Elecsys free PSA 09744908190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Elecsys free PSA (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336217672), received between Jul 25, 2025 and Aug 26, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement1
  • Non Reproducible Results1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2025Output Problem; Output Problem

    THE INITIAL REPORTER COMPLAINED OF HIGH RESULTS FOR MULTIPLE PATIENT SAMPLES TESTED FOR ELECSYS FREE PSA (FREE PSA) ON A COBAS E 801 ANALYTICAL UNIT. THE FREE PSA RESULTS WERE HIGHER THAN THE ELECSYS TOTAL PSA (TOTAL PSA) RESULTS. THIS MEDWATCH WILL COVER FREE PSA. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE TOTAL PSA RESULTS. EXAMPLES OF QUESTIONABLE RESULTS WERE PROVIDED FOR 3 PATIENT SAMPLES. PATIENT 1 FREE PSA RESULT WAS 0.0176 NG/ML. THE TOTAL PSA RESULT WAS 0.0073 NG/ML. PATIENT 2 FREE PSA RESULT WAS 0.0166 NG/ML. THE TOTAL PSA RESULT WAS 0.006 NG/ML WITH A DATA FLAG. ON (B)(6) 2025 PATIENT 3 FREE PSA RESULT WAS 0.0203 NG/ML. THE TOTAL PSA RESULT WAS 0.0110 NG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22893304 ↗
  • MalfunctionJul 25, 2025Incorrect Measurement; Non Reproducible Results; Incorrect Measurement; Non Reproducible Results

    WE RECEIVED AN ALLEGATION ABOUT QUESTIONABLE RESULTS FOR 2 SAMPLES FROM THE SAME PATIENT TESTED WITH ELECSYS TOTAL PROSTATE SPECIFIC ANTIGEN (TOTAL PSA) ASSAY AND ELECSYS FREE PROSTATE SPECIFIC ANTIGEN (FREE PSA) ASSAY ON A COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH WILL APPLY TO THE FREE PSA ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE TOTAL PSA ASSAY. SAMPLE 1: TOTAL PSA: INITIAL RESULT: 0.0821 NG/ML. FREE PSA: INITIAL RESULT: 1.99 NG/ML. THIS FREE PSA RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY. THE FREE PSA RESULT WAS GREATER THAN THE TOTAL PSA RESULT, PROMPTING THE REPEAT OF THE SAMPLE THE NEXT DAY. THE REPEAT RESULTS WERE IDENTICAL TO THE INITIAL RESULTS. ON (B)(6) 2025, A NEW BLOOD SAMPLE (SAMPLE 2) WAS TAKEN FROM THE PATIENT AND TESTED: TOTAL PSA RESULT: 0.0807 NG/ML. FREE PSA RESULT: 1.99 NG/ML. OTHER TOTAL PSA AND FREE PSA TESTS WERE PERFORMED AT A DIFFERENT SERVICE LABORATORY USING SIEMENS ATELLICA SOLUTION, RESULTING IN THE FOLLOWING: TOTAL PSA RESULT: 1.220 UG/L. FREE PSA RESULT: 0.023 UG/L. THESE RESULTS WERE DEEMED TO BE CORRECT. IT IS NOT KNOWN IF SAMPLE 2 WAS REPEATED ON SIEMENS TESTING OR IF ANOTHER NEW S

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612925 ↗

Other PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS devices