Alcon, Infiniti, UltraVit 8065751196Recalls & reported problems

Alcon Laboratories, Inc. · Infiniti Vision System UltraVit 23 GA Anterior Vitrectomy Pak · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00380657511969), received between Feb 3, 2022 and May 17, 2024 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut7
  • Adverse Event Without Identified Device or Use Problem1
  • Complete Blockage1
  • Inability to Irrigate1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information1
  • Rupture1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 17, 2024Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE CUTTING FAILURE OF THE PROBE OCCURRED BEFORE CATARACT SURGERY (WITH INTRAOCULAR LENS IMPLANT (IOL)). THE CONDITION OF ASPIRATION AND ACTUATION WAS UNKNOWN. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH NEW PRODUCT. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 19338370 ↗
  • MalfunctionApr 30, 2024Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE FAILURE OF PRODUCT IN THE CUT AND IT WAS ALSO STUCK BEFORE VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19215769 ↗
  • MalfunctionDec 5, 2023Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT CUTTING AND ACTUATION FAILURE OF THE CUTTER DURING CATARACT SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18271973 ↗
  • MalfunctionDec 5, 2023Failure to Cut; Suction Failure; Failure to Cut; Suction Failure

    A NURSE REPORTED THAT THE VITRECTOMY PROBE HAD NO CUTTING PERFORMANCE AND NO ASPIRATION BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265108 ↗
  • MalfunctionDec 5, 2023Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE CUTTER DID NOT CUT PROPERLY DURING SURGERY. THE PROCEDURE WAS COMPLETED AFTER THE CUTTER WAS REPLACED WITH ANOTHER ONE. THERE WAS NO PATIENT INJURY REPORTED. THE PROCEDURE TYPE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18264358 ↗
  • InjuryJul 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NURSE REPORTED THAT, FOREIGN MATERIAL WAS FOUND AT TIP OF THE PROBE.ADDITIONAL INFORMATION REQUESTED AND RECEIVED THAT, POSTERIOR CAPSULAR RUPTURE (PCR) OCCURRED DURING THE CATARACT SURGERY, BUT NO ADVERSE EFFECTS WERE REPORTED TO THE PATIENT.

    Patient problems: Rupture

    MAUDE report 17371874 ↗
  • MalfunctionMay 25, 2023Complete Blockage; Failure to Cut; Complete Blockage; Failure to Cut

    A CUSTOMER REPORTED THAT THE VITRECTOMY PROBE NOT WORKING, IT HAD ONE OF THE HOSES BLOCKED AND DOES NOT CUT DURING VITRECTOMY SURGERY. ASPIRATION STATUS AND PATIENT HARM WERE NOT REPORTED. THE SURGERY TYPE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17002865 ↗
  • MalfunctionAug 12, 2022Failure to Cut; Failure to Cut

    A PHARMACIST REPORTED VITRECTOME KNIFE DID NOT WORK, DID NOT CUT DURING VITRECTOMY PROCEDURE. PROCEDURE WAS COMPLETED BY USING ANOTHER KNIFE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15218424 ↗
  • MalfunctionFeb 3, 2022Inability to Irrigate; Inability to Irrigate

    A NURSE REPORTED WATER DIDN'T COME OUT FROM THE IRRIGATION LINE DURING A CATARACT SURGERY. THE PRODUCT WAS REPLACED AND THE SURGERY WAS COMPLETED. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13441124 ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →