Alcon, Intrepid 8065751512Recalls & reported problems
Alcon Laboratories, Inc. · Alcon Intrepid Curved 0.3mm Polymer I/A Tip · Class 2
All models of Alcon, Intrepid (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Alcon Intrepid curved irrigation-aspiration tip is a small hand-held tip used during cataract surgery to wash out and suction away lens fragments. The 35 reports mostly describe the tip breaking, snapping in two, or arriving bent or twisted — sometimes found that way when the sterile package was opened — along with suction that stops working and a missing polymer protector;4 reports involved injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 00380657515127), received between Jan 18, 2022 and Jun 10, 2026 — classified as 4 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Break” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break12
- Suction Failure7
- Material Twisted/Bent6
- Adverse Event Without Identified Device or Use Problem4
- Connection Problem2
- Sharp Edges2
- Complete Blockage1
- Defective Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- Insufficient Information6
- Capsular Bag Tear2
- Rupture2
- Corneal Edema1
- Eye Burn1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2026Material Twisted/Bent
A CUSTOMER REPORTED THAT BEFORE THE CATARACT SURGERY THE IRRIGATION AND ASPIRATION TIPS WERE FOUND TO BE BENT. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ALTERNATIVE PRODUCT WITHOUT ANY PATIENT CONTACT AND HEALTH IMPACT. THIS COMPLAINT IS PERTAINING THE ONE OF TWO REPORTS RECEIVED FROM THE INITIAL REPORTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25455922 ↗MalfunctionMar 21, 2026Suction Failure
A PHYSICIAN REPORTED THAT DURING SURGERY, AN OPHTHALMIC TIP EXHIBITED POOR ASPIRATION. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER THE PRODUCT WAS REPLACED WITH ANOTHER ONE, AND THERE WAS NO PATIENT HARM. THE PROCEDURE TYPE WAS UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24656947 ↗MalfunctionFeb 12, 2026Break
A NURSE REPORTED THAT DURING CATARACT SURGERY IRRIGATION AND ASPIRATION (I/A) TIP POLYMER PROTECTOR ON THE FRONT WAS MISSING. PROCEDURE WAS COMPLETED BY USING ANOTHER I/A TIP. PATIENT HARM WAS NOT REPORTED.
Patient problems: Insufficient Information
MAUDE report 24321023 ↗MalfunctionFeb 12, 2026Break
A CUSTOMER REPORTED THAT THE THREADED TOP OF THE OPHTHALMIC IRRIGATION¿ASPIRATION TIP BROKE INTO TWO PIECES. PROCEDURE DETAILS AND PATIENT IMPACT HAVE NOT BEEN PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Insufficient Information
MAUDE report 24320320 ↗MalfunctionOct 31, 2025Suction Failure; Suction Failure
A CUSTOMER REPORTED THAT AN OPTHALMIC IRRIGATION AND ASPIRATION TIP WOULD NOT EXHIBIT THE SUCTION FUNCTION DURING THE CATARACT SURGERY. DETAILS OF THE PATIENT IMPACT HAVE NOT BEEN REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 23438538 ↗MalfunctionJun 13, 2025Suction Failure; Suction Failure
A PHYSICIAN REPORTED THAT DURING THE OPHTHALMIC SURGERY, OPHTHALMIC IRRIGATION AND ASPIRATION TIP COULD NOT BE USED ON A PATIENT PLANNED ONLY IA OPERATION DUE TO IA ASPIRATION FAILURE. THE SURGERY WAS COMPLETED AFTER REPLACING THE WHOLE SET WITH ANOTHER ONE. THE SAMPLE WAS NOT AVAILABLE BECAUSE IT WAS USED ON A PATIENT WITH AN INFECTION AND DISCARDED. THE INVOLVED EYE AND THE PATIENT IDENTIFIER WERE NOT AVAILABLE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22202350 ↗MalfunctionMar 4, 2025Suction Failure; Suction Failure
A PHYSICIAN REPORTED THAT DURING SURGERY, THE OPHTHALMIC IRRIGATION ASPIRATION TIP WAS DEFORMED AND THE ASPIRATION WAS POOR. THE SURGERY WAS COMPLETED BY USING NEW IRRIGATION AND ASPIRATION TIP. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21512701 ↗MalfunctionFeb 25, 2025Material Twisted/Bent; Material Twisted/Bent
A PHYSICIAN REPORTED THAT BEFORE CATARACT SURGERY AN OPHTHALMIC IRRIGATION AND ASPIRATION TIP WAS FOUND BENT. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ALTERNATIVE TIP AND THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21457603 ↗MalfunctionDec 9, 2024Adverse Event Without Identified Device or Use Problem; Sharp Edges; Adverse Event Without Identified Device or Use Problem; Sharp Edges
A CUSTOMER REPORTED THAT AN OPHTHALMIC TIP WAS USED TO PERFORM A SURGERY. THE PATIENT EXPERIENCED TEAR IN THE CAPSULE. THE SURGERY WAS COMPLETED. PROCEDURE DETAILS WAS NOT REPORTED.
