Skip to content

Alcon, Intrepid 8065752146Recalls & reported problems

Alcon Laboratories, Inc. · Alcon Intrepid 0.3mm Polymer I/A Coaxial Conical Curved · Class 2

All models of Alcon, Intrepid (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00380657521463), received between Sep 27, 2023 and Jun 19, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Complete Blockage” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Blockage3
  • Suction Failure3
  • Break2
  • Material Twisted/Bent2
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information4

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Material Twisted/Bent

    A PHYSICIAN REPORTED THAT BEFORE A SURGERY AN OPHTHALMIC TIP WAS FOUND TO BE BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25561905 ↗
  • MalfunctionFeb 20, 2026Suction Failure

    A PHYSICIAN REPORTED THAT ASPIRATION FAILED FROM THE IRRIGATION AND ASPIRATION HANDPIECE DURING SURGERY OF ULTRASOUND PHASE. PROCEDURE WAS COMPLETED ON THE SAME DAY. PATIENT IMPACT AND PROCEDURE DETAILS HAVE NOT BEEN REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 24399226 ↗
  • MalfunctionFeb 18, 2026Suction Failure

    A PHYSICIAN REPORTED THAT ASPIRATION FAILED FROM THE IRRIGATION AND ASPIRATION HANDPIECE DURING SURGERY OF ULTRASOUND PHASE. PROCEDURE WAS COMPLETED ON THE SAME DAY. PATIENT IMPACT AND PROCEDURE DETAILS HAVE NOT BEEN REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 24376751 ↗
  • MalfunctionSep 16, 2025Complete Blockage; Suction Problem; Complete Blockage; Suction Problem

    THE PHYSICIAN REPORTED THAT DURING THE CATARACT SURGERY THE IRRIGATION AND ASPIRATION HANDPIECE WOULD NOT INCREASING THE SUCTION PRESSURE AND IT WAS CLOGGED. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE WITHOUT ANY PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23062686 ↗
  • MalfunctionMay 28, 2025Break; Break

    A PHYSICIAN REPORTED THAT BEFORE CATARACT SURGERY, THE TIP OF THE PRODUCT WAS BROKEN. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095082 ↗
  • MalfunctionJan 9, 2025Complete Blockage; Complete Blockage

    A CUSTOMER REPORTED THAT, THE OPHTHALMIC IRRIGATION AND ASPIRATION HANDPIECE HAS CLOGGED APPROXIMATELY THREE TIMES AND THE SURGEON FLUSHED THE DEVICE THEN IT WORKED PROPERLY. THE SURGERY WAS SLIGHTLY DELAYED, BUT THERE WAS NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21109249 ↗
  • MalfunctionSep 27, 2024Complete Blockage; Complete Blockage

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING SURGERY, AN OPHTHALMIC HANDPIECE TIP WAS CLOGGED. PROCEDURE DETAILS AND PATIENT IMPACT HAVE NOT BEEN PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED; NONE HAS BEEN RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 20320409 ↗
  • MalfunctionSep 3, 2024Material Twisted/Bent; Material Twisted/Bent

    A HEALTH CARE PROFESSIONAL REPORTED THAT OPHTHALMIC IRRIGATION AND ASPIRATION HANDPIECE TIP WAS FOUND TWISTED IN PACKAGING. PATIENT AND PROCEDURE DETAILS WERE NOT REPORTED. NO PATIENT IMPACT WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20126504 ↗
  • MalfunctionOct 24, 2023Break; Break

    A PHYSICIAN REPORTED THAT THE IRRIGATION AND ASPIRATION TIP WAS BROKE IMMEDIATELY DURING SETTING JUST BEFORE USING IT DURING A TEST. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE DETAILS OF PATIENT IMPACT WAS NOT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993928 ↗
  • MalfunctionSep 27, 2023Suction Failure; Suction Failure

    A PHYSICIAN REPORTED THAT THE ASPIRATION FAILED WITH OPHTHALMIC HANDPIECE DURING THE SURGERY. THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE INVOLVED EYE AND THE PATIENT IDENTIFIER WERE NOT AVAILABLE. THERE WAS NO PATIENT HARM. THE PROCEDURE TYPE WAS AN UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17823482 ↗

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page