Alcon 8065751447Recalls & reported problems

Alcon Laboratories, Inc. · 23 GA Entry System, Enhanced, 3-CT · Class 2

Quick summary

This Alcon eye-surgery trocar is used to create an entry point during vitrectomy. Reports mostly describe blades being dull or blunt before surgery, sometimes bent or broken, making tissue puncture difficult or requiring another trocar. There are 29 reports, with no deaths or injuries, and no recalls name this device. Reports rose from 3 in the previous 12 months to 19 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00380657514472), received between Mar 18, 2021 and Sep 23, 2026 — classified as 23 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 5 reportsJul 2026: 4 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Dull, Blunt19
  • Fitting Problem2
  • Break1
  • Component Missing1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information2

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Dull, Blunt

    A nurse reported that ophthalmic trocar blade was found to be blunt and not sharp before the vitrectomy surgery. Surgery has been completed on the same day without any patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866227 ↗
  • MalfunctionSep 18, 2026Dull, Blunt

    A nurse reported that before the vitrectomy surgery, ophthalmic blade was blunt and not sharp. Patient details were not reported. It is also unclear whether the surgery was completed. There was no further patient impact was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50818236 ↗
  • MalfunctionSep 7, 2026Dull, Blunt

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50651699 ↗
  • MalfunctionJul 10, 2026Dull, Blunt

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25777709 ↗
  • MalfunctionJul 10, 2026Dull, Blunt

    A PHYSICIAN REPORTED THAT BEFORE VITRECTOMY SURGERY AN OPHTHALMIC TROCAR BLADE WAS FOUND TO BE BENT. THE SURGERY WAS COMPLETED ON THE SAME DAY BY USING ANOTHER ONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25777537 ↗
  • MalfunctionJul 3, 2026Component Missing; Dull, Blunt

    A CUSTOMER REPORTED THAT BEFORE SURGERY AN OPHTHALMIC DEVICE MISSED PARTS AND WAS UNABLE TO PUNCTURE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25704693 ↗
  • MalfunctionJul 3, 2026Dull, Blunt

    A PHYSICIAN REPORTED THAT BEFORE VITRECTOMY SURGERY AN OPHTHALMIC TROCAR BLADE WAS NOT SHARP. THERE WAS NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25704113 ↗
  • MalfunctionJun 19, 2026Dull, Blunt

    A HEALTHCARE PROFESSIONAL REPORTED THAT BEFORE VITRECTOMY SURGERY AN OPHTHALMIC TROCAR BLADE WAS FOUND BENT. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ALTERNATIVE PRODUCT WITHOUT A PATIENT HEALTH IMPACT. THE DETAILS OF THE PROCEDURE HAVE NOT BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25560980 ↗
  • MalfunctionJun 11, 2026Dull, Blunt

    A HEALTHCARE PROFESSIONAL REPORTED THAT BEFORE VITRECTOMY SURGERY, AN OPHTHALMIC ENTRY SYSTEM TROCAR WAS FOUND TO BE BLUNT. THE SURGERY WAS COMPLETED ON THE SAME DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25470133 ↗
  • MalfunctionJun 9, 2026Dull, Blunt

    A NURSE REPORTED THAT BEFORE VITRECTOMY SURGERY, AN OPHTHALMIC TROCAR BLADE WAS BLUNT. THE SURGERY WAS COMPLETED ON THE SAME DAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25447553 ↗
  • MalfunctionJun 9, 2026Dull, Blunt

    A NURSE REPORTED THAT BEFORE VITRECTOMY SURGERY OPHTHALMIC TROCAR BLADE WAS DULL. THE SURGERY WAS COMPLETED ON THE SAME DAY AND THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25447354 ↗
  • MalfunctionJun 9, 2026Dull, Blunt

    A NURSE REPORTED THAT DURING THE VITRECTOMY SURGERY AN OPHTHALMIC TROCAR WAS NOT SHARP. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ALTERNATIVE PRODUCT WITHOUT THE PATIENT HEALTH IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25444205 ↗
  • MalfunctionJan 13, 2026Dull, Blunt

