CONSTELLATION ULTRAVIT ENGAUGE 8065751571Recalls & reported problems

Alcon Laboratories, Inc. · CONSTELLATION 23 GA ULTRAVIT VITRECTOMY PROBE, 7500 CPM · Class 2

All models of CONSTELLATION ULTRAVIT ENGAUGE (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Alcon Constellation UltraVit vitrectomy probe is a surgical instrument used to cut and suction tissue inside the eye during vitrectomy surgery. The 28 reports mostly describe the probe failing to cut or losing suction, which surgeons say slowed or disrupted the operation and in some cases required swapping the instrument mid-surgery. There are no deaths and no injuries reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00380657515714), received between Mar 29, 2022 and Sep 17, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Cut23
  • Suction Failure5
  • Suction Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • Insufficient Information5

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Failure to Cut

    A physician reported that the probe has poor cutting during vitrectomy surgery. The surgery was completed on the same day by replacing the product with new one. There was no information about patient impact.

    Patient problems: Insufficient Information

    MAUDE report 50804579 ↗
  • MalfunctionJul 21, 2026Failure to Cut

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25888227 ↗
  • MalfunctionJul 3, 2026Failure to Cut

    A PHYSICIAN REPORTED THAT THE VITRECTOMY PROBE HAD NO CUTTING WHICH AFFECTED THE SURGICAL EFFICIENCY DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE INSTRUMENT. THERE WAS PATIENT IMPACT. ADDITIONAL INFORMATION WAS REQUESTED BUT NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 25703465 ↗
  • MalfunctionFeb 3, 2026Failure to Cut

    A NURSE REPORTED THAT THE PROBE HAD POOR CUTTING DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24243660 ↗
  • MalfunctionJan 12, 2026Failure to Cut

    A NURSE REPORTED THAT THE VITRECTOMY PROBE HAD POOR CUTTING DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24039869 ↗
  • MalfunctionJan 6, 2026Suction Problem

    A PHYSICIAN REPORTED THAT PROBE COULD NOT ASPIRATE DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY BY CHANGING WITH NEW PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23971801 ↗
  • MalfunctionNov 26, 2025Suction Problem; Suction Problem

    A NURSE REPORTED THAT THE PROBE HAD NO VACUUM DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23656250 ↗
  • MalfunctionNov 17, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT THE VITRECTOMY PROBE HAD POOR CUTTING BEFORE THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY BY REPLACING PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23577187 ↗
  • MalfunctionOct 29, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT VITRECTOMY PROBE HAD POOR CUTTING BEFORE THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23412188 ↗
  • MalfunctionSep 2, 2025Suction Problem; Suction Problem

    A NURSE REPORTED THAT PROBE WAS OCCLUDED AND NO VACUUM PRESSURE BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931734 ↗
  • MalfunctionMay 7, 2025Suction Failure; Suction Failure

    A NURSE REPORTED THAT THE VITRECTOMY PROBE HAD NO ASPIRATION DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21974910 ↗
  • MalfunctionJan 1, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT THE VITRECTOMY PROBES HAD NO CUTTING BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCTS WITH ANOTHER ONE. THERE WAS NO IMPACT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21051251 ↗
  • MalfunctionJan 1, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT THE VITRECTOMY PROBES HAD NO CUTTING BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCTS WITH ANOTHER ONE. THERE WAS NO IMPACT ON PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21051250 ↗
  • MalfunctionAug 28, 2024Failure to Cut; Suction Failure; Failure to Cut; Suction Failure

    A NURSE REPORTED THAT THE PROBE HAD NO ASPIRATION AND CUTTING DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20083913 ↗
  • MalfunctionAug 28, 2024Suction Failure; Suction Failure

    A NURSE REPORTED THAT THE VITRECTOMY PROBE HAD NO ASPIRATION DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20083720 ↗
  • MalfunctionAug 22, 2024Suction Failure; Suction Failure

    THE PHYSICIAN REPORTED THAT THERE WAS NO ASPIRATION DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY BY REPLACING THE PROBE. THERE WAS NO PATIENT HARM. THIS REPORT PERTAINS TO PROBE INVOLVED IN THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20024618 ↗
  • MalfunctionJun 21, 2024Failure to Cut; Suction Failure; Failure to Cut; Suction Failure

    A CUSTOMER REPORTED THAT THE VITRECTOMY PROBE HAD NO CUTTING AND NO ASPIRATION DURING CALIBRATION. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH NEW ONE. THE PROCEDURE DETAILS WERE UNKNOWN. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19583436 ↗
  • MalfunctionApr 15, 2024Failure to Cut; Failure to Cut

    THE NURSE REPORTED THAT PROBE COULD NOT CUT BEFORE THE SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER PROBE. THERE WAS NO PATIENT HARM. THE PROCEDURE TYPE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19103573 ↗
  • MalfunctionApr 10, 2024Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE POOR CUTTING DURING VITRECTOMY AND CATARACT COMBINATION SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 19079882 ↗
  • MalfunctionApr 9, 2024Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE POOR CUTTING DURING CATARACT/VITRECTOMY COMBINATION SURGERY. THE PROCEDURE WAS COMPLETED AFTER REPLACEMENT OF NEW ONE. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19071046 ↗
  • MalfunctionJan 3, 2024Failure to Cut; Failure to Cut

    A NURSE REPORTED PROBE COULD NOT CUT DURING VITRECTOMY SURGERY. A NEW PRODUCT WAS CHANGED TO COMPLETE THE SURGERY. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: Insufficient Information

    MAUDE report 18433715 ↗
  • MalfunctionNov 3, 2023Failure to Cut; Failure to Cut

    A NURSE REPORTED PROBE COULD NOT CUT DURING VITRECTOMY SURGERY. A NEW PRODUCT WAS CHANGED TO COMPLETE THE SURGERY. THERE WAS NO REPORT OF PATIENT HARM.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18063616 ↗
  • MalfunctionNov 2, 2023Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT THE PROBE COULD NOT CUT DURING VITRECTOMY SURGERY. THE ASPIRATION CONDITION WAS UNKNOWN. THE PROCEDURE WAS COMPLETED ON THE SAME DAY. THERE WAS NO HARM TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18062450 ↗
  • MalfunctionOct 19, 2023Failure to Cut; Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT CUTTER DID NOT CUTTING EFFECTIVELY BEFORE THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17962842 ↗
  • MalfunctionJul 5, 2023Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT PROBE COULD NOT CUT DURING THE CATARACT AND VITRECTOMY COMBINATION SURGERY. THE SURGERY WAS COMPLETED BY REPLACING WITH A NEW PRODUCT. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17258173 ↗

Other Vitrectomy, instrument cutter devices