Skip to content

Constellation 8065750888Recalls & reported problems

Alcon Laboratories, Inc. · Constellation Handpiece · Class 2

All models of Constellation (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380657508884), received between Jun 9, 2022 and Sep 26, 2023 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Overheating of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Overheating of Device2
  • Energy Output Problem1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Capsular Bag Tear1
  • Eye Injury1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 26, 2023Energy Output Problem; Energy Output Problem

    A CUSTOMER REPORTED THAT DURING A CATARACT SURGERY PHACOEMULSIFICATION HANDPIECE FAILED TO BREAK CATARACT EFFICIENTLY AND PATIENT'S NUCLEUS DROPPED POSTERIORLY. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE RECEIVED TILL DATE. THIS IS THREE OF FOUR REPORTS FOR THIS ISSUE AND REPRESENTS THE PHACOEMULSIFICATION HANDPIECE.

    Patient problems: Capsular Bag Tear

    MAUDE report 17815846 ↗
  • InjuryJul 15, 2022Overheating of Device; Overheating of Device

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY AN PHACOEMULSIFICATION HANDPIECE TIP WAS HOT AND A THERMAL INJURY OCCURRED TO THE PATIENT IN THE CONJUNCTIVA AND SCLERA PERISCLEROTOMY (COAGULATION). THE SUTURES ALONG WITH USUAL POSTOPERATIVE TREATMENT WERE USED TO SEAL THE WOUND.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Eye Injury

    MAUDE report 15025654 ↗
  • MalfunctionJun 9, 2022Overheating of Device; Overheating of Device

    A PHYSICIAN REPORTED THAT DURING A VITRECTOMY PROCEDURE AN OPHTHALMIC PHACOEMULSIFICATION HANDPIECE TIP WAS GETTING HEATED . PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14641512 ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page