    A NURSE REPORTED THAT BEFORE VITRECTOMY SURGERY IN THE LEFT EYE, AN OPHTHALMIC ENTRY SYSTEM BLADE WAS FOUND TO BE DULL AND NOT SHARP. THE SURGERY WAS COMPLETED THE SAME DAY USING AN ALTERNATIVE ENTRY SYSTEM. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24051098 ↗
  • MalfunctionJul 11, 2025Dull, Blunt; Dull, Blunt

    A PHYSICIAN REPORTED THAT DURING THE VITRECTOMY SURGERY, OPHTHALMIC TROCAR BLADE WAS TOO DULL. THE SURGERY HAS BEEN COMPLETED BY ANOTHER TROCAR. THE PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 22479250 ↗
  • MalfunctionJul 7, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT BEFORE CATARACT AND VITRECTOMY COMBINATION SURGERY, THE OPHTHALMIC TROCAR TIP DAMAGED. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22412595 ↗
  • MalfunctionJan 10, 2025Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT AN OPHTHALMIC KNIFE WAS NOT SHARP BEFORE VITRECTOMY SURGERY. SURGERY WAS COMPLETED AFTER REPLACING A PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21121593 ↗
  • MalfunctionOct 2, 2024Break; Break

    A NURSE REPORTED THAT BEFORE SURGERY THAT A NEWLY OPENED TROCAR WAS BROKEN. PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20350002 ↗
  • MalfunctionApr 23, 2024Dull, Blunt; Dull, Blunt

    A CUSTOMER REPORTED THAT AN OPHTHALMIC ENTRY SYSTEM WAS USED BEFORE THE VITRO RETINAL SURGERY AND BAYONET WAS NOT SHARP. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19161738 ↗
  • MalfunctionDec 11, 2023Fitting Problem; Fitting Problem

    A NURSE REPORTED THAT AN OPHTHALMIC OPERATING NEEDLE UNABLE TO REMOVE BEFORE THE VITRECTOMY SURGERY. THE PATIENT HARM WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18300107 ↗
  • MalfunctionNov 6, 2023Dull, Blunt; Dull, Blunt

    CORRECTED INFORMATION RECEIVED AND STATED THAT THE REPORTED EVENT OF THE PUNCTURE SYSTEM WILL BE LEFT IN THE FLAT PART OF THE SCLERA AS A PASSAGE FOR INTRAOCULAR INSTRUMENTS TO ENTER AND EXIT. THEREFORE, THIS EVENT NO LONGER MEETS REPORTING REQUIREMENTS, BECAUSE A SERIOUS INJURY HAS NOT OCCURRED, AND IT IS NOT LIKELY THAT A RECURRENCE WOULD RESULT IN SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18070929 ↗
  • MalfunctionAug 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A CUSTOMER REPORTED THAT AND OPHTHALMIC ENTRY SYSTEM CANNULA GOT DETACHED DURING THE VITRECTOMY SURGERY AND THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17555946 ↗
  • MalfunctionAug 20, 2021Dull, Blunt; Dull, Blunt

    A NURSE REPORTED THAT BEFORE A VITRECTOMY SURGERY AN OPHTHALMIC TROCAR WAS FOUND TO BE BLUNT. SURGERY WAS COMPLETED BY ALTERNATE TROCAR. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12349050 ↗
  • MalfunctionMar 18, 2021Fitting Problem; Fitting Problem

    A CUSTOMER REPORTED THAT THE TROCAR CAME OUT WHEN EXCHANGING THE VITRECTOR AND LIGHT PIPES DURING VITREORETINAL PROCEDURES; IT WAS REPORTED THAT THE INSIDE OF THE TROCAR WAS GRABBING OR STICKING TO THE INSTRUMENTS BEING PASSED IN AND OUT. THE SURGEON REPOSITIONED THE TROCAR BACK IN PLACE EACH TIME. THE PROCEDURES WERE COMPLETED WITHOUT PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11516896 ↗