VITROS 1612365Recalls & reported problems
Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products BuBc Slides · Class 2
All models of VITROS (71)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10758750009398), received between Feb 12, 2026 and Feb 12, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 13, 2020Class IITerminated
Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides
Z-2874-2020 · Root cause: Other · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2026Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER THAN EXPECTED UNCONJUGATED BILIRUBIN (BU) RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF A NON-VITROS MAS QUALITY CONTROL (QC) FLUID USING VITROS CHEMISTRY PRODUCTS BU SLIDES LOT 0207-0486-4476 ON A VITROS XT 7600 INTEGRATED SYSTEM. MAS BILI LOT 2609 (LEVEL 3) RESULTS OF 6.61, 8.82, 8.35, 8.13, 8.49, 8.59, 8.34, 7.99, 7.92, 7.88, 7.92, 7.85, 7.72, 7.53, 7.60, 7.43, 7.53, 7.19, 7.16, AND 6.66 MG/DL VERSUS THE EXPECTED RESULT OF 12.06 MG/DL (MIDPOINT OF THE MAS PACKAGE INSERT RANGE) BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER-THAN-EXPECTED VITROS BU RESULTS WERE FROM A NON-PATIENT FLUID AND WERE NOT REPORTED FROM THE LABORATORY. THE CUSTOMER CONFIRMED THERE HAVE BEEN NO COMPLAINTS FROM PHYSICIANS REGARDING PATIENT RESULTS. THERE HAS BEEN NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24328120 ↗
FDA clearance and classification
The FDA regulates product code CIG (diazo colorimetry, bilirubin) under 21 CFR 862.1110 as a Class II device. 21 CFR 862.1110 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VITROS 1612365
Has the VITROS 1612365 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2874-2020 (Class II), began on Jul 13, 2020: Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the VITROS 1612365?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 12, 2026. That report lists device problem “Low Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VITROS 1612365?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VITROS 1612365, and is it still sold?
The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 1612365). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DIAZO COLORIMETRY, BILIRUBIN devices
- Dimension® Flex® reagent cartridge TBI 10444957Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN2 recalls4 reports
- Dimension Vista® Flex® reagent cartridge TBIL 10445146Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall7 reports
- VITROS 8383051Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall5 reports
- ARCHITECT 6L45-21Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN46 reports
- ARCHITECT 6L45-22Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN30 reports
- ARCHITECT 6L45-41Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN18 reports
- Bilirubin Total Gen3 5795397190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN16 reports
- ARCHITECT 6L45-42Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page