjuryDec 22, 2025Contamination; Contamination
A PHYSICIAN REPORTED THAT THE PRODUCT WAS CONTAMINATED, AND IT CAUSED ENDOPHTHALMITIS IN THREE PATIENTS AFTER PHAKIC INTRAOCULAR LENS IMPLANTATION SURGERY. THE PROCEDURE WAS COMPLETED WITHOUT REPLACEMENT SINCE THERE WAS NO PRODUCT MALFUNCTION. THIS PATIENT WAS TREATED WITH THREE INTRAVITREAL INJECTIONS. THE CURRENT PATIENT CONDITION IS UNKNOWN. THIS REPORT PERTAINS TO THE THIRD OF THREE PATIENTS INVOLVED IN THESE REPORTED EVENTS. FURTHER INFORMATION HAS BEEN REQUESTED. NONE RECEIVED TILL DATE.
Patient problems: Endophthalmitis
MAUDE report 23863828 ↗InjuryNov 18, 2025Contamination; Contamination
ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THE EVENT OCCURRED AFTER PHAKIC INTRAOCULAR LENS IMPLANTATION SURGERY. THE PROCEDURE WAS COMPLETED WITHOUT REPLACEMENT SINCE THERE WAS NO PRODUCT MALFUNCTION. VITRECTOMY PROCEDURE WAS PERFORMED TO THE PATIENT AS INFECTION WAS NOT UNDER CONTROL THROUGH MEDICATIONS. THIS REPORT PERTAINS TO THE FIRST OF THREE PATIENTS INVOLVED IN THESE REPORTED EVENTS. FURTHER INFORMATION HAS BEEN REQUESTED. NONE RECEIVED TILL DATE.
Patient problems: Endophthalmitis
MAUDE report 23579362 ↗MalfunctionJan 9, 2025Product Quality Problem; Product Quality Problem
A PHYSICIAN REPORTED THAT STIFFNESS OF A SLEEVE FELT WEAK AND ITS THICKNESS FELT THIN DURING CATARACT SURGERY WITH INTRA OCULAR LENS IMPLANT. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21120976 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page