Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752436Recalls & reported problems

Alcon Laboratories, Inc. · 23 GA TOTAL PLUS Vitrectory Pak, Beveled 10,000 CPM ULTRAVIT Probe Wide Angle Endoilluminator · Class 2

All models of Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Alcon Constellation vitrectomy cutter has 79 adverse event reports, with no deaths and 9 injuries. Reports mostly describe cutting failure, suction problems, leaks, or blocked infusion lines during eye surgery. One open recall began April 2025 because some units may have been damaged during manufacturing, risking sterility. Reports in the most recent 12 months (21) were lower than the prior 12 months (32).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
79
Deaths reported
0
Injuries reported
9
Recalls naming it
1

Summary

The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 00380657524365), received between Jul 8, 2021 and Sep 30, 2026 — classified as 9 injuries, 70 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (55 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 6 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 5 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 4 reportsSep 2025: 4 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 8 reportsJan 2026: 3 reportsFeb 2026: 7 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 28, 2025Class IIOpen, Classified

    Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.

    Z-2108-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Cut55
  • Suction Problem9
  • Suction Failure5
  • Ambient Noise Problem3
  • Disconnection2
  • Fluid/Blood Leak2
  • Leak/Splash2
  • No Flow2

Patient problems

  • No Clinical Signs, Symptoms or Conditions62
  • Insufficient Information12
  • Retinal Tear3
  • Retinal Detachment2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Endophthalmitis1
  • Ischemia1

Recent adverse event reports

The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Failure to Cut

    A healthcare professional reported that cutter did not cut during the vitrectomy surgery. The aspiration condition was normal. The other surgery details and patient impact details were unknown.Additional information requested and received that the surgery was completed by replacing the cutter. Theare was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943687 ↗
  • MalfunctionMar 6, 2026Failure to Cut

    A NURSE REPORTED THAT CUTTING FAILURE OF THE PROBE OCCURRED DURING THE VITRECTOMY SURGERY. FIVE THOUSAND CUTS PER MINUTE WAS NOT COMMENTED IF TRIED AND VACUUM OR ASPIRATION APPEARED NORMAL THROUGHOUT THE SURGERY. MAXIMUM CUT RATE AND MAXIMUM VACUUM WERE USED AND NO ERROR MESSAGE WAS DISPLAYED ON THE MACHINE. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24530579 ↗
  • MalfunctionFeb 25, 2026Suction Problem; Failure to Cut

    A HEALTHCARE PROFESSIONAL REPORTED THAT THE CUT RATE WAS NORMAL IN THE BEGINNING AFTER A WHILE CUTTING WAS REDUCED DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY USING NEW PAK. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE ASPIRATION WAS QUITE NOT NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24440196 ↗
  • MalfunctionFeb 19, 2026Failure to Cut

    A NURSE REPORTED THAT THE CUTTER/PROBE WAS NOT FUNCTIONING EFFICIENTLY (CUTTING ISSUE) EVEN REDUCED THE 5000 CUTS PER MINUTE (CPM) DURING THE VITRECTOMY SURGERY. THE OTHER SURGERY DETAILS WERE UNKNOWN. THE SURGERY WAS COMPLETED ON SAME DAY BY REPLACING THE PAK. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24386480 ↗
  • MalfunctionFeb 16, 2026Failure to Cut

    THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24345752 ↗
  • MalfunctionFeb 11, 2026Obstruction of Flow

    A CUSTOMER REPORTED VIA HEALTH AUTHORITY THAT THE INFUSION HEAD WAS NOT UNOBSTRUCTED (BLOCKED) BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE INFUSION CANNULA WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24312895 ↗
  • MalfunctionFeb 10, 2026Failure to Cut

    A NON-HEALTH CARE PROFESSIONAL REPORTED THAT CUTTER STOPPED WORKING PROPERLY AND EVEN AFTER REDUCING THE CUT RATE THERE WAS NO IMPROVEMENT DURING SURGERY. THE PROCEDURE TYPE AND COMPLETION DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24302465 ↗
  • MalfunctionFeb 4, 2026Obstruction of Flow

    A PHARMACIST REPORTED THAT THE INFUSION LINE WAS NOT WORKING, APPEARED TO BE BLOCKED RIGHT FROM THE START (BEFORE) VITREO RETINAL SURGERY. A NEW CASSETTE WAS OPENED TO RETRIEVE THE INFUSION LINE AND TO COMPLETE THE SURGERY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24256259 ↗
  • MalfunctionFeb 2, 2026Failure to Cut

    A NURSE REPORTED THAT PROBE COULD NOT CUT BEFORE VITRECTOMY SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24232452 ↗
  • MalfunctionJan 13, 2026Failure to Cut

    A CUSTOMER REPORTED THAT THE PRODUCT (PROBE) COULD NOT CUT BEFORE INTRA OCULAR LENS IMPLANTATION. IT WAS CHANGE TO NEW ONE AND PROCEDURE COMPLETED ON THE SAME DAY. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24053240 ↗
  • MalfunctionJan 8, 2026Failure to Cut

    A CUSTOMER REPORTED CUTTER WOULDN¿T CUT AT AN UNKNOWN TIMING. THE PROCEDURE TYPE AND SURGERY DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24001245 ↗
  • InjuryJan 6, 2026Failure to Cut

    A CUSTOMER REPORTED THAT DURING A PARS PLANA VITRECTOMY PROCEDURE ON THE LEFT EYE, A VITRECTOMY PROBE STOPPED CUTTING AND CONTINUED SUCTION, CAUSING A RETINAL DETACHMENT. THE SURGERY WAS COMPLETED ON THE SAME DAY WITH AN ALTERNATIVE CUTTER. THE PATIENT¿S CURRENT CONDITION WAS UNKNOWN. THIS REPORT PERTAINED TO THE VITRECTOMY PROBE INVOLVED IN THE REPORTED EVENT.

