Skip to content

ARCHITECT 6L45-41Recalls & reported problems

Abbott Laboratories · Total Bilirubin · Class 2

All models of ARCHITECT (51)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00380740011383), received between Aug 14, 2017 and Nov 2, 2020 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Low Test Results9
  • High Test Results7
  • Incorrect Or Inadequate Test Results2

Patient problems

  • No Known Impact Or Consequence To Patient16
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 2, 2020High Test Results; High Test Results

    THE CUSTOMER OBTAINED A FALSELY ELEVATED ARCHITECT TOTAL BILIRUBIN RESULT FOR ONE PATIENT. SAMPLE ID (B)(6) GENERATED 8.49 AND REPEAT 0.21 MG/DL. A SERUM SAMPLE FROM THE SAME PATIENT WAS ALSO TESTED, AND GENERATED 0.18 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10771438 ↗
  • MalfunctionJul 22, 2020Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED TOTAL BILIRUBIN ON ARCHITECT C8000 PROCESSING MODULE FOR MULTIPLE PATIENTS. ONE EXAMPLE PATIENT RESULT PROVIDED: SID (B)(4) TOTAL BILIRUBIN RESULT=0.20 MG/DL/ REPEATED=0.20 MG/DL/ REPEATED=0.50 MG/DL (NORMAL REFERENCE RANGE=0.2 TO 1.2 MG/DL). THE PERCENT BIAS BETWEEN THE TWO RESULTS ARE 60%. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10309688 ↗
  • MalfunctionMay 6, 2020High Test Results; High Test Results

    THE CUSTOMER OBSERVED ERRATIC TOTAL BILIRUBIN RESULTS ON ONE PATIENT ON THE ARCHITECT C8000 ANALYZER, SERIAL NUMBER (B)(4). THE FOLLOWING DATA WAS PROVIDED: TOTAL BILIRUBIN = 5.4MMOL/L, DIRECT BILIRUBIN = 6.6. THE SAMPLE WAS TESTED ON ANOTHER ARCHITECT C8000, SERIAL NUMBER (B)(4) - TOTAL BILIRUBIN = 6.5MMOL/L, DIRECT BILIRUBIN = 7.0. THE SAMPLE WAS THEN RETESTED ON SN (B)(4) - TOTAL BILIRUBIN = 1.9MMOL/L, DIRECT BILIRUBIN = 5.9. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10028824 ↗
  • MalfunctionApr 8, 2020High Test Results; High Test Results

    THE CUSTOMER OBSERVED ERRATIC TOTAL BILIRUBIN RESULTS ON ONE PATIENT ON THE ARCHITECT C8000 ANALYZER, SERIAL NUMBER (B)(4). THE FOLLOWING DATA WAS PROVIDED: TOTAL BILIRUBIN = 5.4MMOL/L, DIRECT BILIRUBIN = 6.6. THE SAMPLE WAS TESTED ON ANOTHER ARCHITECT C8000, SERIAL NUMBER (B)(4) - TOTAL BILIRUBIN = 6.5MMOL/L, DIRECT BILIRUBIN = 7.0. THE SAMPLE WAS THEN RETESTED ON SN (B)(4) - TOTAL BILIRUBIN = 1.9MMOL/L, DIRECT BILIRUBIN = 5.9. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9939207 ↗
  • MalfunctionFeb 25, 2020Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSELY DEPRESSED TOTAL BILIRUBIN RESULTS ON APPROXIMATELY 30 PATIENTS GENERATED ON THE ARCHITECT ANALYZER. THE CUSTOMER STATED THE LEVEL I CONTROL FOR TOTAL BILIRUBIN WAS OUT OF RANGE LOW IN THE AFTERNOON. THE MORNING QC HAD BEEN ACCEPTABLE. THE CUSTOMER REPEATED ALL PATIENT SAMPLES FROM THE MORNING. THE RESULTS PROVIDE WERE FOR 12 SAMPLES: ON (B)(6) ORIGINAL = 10.8 /AMENDED =14 (CURRENT ACCEPTANCE RANGE <20UMOL/L); SID (B)(6) = 3.1 /AMENDED= 6; SID (B)(6) =2.4 / AMENDED = 5; SID (B)(6) = 4.4 /AMENDED= 7; SID (B)(6) = 9.9 / AMENDED =14; SID (B)(6) = 5.3 / AMENDED= 8; SID (B)(6) = 3 / RERUN: 5; SID (B)(6) = 13 / RERUN = 16; SID (B)(6) = 7 / RERUN = 10; SID (B)(6) = 12 / RERUN = 15; SID (B)(6) = 7 / RERUN = 9; SID (B)(6) = 9 / RERUN = 12UMOL/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9751255 ↗
  • MalfunctionFeb 13, 2020Low Test Results; Low Test Results

    DURING A PATIENT COMPARISON STUDY THE CUSTOMER OBSERVED FALSELY DEPRESSED TOTAL BILIRUBIN RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR SEVERAL SAMPLES UPON USE OF CALIBRATOR LOT 05647FD01. THE BIASES OBSERVED WITH THIS CALIBRATOR LOT RANGED FROM 2.9% TO 22.5%. THE CUSTOMER ALSO REPORTED AN OBSERVED NEGATIVE SHIFT IN QC RESULTS DURING THIS TIME. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9707745 ↗
  • MalfunctionOct 2, 2019Low Test Results; Low Test Results

