Skip to content

VITROS 1933027Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products PSA Calibrators · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10758750000319), received between Oct 5, 2023 and Oct 5, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 5, 2023Low Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED VITROS PROSTATIC SPECIFIC ANTIGEN (PSA) RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD LOT 85330 QUALITY CONTROL FLUID AND A SINGLE LEVEL OF NON-VITROS FUJI REBIO TUMOR MARKER CONTROLS LOT 49820 USING VITROS PSA REAGENT LOT 4340 TESTED ON A VITROS XT 7600 INTEGRATED SYSTEM. THE BIORAD RESULTS WERE LOWER THAN EXPECTED WHEN COMPARED TO THE CUSTOMER¿S BASELINE MEAN AND THE FUJI REBIO RESULT WAS LOWER THAN EXPECTED WHEN COMPARED TO THE PACKAGE INSERT MEAN. BIORAD LEVEL 2 VITROS PSA RESULTS OF 1.43 AND 1.45 NG/ML VS. THE EXPECTED PSA RESULT OF 1.98 NG/ML FUJI REBIO LEVEL 1 VITROS PSA RESULT OF 2.31 NG/ML VS. THE EXPECTED PSA RESULT OF 3.21 NG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER THAN EXPECTED RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL SAMPLES. THE CUSTOMER MADE NO INDICATION THAT PATIENT SAMPLES WERE AFFECTED HOWEVER, THE INVESTIGATION CANNOT CONFIRM PATIENT RESULTS WOULD NOT BE AFFECTED IN THE EVENT WERE TO RECUR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17882022 ↗

Other PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS devices

All PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page