    Patient problems: Retinal Detachment

    MAUDE report 23973595 ↗
  • MalfunctionDec 30, 2025Failure to Cut; Failure to Cut

    A CUSTOMER REPORTED THAT THE CUTTER STOPPED CUTTING DURING PARS PLANAR VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS A CONTACT WITH THE PATIENT BUT NO IMPACT TO THE PATIENT. THIS REPORT PERTAINS TO VITRECTOMY CUTTER INVOLVED IN THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23922624 ↗
  • MalfunctionDec 26, 2025Failure to Cut; Ambient Noise Problem; Failure to Cut; Ambient Noise Problem

    A CUSTOMER REPORTED THAT AFTER REDUCING THE CUT RATE, THE CUTTER WAS FUNCTIONING AND THERE WAS ABNORMAL NOISE FROM THE CUTTER DURING VITRECTOMY SURGERY FOR THE LEFT EYE. THE SURGERY WAS COMPLETED BY REDUCING THE CUT RATE. THERE WAS PATIENT CONTACT, BUT NO IMPACT OCCURRED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23901293 ↗
  • MalfunctionDec 25, 2025Failure to Cut; Failure to Cut

    NEW INFORMATION HAS BEEN RECEIVED INDICATING THAT THE PATIENT LEFT EYE WAS INVOLVED IN THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23892352 ↗
  • MalfunctionDec 24, 2025Failure to Cut; Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT CUTTER DID NOT CUT DURING VITRECTOMY SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS CONTACT WITH THE PATIENT BUT NO IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23887770 ↗
  • MalfunctionDec 18, 2025Failure to Cut; Failure to Cut

    A HEALTHCARE PROFESSIONAL REPORTED THAT THE VITRECTOMY PROBES THAT STOPPED CUTTING DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER PROBE. THERE WAS A PATIENT CONTACT BUT NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23836227 ↗
  • InjuryDec 10, 2025Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION INDICATED THAT THE PREVIOUSLY ABORTED PROCEDURE WAS EVENTUALLY COMPLETED, HOWEVER, THE EXACT DATE AND TIME OF COMPLETION REMAIN UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 23765813 ↗
  • InjuryDec 10, 2025Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION INDICATED THAT THE PREVIOUSLY ABORTED PROCEDURE WAS EVENTUALLY COMPLETED, HOWEVER, THE EXACT DATE AND TIME OF COMPLETION REMAIN UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 23765810 ↗
  • InjuryDec 10, 2025Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION INDICATED THAT THE PREVIOUSLY ABORTED PROCEDURE WAS EVENTUALLY COMPLETED, HOWEVER, THE EXACT DATE AND TIME OF COMPLETION REMAIN UNKNOWN. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NONE RECEIVED TILL DATE.

    Patient problems: Insufficient Information

    MAUDE report 23765806 ↗
  • MalfunctionNov 5, 2025Failure to Cut; Failure to Cut

    A HEALTH CARE PROFESSIONAL REPORTED THAT THE CUTTER FAILED TO CUT IN AN UNKNOWN TIMING. THE PROCEDURE TYPE WAS UNKNOWN. THE OTHER SURGERY DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 23477680 ↗
  • MalfunctionOct 27, 2025Failure to Cut; Failure to Cut

    ADDITIONAL INFORMATION WAS RECEIVED THAT THERE WAS PATIENT CONTACT, BUT NO IMPACT TO THE LEFT EYE OF THE FEMALE ADULT PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23390504 ↗
  • MalfunctionOct 16, 2025Failure to Cut; Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE VITRECTOMY CUTTER WAS DELIVERING POOR CUTTING SPEED IN BETWEEN THE VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WAS PATIENT CONTACT WITH NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23319898 ↗
  • MalfunctionOct 16, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT ALTHOUGH THE PRIMING OF THE PAK WAS NORMAL, AFTER A FEW MINUTES INTO THE PARS PLANA VITRECTOMY (PPV) SURGERY IN THE RIGHT EYE OF A FEMALE ADULT PATIENT, VITRECTOMY CUTTER/PROBE WAS NOT WORKING EFFICIENTLY. RE-PRIMING WAS SUCCESSFUL BUT THE VITRECTOR WAS CUTTING AND VACUUMING BUT NOT PERFORMING AS EXPECTED. THE MACHINE WAS REPLACED WITH A DIFFERENT MACHINE AND A NEW PACK WAS OPENED AND USED BUT THE PERFORMANCE ISSUE PERSISTED. THE VITRECTOR CUTTING AND VACUUM FUNCTIONS REMAINED SUBOPTIMAL THROUGHOUT THE EARLY PART OF THE PROCEDURE. THE SURGEON CONFIRMED THAT THE CUTTER BEGAN TO FUNCTION UP TO SPEED ONLY DURING THE LATTER HALF OF THE PROCEDURE, PRIOR TO CLOSURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO ISSUES. THERE WAS PATIENT CONTACT BUT NO IMPACT TO THE PATIENT. ADDITIONALLY, IT WAS CONFIRMED THAT THERE WAS ONLY POOR CUTTING AND ASPIRATION WAS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23308418 ↗
  • MalfunctionOct 15, 2025Suction Problem; Failure to Cut; Suction Problem; Failure to Cut

    ADDITIONAL INFORMATION REQUESTED AND RECEIVED THAT THE EVENT OCCURRED BEFORE VITRECTOMY SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 23297947 ↗

Other Vitrectomy, instrument cutter devices

All Vitrectomy, instrument cutter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page