    THE CUSTOMER REPORTED MULTIPLE FALSE DECREASED TOTAL BILIRUBIN RESULTS WHEN PROCESSING ON THE ARCHITECT C16000. INITIAL RESULTS RANGED FROM 1.9 TO 38.4 AND RETEST RESULTS RANGED FROM 8 TO 46. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9143743 ↗
  • MalfunctionMay 7, 2019High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED TOTAL BILIRUBIN ON ONE PATIENT GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: SID (B)(6) = 107UMOL/L / REPEATED ON DIFFERENT ANALYZER = 8.5 (REFERENCE RANGE = <25UMOL/L). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8587548 ↗
  • MalfunctionApr 18, 2019Low Test Results; Low Test Results

    WHILE USING THE CLINICAL CHEMISTRY TOTAL BILIRUBIN ASSAY, THE CUSTOMER INDICATED THAT DECREASED BILIRUBIN RESULTS WERE GENERATED AFTER CALIBRATION WAS COMPLETED WITH A NEW LOT OF CALIBRATOR. THE CUSTOMER PROVIDED THE FOLLOWING DATA FOR AN ADULT PATIENT (PROVIDED REFERENCE RANGE 3 - 21 UMOL/L): NOTE: THE RETEST RESULTS WERE GENERATED AFTER CALIBRATION WAS COMPLETED WITH NEW LOT OF CALIBRATOR. THE CUSTOMER INDICATED THE INITIAL RESULTS WERE CONSIDERED CORRECT. SID (B)(6), INITIAL 10.7, AFTER RETEST 7.0, SID (B)(6), INITIAL 5.4, RETEST 3.0, SID (B)(6), INITIAL 9.3, RETEST 4.9, SID (B)(6), INITIAL 7.0, RETEST 3.3, SID (B)(6), INITIAL 7.1, RETEST 4.1, SID (B)(6), INITIAL 10.6, RETEST 5.3, SID (B)(6), INITIAL 13.2, RETEST 9.3, SID (B)(6), INITIAL 17.9 , RETEST 11.4, SID (B)(6), INITIAL 26.6 , RETEST 20.4, SID (B)(6), INITIAL 21.1, RETEST 13.2. THERE HAS BEEN NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8527450 ↗
  • MalfunctionApr 2, 2019High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED TOTAL BILIRUBIN RESULTS ON TWO PATIENTS GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: SID (B)(6) INITIAL = 152.2 / REPEAT 6.0UMOL/L; SID (B)(6) INITIAL = 110.9 / REPEAT =13.3UMOL//L. THE MEDIC QUESTIONED THE RESULTS PRIOR TO CLINICAL DECISIONS BEING MADE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8475150 ↗
  • MalfunctionMar 8, 2019Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSELY DEPRESSED TOTAL BILIRUBIN RESULTS ON 44 PATIENTS GENERATED ON THE ARCHITECT ANALYZER. THE CUSTOMER REPORTED A RESULT OF <0.1MG/DL; THEN DURING HER LUNCH ANOTHER TECH HAD 2 MORE RESULT <0.1 AND QUESTIONED THEM. THEY REPLACED THE KIT AND RETESTED THE THREE SAMPLES WITH ALL GENERATING 0.2 (0.3-1.2MG/DL). THE CUSTOMER THEN REPEATED ALL REMAINING 44 SAMPLES FROM (B)(6) 2019 BETWEEN 07:00 TO 13:30, RESULTS RANGED FROM 0.1 TO 1.0. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8405229 ↗
  • MalfunctionNov 2, 2018Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED RESULTS FOR A PEDIATRIC PATIENT WHILE USING CLINICAL CHEMISTRY TOTAL BILIRUBIN REAGENTS WITH THE ARCHITECT C8000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED. THE CUSTOMER USES NORMAL RANGE LESS THAN OR EQUAL TO 1.9 NG/ML. SID (B)(4), INITIAL 0.7 NG/DL, REPEAT 6.8, 6.8 NG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8034914 ↗
  • MalfunctionOct 10, 2018Low Test Results; Low Test Results

    THE CUSTOMER STATED THAT A FALSELY DECREASED CLINICAL CHEMISTRY TOTAL BILIRUBIN RESULT OF 1.7 MG/DL WAS GENERATED FOR A PEDIATRIC PATIENT SAMPLE TESTED ON THE ARCHITECT C8000 SYSTEM. THE RESULT WAS REPORTED TO THE PHYSICIAN ON (B)(6) 2018. THE SAME SAMPLE WAS RETESTED ON (B)(6) 2018 AND THE RESULT WAS 9.6 MG/DL. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7953181 ↗
  • MalfunctionJul 30, 2018Low Test Results; Low Test Results

