ARCHITECT 2P24-40Recalls & reported problems
Abbott Laboratories · ARCHITECT c4000 PROCESSING MODULE INTEGRATED · Class 1
All models of ARCHITECT (51)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Abbott ARCHITECT processing module is used in clinical chemistry testing. Its 646 reports mostly describe falsely high or low lab results, such as sodium, magnesium, potassium and calcium readings. The reports include 1 death and 8 injuries, and they fell from 49 in the previous 12 months to 25 in the last 12 months. One open recall started in October 2023; it covers parts containing natural rubber latex that were not labeled as such.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 645
- Deaths reported
- 1
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 645 adverse event reports that list this exact device (DI 00380740003753), received between Sep 16, 2016 and Sep 14, 2026 — classified as 1 death, 8 injuries, 636 malfunctions.
The most frequently reported specific device problem is “High Test Results” (305 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 19, 2023Class IIOpen, Classified
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
Z-0682-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results305
- Low Test Results214
- Incorrect, Inadequate or Imprecise Result or Readings19
- Fluid/Blood Leak18
- High Readings15
- Leak/Splash15
- Smoking11
- Low Readings9
Patient problems
- No Clinical Signs, Symptoms or Conditions343
- No Known Impact Or Consequence To Patient241
- No Consequences Or Impact To Patient39
- No Patient Involvement10
- Injury3
- No Code Available3
- Needle Stick/Puncture2
- Abrasion1
Recent adverse event reports
The 25 most recent of 645 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Thermal Decomposition of Device
The customer observed error 3598 when processing on the ARCHITECT c4000 Processing Module. Further troubleshooting was performed and found cpu11 to cnn5 cable have burnt mark. There was no impact to patient management or user safety.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50753451 ↗MalfunctionMay 15, 2026Leak/Splash
THE FIELD SERVICE REPRESENTATIVE (FSR) WAS TROUBLESHOOTING A LEAK INSIDE THE ARCHITECT C4000 PROCESSING MODULE. DURING THIS ACTIVITY, LIQUID SPLASHED FROM THE EXTERNAL WASTE PUMP WHILE THE FSR WAS RESEATING THE CONNECTOR BETWEEN THE TUBING AND THE EXTERNAL WASTE PUMP. THE LIQUID SPLASH CONTACTED THE FSR¿S FACE, HANDS, EYEGLASSES, AND CLOTHING. AT THE TIME OF THE INCIDENT, THE FSR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), INCLUDING EYEGLASSES, A LAB COAT, AND GLOVES. THERE WAS NO MEDICAL TREATMENT WAS REQUIRED OR PROVIDED BY A PHYSICIAN. THERE WAS NO IMPACT TO THE OPERATOR, AND NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25183943 ↗MalfunctionApr 1, 2026High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS FROM ALINITY C PROCESSING MODULE ON MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE: ON (B)(6) 2026 SID (B)(6): 59 YRS OLD FEMALE: INITIAL=7.5 MG/DL /REPEATED=1.5MG/DL. ON (B)(6) 2026 SID (B)(6): 80 YRS OLD MALE: INITIAL=6.9 MG/DL /REPEATED=1.8 MG/DL. CUSTOMER REFERENCE RANGE FOR MAGNESIUM=1.6 TO 2.6 MG/DL; CRITICAL RANGE=< 1.0 OR > 5.0 MG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24755285 ↗MalfunctionMar 25, 2026Imprecise or Erratic Results
THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C4000 INSTRUMENT. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2026, FOR SID (B)(6), THE INITIAL MAGNESIUM RESULT WAS 6.58 MG/DL (PATIENT PANEL), THE SAMPLE WAS REPEATED WITH OTHER SAMPLES RUNNING AND THE RESULT WAS 1.31 MG/DL. THE SAMPLE WAS REPEATED WITH NO OTHER SAMPLES RUNNING AND THE RESULT WAS 1.91 MG/DL (REPORTED). ON (B)(6) 2026, FOR SID (B)(6), THE INITIAL MAGNESIUM RESULT WAS 0.94 MG/DL (PATIENT PANEL), THE SAMPLE WAS REPEATED WITH OTHER SAMPLES RUNNING AND THE RESULT WAS 4.25 MG/DL. THE SAMPLE WAS REPEATED WITH NO OTHER SAMPLES RUNNING AND THE RESULT WAS 3.97 MG/DL (REPORTED). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24691201 ↗MalfunctionMar 19, 2026Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED ARCHITECT CALCIUM RESULTS GENERATED ON THE ARCHITECT C4000 INSTRUMENT. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT 2 MG/DL, REPEATED NORMAL (THE VALUE WAS NOT PROVIDED). EXPECTED VALUES: 9.0 - 10.2 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24640033 ↗MalfunctionMar 19, 2026High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C4000 INSTRUMENT FOR A 25-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (NORMAL RANGE 1.6 - 2.6 MG/DL): SID (B)(6) 2026 RESULTED 8.0 REPEATED 1.8, 1.9 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24637410 ↗MalfunctionMar 9, 2026High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED POTASSIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR TWO PATIENTS. THE SAMPLES WERE REPEATED WITH LOWER RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6), INITIAL POTASSIUM RESULT= 7.923 MMOL/L; REPEAT RESULT= 4.767 MMOL/L. SID (B)(6), INITIAL POTASSIUM RESULT= 7.977 MMOL/L; REPEAT RESULT= 4.561 MMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547531 ↗MalfunctionFeb 20, 2026High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CALCIUM2 RESULTS GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR TWO PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGES AT DIFFERENT FACILITIES - ARCHITECT 2.25¿2.55 MMOL/L, ALINITY 2.1¿2.55 MMOL/L): (B)(6) 2026 SAMPLE ID (B)(6), (86-YEAR-OLD FEMALE) INITIAL RESULT = 4.47 MMOL/L, (B)(6) 2026 SAME SAMPLE WAS REPEATED ON ALINITY ANALYZER AT ANOTHER LABORATORY AND RESULT = 2.35 MMOL/L. (B)(6) 2026 SAMPLE ID (B)(6), (77-YEAR-OLD FEMALE) INITIAL RESULT = 5.18 MMOL/L, REPEAT RESULT = 5.26 MMOL/L. (B)(6) 2026 NEW SAMPLE OBTAINED AND RESULT ON THE RADIOMETER = 1.16 MMOL/L. (B)(6) 2026 SAME SAMPLE WAS REPEATED ON ALINITY ANALYZER AT ANOTHER LABORATORY AND RESULT = 2.36 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24402603 ↗MalfunctionFeb 9, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED MAGNESIUM RESULT GENERATED ON THE ARCHITECT C4000 ANALYZER FOR 2 PATIENTS. RESULTS WERE NOT REPORTED TO MEDICAL PROVIDER. THE FOLLOWING DATA WAS PROVIDED FOR ONE PATIENT. INITIAL RESULT = 0.4 MMOL/L, REPEAT RESULTS = 0.8 MMOL/L. THE CUSTOMER USES REFERENCE RANGE: 0.74-1.07 MMOL/L . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24293118 ↗MalfunctionFeb 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED DISCREPANT PATIENT RESULTS FOR MULTIPLE ASSAYS FOR 2 PATIENTS WHEN PROCESSED ON THE ARCHITECT ANALYZER. RESULTS WERE NOT REPORTED TO MEDIAL PROVIDER. THE FOLLOWING DATA PROVIDED. SID (B)(6). CALCIUM, INITIAL RESULT = 4.6 MG/DL, REPEAT RESULT = 9.1 MG/DL, REFERENCE RANGE= 8.5 ¿ 10.5 MG/DL. CO2, INITIAL RESULT = 5 MMOL/L, REPEAT RESULT = 23 MMOL/L, REFERENCE RANGE= 20 ¿ 31 MMOL/L GLUCOSE, INITIAL RESULT = 20 MG/DL, REPEAT RESULT = 101 MG/DL, REFERENCE RANGE= 70 ¿ 100 MG/DL. SODIUM, INITIAL RESULT = 110 MMOL/L, REPEAT RESULT = 140 MMOL/L, REFERENCE RANGE= 135 ¿ 145 MMOL/L. SID (B)(6) CALCIUM, INITIAL RESULT = 5.8 MG/DL, REPEAT RESULT = 9.2 MG/DL, REFERENCE RANGE= 8.5 ¿ 10.5 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24245738 ↗MalfunctionJan 23, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED MAGNESIUM RESULT GENERATED ON THE ARCHITECT C4000 ANALYZER FOR 2 PATIENTS. RESULTS WERE NOT REPORTED TO MEDICAL PROVIDER. THE FOLLOWING DATA WAS PROVIDED FOR ONE PATIENT. INITIAL RESULT = 0.4 MMOL/L, REPEAT RESULTS = 0.8 MMOL/L THE CUSTOMER USES REFERENCE RANGE: 0.74-1.07 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24160614 ↗MalfunctionJan 22, 2026Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED POTASSIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR A 62-YEAR-OLD FEMALE TAKING A POTENT NEUROLEPTIC MEDICATION. THE PATIENT SAMPLE WAS REPEATED ON THE BLOOD GAS ANALYZER WITH A HIGHER RESULT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 3.4 - 5.1 MMOL/L): SID (B)(6), INITIAL POTASSIUM RESULT= 1.5 MMOL/L, WHICH WAS CONFIRMED BY A SECOND BLOOD DRAW; THE SAMPLE WAS TEST AT ANOTHER FACILITY WITH THE SAME RESULTS; REPEAT RESULTS ON THE BLOOD GAS ANALYSIS (CAPILLARY BLOOD DRAW)= 4 MMOL/L. THE PATIENT WAS RETESTED ON (B)(6) 2026 WITH A RESULT OF 2.5 MMOL/L. THE PATIENT RECEIVED IV POTASSIUM. THE CUSTOMER CONFIRMED THE IV POTASSIUM THE PATIENT RECEIVED WASN'T CONSIDERED UNNECESSARY. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24153891 ↗MalfunctionJan 5, 2026High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR A 79-YEAR-OLD MALE. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 1.6 - 2.6 MG/DL): SID (B)(6), INITIAL MAGNESIUM RESULT= 6.6 MG/DL; REPEAT RESULT= 2.1 MG/DL . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23960246 ↗MalfunctionDec 29, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED POTASSIUM RESULT GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR MULTIPLE SAMPLES. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED NORMAL RANGE: 3.5 TO 5.1 MMOL/L): (B)(6) 2025, SID (B)(6): INITIAL RESULT = 6.65 MMOL/L, REPEAT = 4.83 MMOL/L, (B)(6) 2025, SID (B)(6): INITIAL RESULT = 6.88 MMOL/L, REPEAT = 5.03 MMOL/L, (B)(6) 2025, SID (B)(6): INITIAL RESULT = 6.68 MMOL/L, REPEAT = 4.94 MMOL/L , NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23907296 ↗MalfunctionDec 23, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED POTASSIUM RESULTS GENERATED BY THE ARCHITECT C4000 PROCESSING MODULE FOR MULTIPLE PATIENTS. THE SAMPLES WERE REPEATED, AND NORMAL LOWER RESULTS WERE OBTAINED. THE CUSTOMER PROVIDED ONE EXAMPLE: CUSTOMER¿S NORMAL RANGE: 3.5 TO 5.1 MMOL/L. INITIAL POTASSIUM RESULT = 7.8 MMOL/L, REPEAT RESULT = 3.0 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23882128 ↗MalfunctionDec 22, 2025Patient Data Problem; Patient Data Problem
