Skip to content

Alcon, TurboSonics, ABS, MicroTip 8065790020Recalls & reported problems

Alcon Laboratories, Inc. · 30° Round, 0.9 mm TurboSonics® ABS® MicroTip™ Tip · Class 2

All models of Alcon, TurboSonics, ABS, MicroTip (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380657900206), received between Nov 14, 2024 and Nov 14, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Suction Failure” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2023Class IIOpen, Classified

    Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench.

    Z-2410-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2024Suction Failure; Suction Failure

    A PHYSICIAN REPORTED THAT NO ASPIRATION WAS APPLIED AT ALL DURING ULTRASOUND (US) AND IRRIGATION ASPIRATION (IA) PHASE OF CATARACT [WITH INTRAOCULAR LENS (IOL) IMPLANT] SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY. THE PAK AND TIP WAS REPLACED, BUT DID NOT IMPROVE THE PROBLEM, SO ASPIRATION WAS PERFORMED MANUALLY BY OTHER COMPANY NEEDLE. THERE WAS NO PATIENT IMPACT. THIS REPORT PERTAINS TO PHACOEMULSIFICATION TIP INVOLVED IN THIS REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20681999 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alcon, TurboSonics, ABS, MicroTip 8065790020

Has the Alcon, TurboSonics, ABS, MicroTip 8065790020 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2410-2023 (Class II), began on Jun 28, 2023: Reports of patient injury during use in procedures due to potential for plastic debris from over-tightening with included plastic wrench. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Alcon, TurboSonics, ABS, MicroTip 8065790020?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 14, 2024. That report lists device problem “Suction Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alcon, TurboSonics, ABS, MicroTip 8065790020?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alcon, TurboSonics, ABS, MicroTip 8065790020, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Alcon, TurboSonics, ABS, MicroTip, model 8065790020). Its record shows it as in commercial distribution.