    A PHYSICIAN QUESTIONED AN ARCHITECT TOTAL BILIRUBIN RESULT OF 0.5 MG/DL FOR A NEONATAL PATIENT. THE SAMPLE WAS SHORT VOLUME BUT NO ERROR MESSAGE WAS GENERATED. A NEW SAMPLE WAS DRAWN AND THE ARCHITECT TOTAL BILIRUBIN RESULT WAS 11.8 MG/DL. THE CUSTOMER STATED NO ADVERSE IMPACT TO PATIENT MANAGEMENT OCCURRED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7731270 ↗
  • MalfunctionMay 22, 2018High Test Results; High Test Results

    THE CUSTOMER REPORTED IMPRECISE AND FALSELY ELEVATED ARCHITECT TOTAL BILIRUBIN RESULTS FOR ONE PATIENT SAMPLE. THE SAMPLE WAS TESTED MULTIPLE TIMES ON MULTIPLE ANALYZERS GENERATING THE FOLLOWING RESULTS: (B)(6). THE CUSTOMER USES A NORMAL RANGE OF LESS THAN 20 UMOL/L. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7534000 ↗
  • MalfunctionApr 5, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    THE CUSTOMER OBSERVED MULTIPLE FALSELY ELEVATED TOTAL BILIRUBIN RESULTS WHILE USING THE CLINICAL CHEMISTRY TOTAL BILIRUBIN REAGENTS. THE FOLLOWING DATA WAS PROVIDED. SID (B)(6) INITIAL 1.636 MG/DL, REPEAT 0.509 MG/DL, SID (B)(6) INITIAL 1.807 MG/DL, REPEAT 0.717 MG/DL, SID (B)(6) INITIAL 1.646 MG/DL, REPEAT 0.493 MG/DL, SID (B)(6) INITIAL 1.795 MG/DL, REPEAT 0.674 MG/DL, SID (B)(6) INITIAL 1.582 MG/DL, REPEAT 0.443 MG/DL, SID (B)(6) INITIAL 1.581 MG/DL, REPEAT 0.431 MG/DL, SID (B)(6) INITIAL 2.088 MG/DL, REPEAT 1.045 MG/DL, SID (B)(6) INITIAL 1.711 MG/DL, REPEAT 0.578 MG/DL, SID (B)(6) INITIAL 1.654 MG/DL, REPEAT 0.512 MG/DL, SID (B)(6) INITIAL 1.778 MG/DL, REPEAT 0.66 MG/DL, SID (B)(6) INITIAL 1.835 MG/DL, REPEAT 0.781 MG/DL, SID (B)(6) INITIAL 1.648 MG/DL, REPEAT 0.528 MG/DL, SID (B)(6) INITIAL 1.623 MG/DL, REPEAT 0.453 MG/DL, SID (B)(6) INITIAL 1.605 MG/DL, REPEAT 0.472 MG/DL, SID (B)(6) INITIAL 1.474 MG/DL, REPEAT 0.34 MG/DL, SID (B)(6) INITIAL 1.699 MG/DL, REPEAT 0.592 MG/DL, SID (B)(6) INITIAL 1.548 MG/DL, REPEAT 0.427 MG/DL, SID (B)(6) INITIAL 1.734 MG/DL, REPEAT 0.608 MG/DL, SID (B)(6) INITIAL 1.425 MG/DL, REPEAT 0.278 MG/DL, SID (B)(6) INITIAL 1.756 MG/DL, REPEAT 0.625 MG/

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7400545 ↗
  • MalfunctionMar 15, 2018Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    THE CUSTOMER OBSERVED MULTIPLE FALSELY ELEVATED TOTAL BILIRUBIN RESULTS WHILE USING THE CLINICAL CHEMISTRY TOTAL BILIRUBIN REAGENTS. THE FOLLOWING DATA WAS PROVIDED. (B)(6). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7342841 ↗
  • MalfunctionAug 14, 2017High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED BILIRUBIN RESULTS WHILE USING THE CLINICAL CHEMISTRY TOTAL BILIRUBIN ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING EXAMPLE DATA FROM (B)(6) 2017: PATIENT 1: INITIAL 1.4, RETEST 0.5. PATIENT 2: 2, RETEST 1.1. PATIENT 3: 1.3, RETEST 0.4. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6791660 ↗

FDA clearance and classification

The FDA regulates product code CIG (diazo colorimetry, bilirubin) under 21 CFR 862.1110 as a Class II device. 21 CFR 862.1110 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ARCHITECT 6L45-41

Has the ARCHITECT 6L45-41 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740011383), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ARCHITECT 6L45-41?

The FDA's MAUDE database lists 18 adverse event reports for this device, received between Aug 14, 2017 and Nov 2, 2020. In those reports, the most common specific device problems are “Low Test Results” (9), “High Test Results” (7) and “Incorrect Or Inadequate Test Results” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ARCHITECT 6L45-41?

None of the reports are classified as deaths, 0 as injuries and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ARCHITECT 6L45-41, and is it still sold?

The FDA's device database lists Abbott Laboratories as the company responsible for it (brand ARCHITECT, model 6L45-41). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DIAZO COLORIMETRY, BILIRUBIN devices

All DIAZO COLORIMETRY, BILIRUBIN devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page