THE CUSTOMER OBSERVED THE ARCHITECT C4000 PRINTOUT SHOWED THE INCORRECT PATIENT DEMOGRAPHICS, WHEN COMPARED WITH INFORMATION ON BARCODE LABEL FOR A SAMPLE. THE BARCODE LABEL CONTAINS THE CORRECT PATIENT DEMOGRAPHICS. PATIENT 1: SID (B)(6) RUN (B)(6) 2025 WITH PATIENT ID (B)(6) INITIALS (B)(6) FOR IRON TEST. THIS PATIENT WAS HOSPITALIZED A MONTH AGO AND HAS NOT BEEN BACK FOR ANYTHING SINCE. PATIENT 2: CORRECT PATIENT ID (B)(6) WITH PATIENT INITIALS LL WHICH ALSO HAD AN IRON TEST CUSTOMER MANUALLY ORDERED TEST FOR CORRECT PATIENT ID (B)(6) AND REPORTED THE RESULTS WITH CORRECT DEMOGRAPHICS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23873054 ↗MalfunctionDec 19, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 1.6 - 2.6 MG/DL): (B)(6) 2025, SID (B)(6), INITIAL MAGNESIUM RESULT= 2.40 MG/DL; (B)(6) 2025, REPEAT RESULT= 6.93 MG/DL. THE PATIENT IS A 34-YEAR-OLD FEMALE THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23849099 ↗MalfunctionDec 17, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C4000 INSTRUMENT FOR ONE 53-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (MMOL/L): (B)(6) INITIAL RESULT 2.86 MMOL/L, REPEATED 0.87 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23826872 ↗MalfunctionDec 9, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED MAGNESIUM RESULT GENERATED ON AN ARCHITECT C4000 ANALYZER FOR ONE PATIENT. SID (B)(6) INITIAL RESULT = 5.98 MG/DL, REPEAT RESULTS = 1.98 MG/DL AND 1.90 MG/DL (NORMAL RANGE 1.6-2.5 MG/DL). THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23757271 ↗MalfunctionNov 26, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM (MG) RESULT GENERATED ON THE ARCHITECT C4000 FOR A 72-YEAR-OLD MALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE 1.6-2.6 MG/DL): (B)(6) 2025 SAMPLE ID (B)(6), INITIAL RESULT = 7.23 MG/DL, REPEAT RESULT = 1.91 MG/DL, REPEAT RESULT ON ANOTHER INSTRUMENT (B)(6) = 1.82 MG/DL . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23660463 ↗MalfunctionNov 24, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DECREASED RESULTS ON MULTIPLE ASSAYS GENERATED BY THE ARCHITECT C4000 PROCESSING MODULE FOR MULTIPLE PATIENTS. THE SAMPLES WERE REPEATED, AND NORMAL RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S NORMAL RANGE: SODIUM: 136 TO 145 MMOL/L. POTASSIUM: 3.5 TO 5.1 MMOL/L. CALCIUM: 8.4 TO 10.2 MG/DL. TOTAL BILIRUBIN: 0.2 TO 1.2 MG/DL. (B)(6) 2025 SID (B)(6): INITIAL SODIUM RESULT = 106 MMOL/L, REPEAT RESULT = 139 MMOL/L. (B)(6) 2025 SID (B)(6): INITIAL POTASSIUM RESULT = 2.2 MMOL/L, REPEAT RESULT = 4.1 MMOL/L. INITIAL SODIUM RESULT = 107 MMOL/L, REPEAT RESULT = 138 MMOL/L. (B)(6) 2025 SID (B)(6): INITIAL SODIUM RESULT = 112 MMOL/L, REPEAT RESULT = 141 MMOL/L. INITIAL TOTAL BILIRUBIN RESULT = < 0.1 MG/DL, REPEAT RESULT = 0.4 MG/DL. (B)(6) 2025 SID (B)(6): INITIAL CALCIUM RESULT = 5.7 MG/DL, REPEAT RESULT = 9.2 MG/DL. (B)(6) 2025 SID (B)(6): INITIAL POTASSIUM RESULT = 2.2 MMOL/L, REPEAT RESULT = 3.9 MMOL/L. INITIAL SODIUM RESULT = 120 MMOL/L, REPEAT RESULT = 138 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23633924 ↗MalfunctionNov 20, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED