Constellation 8065751155Recalls & reported problems
Alcon Laboratories, Inc. · Constellation Vision System 0.9 mm Tipless Phaco Pak · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Alcon Constellation Phaco Pak is a disposable tubing set used in cataract surgery. Its 219 reports often describe suction failing, fluid leaking, and the system being unable to irrigate. There were 14 injuries and no deaths. Reports fell in the last 12 months, with 16 compared with 54 in the year before. There is 1 open recall, which began in April 2025 over some units whose sealed cover may be damaged, putting sterility at risk.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 220
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 220 adverse event reports that list this exact device (DI 00380657511556), received between Mar 19, 2021 and Sep 23, 2026 — classified as 14 injuries, 206 malfunctions.
The most frequently reported specific device problem is “Suction Failure” (56 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 28, 2025Open, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.
Z-2109-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Suction Failure56
- Fluid/Blood Leak40
- Leak/Splash40
- Suction Problem28
- Inability to Irrigate17
- Air/Gas in Device14
- Obstruction of Flow9
- Adverse Event Without Identified Device or Use Problem6
Patient problems
- No Clinical Signs, Symptoms or Conditions182
- Insufficient Information30
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available9
- Toxic Anterior Segment Syndrome (TASS)5
- Uveitis5
- Intraocular Pressure Decreased3
- Capsular Bag Tear2
- Endophthalmitis1
Recent adverse event reports
The 25 most recent of 220 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Suction Problem
A physician reported that ultrasound could not be used due to high aspiration resistance during the surgery. The procedure type was unknown. The surgery was completed on the same day after replacing the product with another one. There was no patient contact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870520 ↗MalfunctionAug 24, 2026Suction Problem
A physician reported that, at the time of surgery, the cassette was found to be defective as it was unable to aspirate. The procedure type was unknown. The procedure was completed on same day after replacing the product with new one. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50401854 ↗MalfunctionMar 3, 2026Obstruction of Flow
A NURSE REPORTED THAT ASPIRATION TUBING WAS BLOCKED DURING CATARACT SURGERY (WITHOUT INTRAOCULAR LENSES IMPLANT (IOL)). THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24493676 ↗MalfunctionFeb 25, 2026Material Separation
A HEALTH CARE PROFESSIONAL REPORTED THAT FLEXIBLE TUBES WAS DISCONNECTED FROM THE METAL TIPS AND STERILIZED DURING CATARACT SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING IT WITH A NEW PRODUCT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442547 ↗MalfunctionFeb 13, 2026Leak/Splash
A PHYSICIAN REPORTED THAT THE LEAKAGE OCCURRED FROM THE CASSETTE DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE PROCEDURE WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT CONTACT AND NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330509 ↗MalfunctionFeb 3, 2026Suction Problem
A PHYSICIAN REPORTED THAT THE ASPIRATION DID NOT WORK DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE PROCEDURE WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24243627 ↗MalfunctionJan 30, 2026Leak/Splash
A NURSE REPORTED THAT CASSETTE HAD LEAKED WATER DURING CATARACT WITHOUT INTRA OCULAR LENS IMPLANT SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24223091 ↗InjuryJan 12, 2026Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AN ERROR MESSAGE WAS DISPLAYED INDICATING THAT THERE WAS A PROBLEM DETECTED WITH DRAINAGE PUMP, INFUSION/IRRIGATION AND SUCTION FUNCTIONS WERE UNAVAILABLE DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANTATION. THE SURGERY WAS CANCELLED, AND THE EQUIPMENT WAS REPAIRED ON THE SAME DAY. REOPERATION HAS BEEN PERFORMED ON THE FOLLOWING DAY AND WAS COMPLETED WITHOUT AFFECTING THE PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. NONE RECEIVED TILL DATE. THIS REPORT PERTAINS TO THE CASSETTE INVOLVED IN THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24033969 ↗MalfunctionJan 8, 2026Suction Problem
A PHYSICIAN REPORTED THAT THE ASPIRATION WAS POOR DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24000457 ↗MalfunctionDec 31, 2025Product Quality Problem; Product Quality Problem
A NON-HEALTH CARE PROFESSIONAL REPORTED THAT IRRIGATION AND ASPIRATION HANDPIECE SLEEVE HAD DEFECT DURING CATARACT SURGERY WITHOUT INTRAOCULAR LENS IMPLANT. THE SURGERY WAS COMPLETED AFTER REPLACING WITH SPARE SLEEVE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23932394 ↗MalfunctionDec 24, 2025Leak/Splash; Leak/Splash