Patient problems: Capsular Bag Tear
MAUDE report 20879392 ↗MalfunctionOct 14, 2024Break; Break
A NURSE REPORTED THAT DURING CATARACT SURGERY, FOUR OF OPHTHALMIC TIPS BROKE OUTSIDE THE EYE AT THE END OF CASE. SURGERY WAS COMPLETED WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20437862 ↗MalfunctionSep 23, 2024Suction Failure; Suction Failure
A PHYSICIAN REPORTED THAT IN AN IRRIGATION AND ASPIRATION TIP EXHIBITED ASPIRATION FAILURE DURING SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. NO PATIENT IMPACT WAS REPORTED. PROCEDURE DETAILS WAS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20277764 ↗MalfunctionSep 12, 2024Break; Break
A PHYSICIAN REPORTED THAT DURING PRE-SURGERY SETUP IRRIGATION AND ASPIRATION TIP WAS NOT ATTACHED WHEN OPENED. IT WAS ALSO REPORTED THAT IT WAS CHECKED INSIDE THE CASE, BUT IT DID NOT SEEM TO HAVE FALLEN OFF. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING WITH ANOTHER PRODUCT. NO PATIENT IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20201640 ↗MalfunctionJul 30, 2024Break; Break
A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY IT WAS FOUND THAT AN OPHTHALMIC TIP FOUND BROKEN AND WAS REMAINED INSIDE. PATIENT IMPACT WAS NOT REPORTED.
Patient problems: Insufficient Information
MAUDE report 19862861 ↗MalfunctionJun 13, 2024Break; Break
A NURSE REPORTED THAT DURING CATARACT SURGERY MANY OPHTHALMIC TIPS BROKE OFF. SURGERY WAS COMPLETED WITH USING AN ALTERATIVE PRODUCT AND THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19526333 ↗MalfunctionJun 4, 2024Material Twisted/Bent; Material Twisted/Bent
A PHYSICIAN REPORTED THAT BEFORE CATARACT SURGERY, AN OPHTHALMIC IRRIGATION AND ASPIRATION TIP WAS FOUND TO BE TWISTED. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE TIP WITH ANOTHER ONE. THERE WAS NO CONTACT WITH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19458267 ↗MalfunctionMay 15, 2024Break; Break
A CUSTOMER REPORTED THAT AN OPHTHALMIC OPERATING IRRIGATION AND ASPIRATION TIP WAS BROKEN. DETAILS OF PROCEDURE AND PATIENT HARM WAS NOT REPORTED. ADDITIONAL DETAILS HAS BEEN REQUESTED AND NONE RECEIVED TILL DATE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19318177 ↗MalfunctionApr 24, 2024Complete Blockage; Suction Failure; Complete Blockage; Suction Failure
A CUSTOMER REPORTED THAT AN OPHTHALMIC OPERATING IRRIGATION AND ASPIRATION TIP HOLE WAS NOT OPENED AND NO ASPIRATION DURING THE CATARACT SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH ALTERNATIVE PRODUCT. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19172475 ↗MalfunctionMar 25, 2024Suction Problem; Connection Problem; Suction Problem; Connection Problem
A PHYSICIAN REPORTED THAT THE ASPIRATION WAS NOT APPLIED AT ALL DURING USE OF IRRIGATION ASPIRATION TIP FROM THE BEGINNING OF THE SURGERY, THAT THE THREADED PART OF THE IRRIGATION ASPIRATION HANDPIECE CONNECTION OF THE POLYMER IRRIGATION ASPIRATION TIP WAS DAMAGED. THE SURGERY WAS COMPLETED AFTER REPLACING THE TIP WITH ANOTHER ONE. THERE'S NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18968327 ↗MalfunctionMar 20, 2024Connection Problem; Connection Problem
A PHYSICIAN REPORTED THAT OPHTHALMIC IRRIGATION AND ASPIRATION TIP CONNECTION FAILURE OCCURRED DURING TEST. THE CATARACT SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE.
