CENTURION, Interpid 8065752900Recalls & reported problems
Alcon Laboratories, Inc. · SMALL PARTS KIT W/ INTREPID* ULTRA SLEEVES 0.9 MM Tipless · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00380657529001), received between Apr 9, 2021 and Nov 27, 2025 — classified as 7 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Particulates” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Particulates2
- Product Quality Problem2
- Insufficient Flow or Under Infusion1
- Material Puncture/Hole1
- Material Split, Cut or Torn1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- Toxic Anterior Segment Syndrome (TASS)5
- Uveitis5
- Insufficient Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 27, 2025Product Quality Problem; Product Quality Problem
A PHYSICIAN REPORTED THAT THE SLEEVE BECAME WIDE AND MADE FOLD WHILE ENTERING IN FROM THE INCISION DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANTATION. THE PROCEDURE WAS COMPLETED BY REPLACING THE PACK WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23664070 ↗MalfunctionAug 1, 2025Product Quality Problem; Product Quality Problem
A PHYSICIAN REPORTED SLEEVES WERE DEFECTIVE AND INCLUDED BOTH A SLEEVE WITH NO HOLES AND A SLEEVE WITH MULTIPLE HOLES DURING CATARACT SURGERY (WITH INTRAOCULAR LENS IMPLANT). THE SURGERY WAS PERFORMED AND COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO REPORT OF PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 22684606 ↗MalfunctionMar 25, 2025Material Puncture/Hole; Material Puncture/Hole
A PHYSICIAN REPORTED THAT THERE WAS A HOLE IN SLEEVE AND LEAKAGE OCCURRED DURING CATARACT/VITRECTOMY COMBINATION SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687504 ↗InjuryJan 3, 2025Particulates; Particulates
A PHYSICIAN REPORTED THAT A THREAD-LIKE FIBER WAS FOUND INSIDE THE PATIENT'S EYE AFTER CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANTATION. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED THAT THE FOREIGN MATERIAL WAS REMOVED FROM THE PATIENT'S EYE BY PERFORMING ADDITIONAL SURGERY. THE CURRENT CONDITION OF THE PATIENT WAS UNKNOWN. ADDITIONAL INFORMATION WAS RECEIVED AND CONFIRMED THAT THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21074823 ↗MalfunctionApr 20, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
A PHYSICIAN REPORTED THE SLEEVE WAS FOUND TORN DURING A CATARACT PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT HARM. THE SAMPLE WAS NOT AVAILABLE FOR RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14159220 ↗InjuryApr 5, 2022Particulates; Insufficient Flow or Under Infusion; Particulates; Insufficient Flow or Under Infusion
A CUSTOMER REPORTED THE SURGEON NOTED WHITE FILAMENT WITHIN THE SHAFT OF THE PHACOEMULSIFICATION SLEEVE DURING QUADRANT REMOVAL IN CATARACT PROCEDURE. THE IRRIGATION FLOW WAS LOWER THAN NORMAL. THE PROCEDURE WAS COMPLETED AFTER THE PRODUCT WAS REPLACED. THERE WAS NO STRUCTURAL DAMAGE OCCURRED TO THE EYE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014369 ↗InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AFTER A CATARACT SURGERY PATIENT EXPERIENCED INFLAMMATION (FIBRIN) AND ENDOPHTHALMITIS. PATIENT WAS GIVEN WITH AN ANTIBACTERIAL AGENT AND THE CONDITION WAS IMPROVING. IT WAS INFORMED THAT THE PATIENT WAS IN CONTACT WITH SURGICAL PROCEDURE PACKS, VISCOELASTICS, OPHTHALMIC PHACOEMULSIFICATION TIP, OPHTHALMIC PHACOEMULSIFICATION HANDPIECE AND IRRIGATION/ASPIRATION TIP. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE SURGEON CONFIRMED THE EVENT WAS TASS (TOXIC ANTERIOR SEGMENT SYNDROME).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS); Uveitis