A PHYSICIAN REPORTED THAT THERE WAS LEAKAGE FROM THE BACKSIDE OF THE CASSETTE DUE TO THE LOOSE RUBBER DURING CALIBRATION. THE PROCEDURE DETAILS WERE UNKNOWN. THERE WAS NO REPORTED INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23887441 ↗MalfunctionDec 13, 2025Suction Problem; Suction Problem
A PHYSICIAN REPORTED THAT ASPIRATION ISSUE OCCURRED DURING ULTRASOUND (US) SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY BY REPLACING THE CASSETTE. THE SURGERY TYPE WAS UNKNOWN. THE PRODUCT WAS CONTACTED WITH PATIENT BUT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23797843 ↗MalfunctionDec 11, 2025Suction Problem; Suction Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT THE CASSETTE DID NOT RESPOND TO ULTRASOUND OR DIATHERMY DURING SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER REPLACING THE CASSETTE WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT. THIS REPORT PERTAINS TO THE CASSETTE INVOLVED IN THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23775400 ↗MalfunctionNov 28, 2025Suction Problem; Suction Problem
A PHYSICIAN REPORTED THAT ASPIRATION FAILURE OCCURRED, AND PROPER ASPIRATION COULD NOT BE ACHIEVED DURING POLISHING AFTER INTRAOCULAR LENS IMPLANT (IOL) INSERTION DURING CATARACT SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23671588 ↗MalfunctionNov 25, 2025Leak/Splash; Leak/Splash
A CUSTOMER REPORTED THAT THE CASSETTE WAS LEAKING BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT CONTACT WITH THE PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646845 ↗MalfunctionNov 7, 2025Inability to Irrigate; Inability to Irrigate
A PHYSICIAN REPORTED THAT IRRIGATION FAILURE OCCURRED AND THE SOLUTION DID NOT FLOW AT ALL DURING CATARACT SURGERY (WITH IOL IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23497964 ↗MalfunctionNov 7, 2025Leak/Splash; Failure to Prime; Leak/Splash; Failure to Prime
A PHYSICIAN REPORTED THAT PRIMING TEST FAILURE OCCURRED AND BALANCED SALT SOLUTION LEAKED FROM THE BOTTOM OF THE CASSETTE DURING CALIBRATION. AN ERROR OCCURRED ON THE CONSOLE BEFORE SETUP, BUT THERE WAS NO ERROR MESSAGE. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23497764 ↗MalfunctionOct 24, 2025Infusion or Flow Problem; Infusion or Flow Problem
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FLUID WAS CONTINUOUSLY FLOWING OUT OF IRRIGATION PORTS OF HANDPIECE DURING CATARACT (WITH INTRAOCULAR LENS IMPLANT) SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED ON SAME DAY. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23378205 ↗MalfunctionOct 23, 2025Suction Problem; Suction Problem
A PHYSICIAN REPORTED THAT ASPIRATION PRESSURE DID NOT INCREASE DURING CATARACT SURGERY (WITH INTRAOCULAR LENSES IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23365731 ↗MalfunctionOct 23, 2025Leak/Splash; Leak/Splash
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE CASSETTE PULLS LIQUID WHEN CONNECTED BEFORE THE SURGERY. THE SURGERY DETAILS AND PATIENT IMPACT DETAILS WERE UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 23364802 ↗MalfunctionSep 15, 2025Obstruction of Flow; Obstruction of Flow
A PHYSICIAN REPORTED THAT THE ULTRASOUND TIP COULD NOT BE OSCILLATED DURING THE SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT. THE PROCEDURE TYPE WAS UNKNOWN. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23048900 ↗MalfunctionSep 12, 2025Obstruction of Flow; Obstruction of Flow
A CUSTOMER REPORTED THAT DURING CATARACT SURGERY (WITHOUT INTRA OCULAR LENS IMPLANTATION), THERE WAS FREQUENT FAILURES WITH AN ALARM AS THERE WAS SOME PROBLEM IN ASPIRATION LINE (ERROR APPEARED ON THE SCREEN STATING THERE WERE PROBLEMS WITH THE SUCTION). THE PROCEDURE WAS COMPLETED ON THE SAME DAY BY CHANGING OF PACKS AS THE PROBLEM OCCURRED REPEATEDLY WITH THE SECOND PACK AS WELL. THERE WAS NO PATIENT CONTACT WITH SUSPECT PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038368 ↗MalfunctionSep 12, 2025Obstruction of Flow; Obstruction of Flow
A CUSTOMER REPORTED THAT DURING CATARACT SURGERY (WITHOUT INTRA OCULAR LENS IMPLANTATION), THERE WAS FREQUENT FAILURES WITH AN ALARM AS THERE WAS SOME PROBLEM IN ASPIRATION LINE (ERROR APPEARED ON THE SCREEN STATING THERE WERE PROBLEMS WITH THE SUCTION). THE PROCEDURE WAS COMPLETED ON THE SAME DAY BY CHANGING OF PACKS AS THE PROBLEM OCCURRED REPEATEDLY WITH THE SECOND PACK AS WELL. THERE WAS NO PATIENT CONTACT WITH SUSPECT PRODUCT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038312 ↗MalfunctionSep 11, 2025Suction Problem; Suction Problem
A PHYSICIAN REPORTED THAT ASPIRATION FAILURE OCCURRED DURING ULTRASOUND (US) AND IRRIGATION/ASPIRATION (IA) MODE OF SURGERY. THE PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE INVOLVED EYE AND THE PATIENT IDENTIFIER ARE NOT AVAILABLE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23024376 ↗MalfunctionSep 8, 2025Suction Failure; Suction Failure
A PHYSICIAN REPORTED ASPIRATION FAILURE OCCURRED WHEN THE HANDPIECE (PHACOEMULSIFICATION TIP) WAS USED DURING ULTRASOUND MODE (US) FOR CATARACT SURGERY (WITH INTRAOCULAR LENS IMPLANT). THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22993960 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page