Patient problems: Insufficient Information
MAUDE report 18939309 ↗InjuryMar 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY AN OPHTHALMIC IRRIGATION ASPIRATION TIPS COMPROMISED THE CAPSULE DURING POLISH MODE OF PHACOEMULSIFICATION AND THE SURGERY WAS COMPLETED. THIS COMPLAINT IS PERTAINING THE ONE OF TWO REPORTS RECEIVED FROM THE INITIAL REPORTER.
Patient problems: Rupture
MAUDE report 18893393 ↗InjuryMar 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY AN OPHTHALMIC IRRIGATION ASPIRATION TIPS COMPROMISED THE CAPSULE DURING POLISH MODE OF PHACOEMULSIFICATION. SURGERY WAS COMPLETED BY IMPLANTING A THREE PIECE LENS INSTEAD OF THE INITIALLY PLANNED LENS. THIS COMPLAINT IS PERTAINING THE TWO OF TWO REPORTS RECEIVED FROM THE INITIAL REPORTER.
Patient problems: Rupture
MAUDE report 18863980 ↗MalfunctionFeb 15, 2024Material Twisted/Bent; Material Twisted/Bent
A PHYSICIAN REPORTED THAT BEFORE CATARACT SURGERY AN OPHTHALMIC TIP WAS ABNORMALLY BENT. THE SURGERY WAS COMPLETED AFTER REPLACING THE TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18709912 ↗InjuryNov 14, 2023Sharp Edges; Sharp Edges
A CUSTOMER REPORTED THAT A SEVENTY FIVE YEARS OLD FEMALE PATIENT, WHILE USING AN OPHTHALMIC TIP DURING A PHACO/INTRA OCULAR LENS (IOL) SURGERY WHEN POLISHING THE POSTERIOR CAPSULE AFTER CORTEX REMOVAL A CAPSULE TEAR OCCURRED AS THE TIP HAD SHARP EDGES. THE SURGERY WAS COMPLETED BY PERFORMING AN ANTERIOR VITRECTOMY AND A SULCUS LENS WAS PLACED. THERE WAS NO PATIENT HARM.
Patient problems: Capsular Bag Tear
MAUDE report 18129057 ↗InjuryOct 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION RECEIVED INDICATING THAT PATIENT DID NOT EXPERIENCED CORNEAL BURN. NUMBER OF PATIENTS INVOLVED IS 15.
Patient problems: Corneal Edema; Eye Burn
MAUDE report 18011858 ↗MalfunctionOct 4, 2023Material Twisted/Bent; Material Twisted/Bent
A PHYSICIAN REPORTED THAT DURING A SURGERY WHILE USING AN OPHTHALMIC TIP THE SHAFT WAS FOUND BENT WHEN IT WAS OPENED AND CONNECTED TO THE IRRIGATION AND ASPIRATION HANDPIECE. HE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. PATIENT HARM WAS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17867691 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page