SODIUM AND CHLORIDE RESULTS GENERATED ON A PATIENT WHEN PROCESSED ON THE ARCHITECT C4000 ANALYZER. RESULTS WERE REPORTED TO THE MEDICAL PROVIDER. THE FOLLOWING DATA WAS PROVIDED. ON (B)(6) 2025, INITIAL SODIUM RESULTS=160 MMOL/L, REPEAT RESULTS FROM ANOTHER ANALYZER = 142 MMOL/L. ON (B)(6) 2025, INITIAL CHLORIDE RESULTS= 134 MMOL/L, REPEAT RESULTS FROM THIS ANALYZER AND ANOTHER ANALYZER WERE 109 MMOL/L. THE CUSTOMER USES REFERENCE RANGE FOR SODIUM (135-145 MMOL/L) & CHLORIDE (98-110 MMOL/L). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609020 ↗MalfunctionNov 10, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: CUSTOMER¿S REFERENCE RANGE 1.6-2.6 MG/DL. (B)(6) 2025, SID (B)(6), 72-YEAR-OLD FEMALE, INITIAL MAGNESIUM RESULT= 4.95 MG/DL; REPEAT RESULT= 1.09 MG/DL THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23512005 ↗MalfunctionNov 5, 2025Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED SODIUM ON THE ARCHITECT C4000 PROCESSING MODULE FOR A 39-YEAR-OLD MALE. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 137-147 MMOL/L): SID (B)(6), INITIAL SODIUM RESULT= 103 MMOL/L; REPEAT RESULT (AUTOMATIC)= 111 MMOL/L; REPEAT RESULT ON SAME SAMPLE, SAMPLE INSTRUMENT= 138 MMOL/L; REPEAT RESULT ON ANOTHER INSTRUMENT (C402434)= 137 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478605 ↗MalfunctionOct 30, 2025Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED TOTAL BILIRUBIN RESULTS ON ARCHITECT C4000 PROCESSING MODULE FOR SEVERAL PATIENTS. THE RESULTS PROVIDED WERE: SID (B)(6), INITIAL=<0.2 MG/DL /REPEATED=0.3 MG/DL, SID (B)(6), INITIAL=<0.2 MG/DL /REPEATED=1.5 MG/DL, SID (B)(6), INITIAL=<0.2 MG/DL /REPEATED=0.3 MG/DL. LABORATORY REFERENCE RANGE FOR TOTAL BILIRUBIN= 0.2 TO 1.2 MG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23427839 ↗
FDA clearance and classification
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ARCHITECT 2P24-40
Has the ARCHITECT 2P24-40 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0682-2024 (Class II), began on Oct 19, 2023: The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ARCHITECT 2P24-40?
The FDA's MAUDE database lists 645 adverse event reports for this device, received between Sep 16, 2016 and Sep 14, 2026. In those reports, the most common specific device problems are “High Test Results” (305), “Low Test Results” (214) and “Incorrect, Inadequate or Imprecise Result or Readings” (19); the most common patient problems are “Injury” (3), “Needle Stick/Puncture” (2) and “Abrasion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARCHITECT 2P24-40?
1 report is classified as a death, 8 as injuries and 636 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARCHITECT 2P24-40 FDA approved?
Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Laboratories.
Who makes the ARCHITECT 2P24-40, and is it still sold?
The FDA's device database lists Abbott Laboratories as the company responsible for it (brand ARCHITECT, model 2P24-40). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls677 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE680 reports
- VITROS 6802413Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE12 recalls68 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page