Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618Recalls & reported problems
Alcon Laboratories, Inc. · Constellation Vision System 7500 CPM UltraVit Probe Wide Angle Endoilluminator 25+ TotalPlus Vitrectomy Pak EdgePlus Blade Valved Entry System · Class 2
All models of Constellation, Ultravit, 25+, TotalPlus, EdgePlus (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00380657516186), received between Aug 11, 2021 and Sep 26, 2024 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 15, 2024Class IIOpen, Classified
Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
Z-2674-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Cut4
- Fluid/Blood Leak4
- Device Displays Incorrect Message2
- Infusion or Flow Problem2
- No Flow2
- Failure to Calibrate1
- Suction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 26, 2024Suction Problem; Suction Problem
A HEALTHCARE PROFESSIONAL REPORTED THAT THE ASPIRATION WAS POOR DURING VITRECTOMY-RETINAL (VIT-RET) SURGERY. THEY REMOVED THE ENTIRE PAK FROM THE STERILE FIELD, REPLACED WITH NEW AND COMPLETED THE PROCEDURE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20308923 ↗MalfunctionMar 1, 2024Failure to Cut; Failure to Cut
A NURSE REPORTED THAT THERE WAS A MALFUNCTION WITH CUTTER OPERATION; CUTTING DID NOT OCCUR AND TISSUE WAS NOT CUT PROPERLY DURING VITRECTOMY SURGERY. THE SURGERY WAS CONCLUDED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE CONDITION OF ASPIRATION WAS UNKNOWN. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18818089 ↗MalfunctionJan 18, 2024Infusion or Flow Problem; Infusion or Flow Problem
A NURSE REPORTED THAT THE WATER WAS NOT COMING OUT OF THE INFUSION PUMP BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 18533387 ↗MalfunctionJan 8, 2024No Flow; No Flow
A NURSE REPORTED THAT NO AIR FLOW IN FLUID AND AIR EXCHANGE MODE DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY REPLACING WITH ANOTHER ONE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18461806 ↗MalfunctionSep 7, 2022Fluid/Blood Leak; Device Displays Incorrect Message; Fluid/Blood Leak; Device Displays Incorrect Message
A CUSTOMER REPORTED THE CASSETTE WAS LEAKING. THE TIMING AND THE TYPE OF THE SURGERY IS UNKNOWN. THERE IS NO REPORT OF A PATIENT HARM. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REPRESENTATIVE WHO STATED DURING SETUP THE TECHNICIAN NOTED MESSAGE ON THE CONSOLE STATING NO BALANCE SALT SOLUTION (BSS) AND ALSO DISCOVERED THE BSS LEAKED ON THE FLOOR HOWEVER THERE WAS NO HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15370300 ↗MalfunctionSep 7, 2022Fluid/Blood Leak; Device Displays Incorrect Message; Fluid/Blood Leak; Device Displays Incorrect Message
A CUSTOMER REPORTED THE CASSETTE WAS LEAKING. THE TIMING AND THE TYPE OF THE SURGERY IS UNKNOWN. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REPRESENTATIVE THAT TECHNICIAN NOTED THE MESSAGE ON THE CONSOLE STATING NO BALANCED SALT SOLUTION (BSS) DURING CALIBRATION. A NEW FLUIDIC MANAGEMENT SYSTEM (FMS) WAS USED TO COMPLETE THE SET UP FOR PROCEDURE. THERE WAS NO PATIENT HARM. SAMPLE WAS AVAILABLE FOR RETURN. NO FURTHER DETAILS TO PROVIDE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15370001 ↗MalfunctionAug 25, 2022Fluid/Blood Leak; Failure to Calibrate; Fluid/Blood Leak; Failure to Calibrate
A CUSTOMER REPORTED THE CASSETTE WAS LEAKING AND NEVER PRIMED DURING CALIBRATION. THE PRODUCT WAS REPLACED AND THE PROCEDURE WAS COMPLETED. THE PRODUCT WAS NOT AVAILABLE FOR RETURN. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15294658 ↗MalfunctionJul 7, 2022Failure to Cut; Failure to Cut
A HEALTH PROFESSIONAL REPORTED THAT MULTIPLE VITRECTOMY CUTTERS WERE CUTTING INTERMITTENTLY AND SOMETIMES THERE WAS NO CUTTING. IT WAS NOTED THAT THE PROBLEM WAS WITH THE OPHTHALMIC CONSOLE PNEUMATIC ASSEMBLY PANEL. THE PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14946892 ↗MalfunctionJul 7, 2022Failure to Cut; Failure to Cut
A HEALTH PROFESSIONAL REPORTED THAT MULTIPLE VITRECTOMY CUTTERS WERE CUTTING INTERMITTENTLY AND SOMETIMES THERE WAS NO CUTTING. IT WAS NOTED THAT THE PROBLEM WAS WITH THE OPHTHALMIC CONSOLE PNEUMATIC ASSEMBLY PANEL. THE PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14946845 ↗MalfunctionJul 7, 2022Failure to Cut; Failure to Cut
A HEALTH PROFESSIONAL REPORTED THAT MULTIPLE VITRECTOMY CUTTERS WERE CUTTING INTERMITTENTLY AND SOMETIMES THERE WAS NO CUTTING. IT WAS NOTED THAT THE PROBLEM WAS WITH THE OPHTHALMIC CONSOLE PNEUMATIC ASSEMBLY PANEL. THE PROCEDURE DETAILS AND PATIENT IMPACT WERE NOT PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14946826 ↗MalfunctionMay 24, 2022No Flow; No Flow
A NURSE REPORTED BALANCED SALT SOLUTION (BSS) WAS NOT FLEW IN THE INFUSION LINE DURING VITRECTOMY PROCEDURE. THE PRODUCT WAS REPLACED, AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14486636 ↗MalfunctionMay 16, 2022Fluid/Blood Leak; Fluid/Blood Leak
A CUSTOMER REPORTED A CASSETTE WAS LEAKING. THE TIMING OF THE EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN RECEIVED FROM THE COMPANY REPRESENTATIVE WHO STATED THAT THE CASSETTE WAS LEAKING FROM THE BACK DURING PRIMING. THE PRODUCT WAS REPLACED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INVOLVEMENT. THE SAMPLE WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14407485 ↗MalfunctionAug 11, 2021Infusion or Flow Problem; Infusion or Flow Problem
A NURSE REPORTED THE SURGEON TRIED TO PUT AIR INTO THE EYE USING THE FLUID/AIR EXCHANGE FUNCTION, BUT NO AIR CAME OUT FROM THE INFUSION CANNULA DURING A VITRECTOMY PROCEDURE. THE PRODUCT WAS REPLACED WITH ANOTHER ONE AND THE PROCEDURE WAS COMPLETED. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12306050 ↗
FDA clearance and classification
The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618
Has the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2674-2024 (Class II), began on Jul 15, 2024: Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Aug 11, 2021 and Sep 26, 2024. In those reports, the most common specific device problems are “Failure to Cut” (4), “Fluid/Blood Leak” (4) and “Device Displays Incorrect Message” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618?
None of the reports are classified as deaths, 0 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751618, and is it still sold?
The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Constellation, Ultravit, 25+, TotalPlus, EdgePlus, model 8065751618). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation723 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation631 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation550 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls399 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls303 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation367 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page