The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740008987), received between Dec 21, 2018 and Dec 20, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • High Test Results1
  • Leak/Splash1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 20, 2019Leak/Splash; Leak/Splash

    THE CUSTOMER REPORTED THAT A TECH WAS PLACING THE SEPTUM ON THE TOTAL PSA CONJUGATE BOTTLE WHEN HIS FINGER WENT THROUGH THE SEPTUM CAUSING THE SEPTUM TO SINK INTO THE VIAL AND FLUID TO SPLASH ON HIS EYELID. THE TECH THOROUGHLY RINSED HIS EYE, BUT DECLINED ANY FURTHER MEDICAL ATTENTION. THE TECH WAS WEARING A LAB COAT AND GLOVES, BUT NO EYE PROTECTION. THERE WAS NO ADVERSE IMPACT TO THE EMPLOYEE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9505678 ↗
  • MalfunctionDec 21, 2018High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED TOTAL PSA RESULT ON THE ARCHITECT I1000SR AANLYZER. THE FOLLOWING DATA WAS PROVIDED: SID (B)(4) INITIAL 32.64 NG/ML REPEATED 0.12, 0.14 AND 0.12. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8188954 ↗

Other PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS devices

All PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page