MAUDE report 13867826 ↗InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AFTER A CATARACT SURGERY PATIENT EXPERIENCED INFLAMMATION (FIBRIN) AND ENDOPHTHALMITIS. PATIENT WAS GIVEN WITH AN ANTIBACTERIAL AGENT AND THE CONDITION WAS IMPROVING. IT WAS INFORMED THAT THE PATIENT WAS IN CONTACT WITH OPHTHALMIC CATRIDGE, SURGICAL PROCEDURE PACKS, VISCOELASTICS, OPHTHALMIC PHACOEMULSIFICATION TIP, OPHTHALMIC PHACOEMULSIFICATION HANDPIECE AND IRRIGATION/ASPIRATION TIP. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE SURGEON CONFIRMED THE EVENT WAS TASS (TOXIC ANTERIOR SEGMENT SYNDROME).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS); Uveitis
MAUDE report 13867641 ↗InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AFTER A CATARACT SURGERY PATIENT EXPERIENCED INFLAMMATION (FIBRIN) AND ENDOPHTHALMITIS. PATIENT WAS GIVEN WITH AN ANTIBACTERIAL AGENT AND THE CONDITION WAS IMPROVING. IT WAS INFORMED THAT THE PATIENT WAS IN CONTACT WITH OPHTHALMIC CATRIDGE, SURGICAL PROCEDURE PACKS, VISCOELASTICS, OPHTHALMIC PHACOEMULSIFICATION TIP, OPHTHALMIC PHACOEMULSIFICATION HANDPIECE AND IRRIGATION/ASPIRATION TIP. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE SURGEON CONFIRMED THE EVENT WAS TASS (TOXIC ANTERIOR SEGMENT SYNDROME).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS); Uveitis
MAUDE report 13867290 ↗InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AFTER A CATARACT SURGERY PATIENT EXPERIENCED INFLAMMATION (FIBRIN) AND ENDOPHTHALMITIS. PATIENT WAS GIVEN WITH AN ANTIBACTERIAL AGENT AND THE CONDITION WAS IMPROVING. IT WAS INFORMED THAT THE PATIENT WAS IN CONTACT WITH SURGICAL PROCEDURE PACKS, VISCOELASTICS, OPHTHALMIC PHACOEMULSIFICATION TIP, OPHTHALMIC PHACOEMULSIFICATION HANDPIECE AND IRRIGATION/ASPIRATION TIP. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE SURGEON CONFIRMED THE EVENT WAS TASS (TOXIC ANTERIOR SEGMENT SYNDROME). THIS REPORT REPRESENTS SURGICAL PROCEDURE PAK INVOLVED IN THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS); Uveitis
MAUDE report 13867287 ↗InjuryMar 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PHYSICIAN REPORTED THAT AFTER A CATARACT SURGERY PATIENT EXPERIENCED INFLAMMATION (FIBRIN) AND ENDOPHTHALMITIS. PATIENT WAS GIVEN WITH AN ANTIBACTERIAL AGENT AND THE CONDITION WAS IMPROVING. IT WAS INFORMED THAT THE PATIENT WAS IN CONTACT WITH SURGICAL PROCEDURE PACKS, VISCOELASTICS, OPHTHALMIC PHACOEMULSIFICATION TIP, OPHTHALMIC PHACOEMULSIFICATION HANDPIECE AND IRRIGATION/ASPIRATION TIP. ADDITIONAL INFORMATION WAS RECEIVED CLARIFYING THE SURGEON CONFIRMED THE EVENT WAS TASS (TOXIC ANTERIOR SEGMENT SYNDROME). THIS REPORT REPRESENTS SURGICAL PROCEDURE PAK INVOLVED IN THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Toxic Anterior Segment Syndrome (TASS); Uveitis
MAUDE report 13867177 ↗MalfunctionApr 9, 2021Use of Device Problem; Use of Device Problem
A CUSTOMER REPORTED THAT THE INFUSION SLEEVE WAS EASILY CUT AND WAS UNUSABLE DURING A PROCEDURE. THE PRODUCT WAS REPLACED AND PROCEDURE COMPLETED WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11643278 ↗
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page