Infinit, Microsmooth, Alcon 8065750517Recalls & reported problems
Alcon Laboratories, Inc. · 0.9mm Microsmooth NULTRA Infusion Sleeve Kit · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00380657505173), received between Oct 19, 2021 and Jan 1, 2024 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Infusion or Flow Problem6
- Material Discolored5
- Material Too Soft/Flexible5
- Adverse Event Without Identified Device or Use Problem1
- Leak/Splash1
- Suction Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Eye Burn1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 1, 2024Leak/Splash; Leak/Splash
A PHARMACIST REPORTED THAT THE LEAK WAS OBSERVED AT THE BASE OF THE SLEEVE DURING THE CATARACT WITH INTRAOCULAR LENS IMPLANT SURGERY. THE PROCEDURE WAS COMPLETED AFTER THE SLEEVE WAS REPLACED WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18423079 ↗MalfunctionDec 27, 2023Break; Infusion or Flow Problem; Break; Infusion or Flow Problem
AS A RESULT OF AN INTERNAL REVIEW OF THE FILE, IT WAS DETERMINED THAT THE REPORTED PRODUCT ISSUE FOOT SWITCH CABLE AGED CAUSING VALVE DAMAGE AND INABILITY TO PERFUSION DOES NOT REPRESENT A REPORTABLE MALFUNCTION AS PER APPLICABLE MEDICAL DEVICE REGULATIONS. THERE IS NO EVIDENCE OF A LIFE-THREATENING INJURY/ILLNESS OR PERMANENT IMPAIRMENT/DAMAGE, THERE HAS BEEN NO MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE PERMANENT IMPAIRMENT/DAMAGE AND THE INFORMATION PROVIDED DOES NOT REPRESENT A PRODUCT PROBLEM THAT IS LIKELY TO CAUSE THE ABOVE-MENTIONED EVENTS IF THE PRODUCT PROBLEM WERE TO RECUR.
Patient problems: Insufficient Information
MAUDE report 18396128 ↗MalfunctionDec 26, 2023Suction Failure; Suction Failure
A CUSTOMER REPORTED THAT THE CONTROL OF PHACOEMULSIFICATION ENERGY SUCTION FAILED, RESULTING IN AN EXTENDED SURGICAL TIME AND INCREASED INTRAOPERATIVE RISK DURING CATARACT SURGERY (WITH INTRAOCULAR LENSES IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: Insufficient Information
MAUDE report 18393293 ↗InjuryMar 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A NURSE REPORTED THAT OPHTHALMIC OPERATING CONSOLE, HANDPIECE AND TIP WAS USED DURING THE PROCEDURE WHICH CAUSED THE THERMAL INJURY WHICH REQUIRE THE SUTURES. PATIENT CURRENT CONDITION WAS NOT PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ADDITIONAL INFORMATION RECEIVED STATING THAT PATIENT HAD A CORNEAL LEAK DURING THE CATARACT SURGERY WHICH COVERED WITH SEALANT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Eye Burn
MAUDE report 16563417 ↗MalfunctionNov 19, 2021Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible; Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible
A HEALTHCARE PROFESSIONAL REPORTED THE SLEEVES WERE NOT THE SAME AS THEY HAD DIFFERENT SHADES OF COLOR, THE LIGHTER ONES WERE THINNER, BROKE MORE EASILY WHEN THEY WERE BEING PLACED IN THE PHACOEMULSIFICATION (PHACO) NEEDLES, GET CURLED ESPECIALLY WHEN TRYING TO ENTER THE INCISION, AND THERE WAS A SIGNIFICANT REDUCTION IN INFUSION INTO THE EYES. THERE WAS NO PATIENT HARMS REPORTED. THIS IS ONE OF MULTIPLE REPORTS FOR THESE REPORTED EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12843923 ↗MalfunctionNov 19, 2021Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible; Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible
A HEALTHCARE PROFESSIONAL REPORTED THE SLEEVES WERE NOT THE SAME AS THEY HAD DIFFERENT SHADES OF COLOR, THE LIGHTER ONES WERE THINNER, BROKE MORE EASILY WHEN THEY WERE BEING PLACED IN THE PHACOEMULSIFICATION (PHACO) NEEDLES, GET CURLED ESPECIALLY WHEN TRYING TO ENTER THE INCISION, AND THERE WAS A SIGNIFICANT REDUCTION IN INFUSION INTO THE EYES. THERE WAS NO PATIENT HARMS REPORTED. THIS IS ONE OF MULTIPLE REPORTS FOR THESE REPORTED EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12843918 ↗MalfunctionNov 19, 2021Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible; Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible
A HEALTHCARE PROFESSIONAL REPORTED THE SLEEVES WERE NOT THE SAME AS THEY HAD DIFFERENT SHADES OF COLOR, THE LIGHTER ONES WERE THINNER, BROKE MORE EASILY WHEN THEY WERE BEING PLACED IN THE PHACOEMULSIFICATION (PHACO) NEEDLES, GET CURLED ESPECIALLY WHEN TRYING TO ENTER THE INCISION, AND THERE WAS A SIGNIFICANT REDUCTION IN INFUSION INTO THE EYES. THERE WAS NO PATIENT HARMS REPORTED. THIS IS ONE OF MULTIPLE REPORTS FOR THESE REPORTED EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12843829 ↗MalfunctionNov 19, 2021Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible; Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible
A HEALTHCARE PROFESSIONAL REPORTED THE SLEEVES WERE NOT THE SAME AS THEY HAD DIFFERENT SHADES OF COLOR, THE LIGHTER ONES WERE THINNER, BROKE MORE EASILY WHEN THEY WERE BEING PLACED IN THE PHACOEMULSIFICATION (PHACO) NEEDLES, GET CURLED ESPECIALLY WHEN TRYING TO ENTER THE INCISION, AND THERE WAS A SIGNIFICANT REDUCTION IN INFUSION INTO THE EYES. THERE WAS NO PATIENT HARMS REPORTED. THIS IS ONE OF MULTIPLE REPORTS FOR THESE REPORTED EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12843810 ↗MalfunctionOct 19, 2021Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible; Break; Material Discolored; Infusion or Flow Problem; Material Too Soft/Flexible
A HEALTHCARE PROFESSIONAL REPORTED SEVERAL CUSTOMERS INDICATED TWO SLEEVES WERE NOT THE SAME AS THEY HAD DIFFERENT SHADES OF COLOR, THE LIGHTER ONES WERE THINNER, BROKE MORE EASILY WHEN THEY WERE BEING PLACED IN THE PHACOEMULSIFICATION(PHACO) NEEDLES, GET CURLED ESPECIALLY WHEN TRYING TO ENTER THE INCISION, AND THERE WAS A SIGNIFICANT REDUCTION IN INFUSION INTO THE EYES. THERE WAS NO PATIENT HARMS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12657721 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page