Dimension Vista® Flex® reagent cartridge TBIL 10445146Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Total Bilirubin Flex® reagent cartridge, 640 Tests · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00842768015717), received between Aug 22, 2018 and Jan 16, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K061719); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 30, 2024Class IIOpen, Classified
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Z-3065-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results7
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2024Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE FALSELY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME DIMENSION VISTA 500 SYSTEM AND OBTAINED HIGHER RESULT. THE REPROCESSED RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18520164 ↗MalfunctionJan 16, 2024Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE FALSELY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME DIMENSION VISTA 500 SYSTEM AND OBTAINED HIGHER RESULT. THE REPROCESSED RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18519481 ↗MalfunctionJan 16, 2024Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE FALSELY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME DIMENSION VISTA 500 SYSTEM AND OBTAINED HIGHER RESULT. THE REPROCESSED RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18518987 ↗MalfunctionJan 16, 2024Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT SAMPLE RESULT WAS OBTAINED ON A DIMENSION VISTA 500 SYSTEM. THE FALSELY DEPRESSED RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME DIMENSION VISTA 500 SYSTEM AND OBTAINED HIGHER RESULT. THE REPROCESSED RESULT WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) PATIENT RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18518982 ↗MalfunctionDec 20, 2018Low Test Results; Low Test Results
A DISCORDANT, FALSELY DEPRESSED TOTAL BILIRUBIN RESULT WITH A E111 HEMOLYSIS ERROR FLAG WAS OBTAINED ON A NEONATE PATIENT SAMPLE ON THE DIMENSION VISTA INSTRUMENT. THE DISCORDANT RESULT WAS REPORTED TO THE PHYSICIAN(S) WHO QUESTIONED THE RESULT. A NEW SAMPLE DRAWN FROM THE SAME PATIENT WAS PROCESSED THE NEXT DAY ON THE ORIGINAL INSTRUMENT AND A HIGHER RESULT WAS OBTAINED. A CORRECTED REPORT WAS ISSUED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) RESULT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8185331 ↗MalfunctionAug 22, 2018Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 SYSTEM. THE RESULT WAS FLAGGED FOR HEMOLYSIS (E111 FLAG) BUT WAS REPORTED TO THE PHYSICIAN. A NEW SAMPLE DRAWN THE SAME DAY WAS PROCESSED AND A HIGHER RESULT WAS OBTAINED CONSISTENT WITH THE CLINICAL PICTURE AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DEPRESSED TBIL RESULT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806495 ↗MalfunctionAug 22, 2018Low Test Results; Low Test Results
A DISCORDANT FALSELY DEPRESSED TOTAL BILIRUBIN (TBIL) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE DIMENSION VISTA 1500 SYSTEM. THE RESULT WAS FLAGGED FOR HEMOLYSIS (E111 FLAG) BUT WAS REPORTED TO THE PHYSICIAN. A NEW SAMPLE DRAWN THE SAME DAY WAS PROCESSED AND A HIGHER RESULT WAS OBTAINED CONSISTENT WITH THE CLINICAL PICTURE AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT DEPRESSED TBIL RESULT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7806482 ↗
FDA clearance and classification
Cleared through 510(k) K061719 on Aug 25, 2006 (decision: Substantially Equivalent). Submitted by Dade Behring, Inc. as “Dimension Vista Tbil Flex Reagent Cartridge, And Dimension Vista Tbil/Dbil Calibrator; Models K1167, Kc212”. The FDA review took 67 days. FDA record ↗
The FDA regulates product code CIG (diazo colorimetry, bilirubin) under 21 CFR 862.1110 as a Class II device. 21 CFR 862.1110 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K061719. One submission often covers many models, so check the product and lot details in each record.
- Z-1541-2020 · Feb 13, 2020 · TerminatedDimension Vista Total Bilirubin Flex reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is
Common questions about the Dimension Vista® Flex® reagent cartridge TBIL 10445146
Has the Dimension Vista® Flex® reagent cartridge TBIL 10445146 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3065-2024 (Class II), began on Jul 30, 2024: During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Dimension Vista® Flex® reagent cartridge TBIL 10445146?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Aug 22, 2018 and Jan 16, 2024. In those reports, the most common specific device problem is “Low Test Results” (7). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dimension Vista® Flex® reagent cartridge TBIL 10445146?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Dimension Vista® Flex® reagent cartridge TBIL 10445146 FDA approved?
It was cleared through 510(k) K061719 on Aug 25, 2006. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Dimension Vista® Flex® reagent cartridge TBIL 10445146, and is it still sold?
The FDA's device database lists Siemens Healthcare Diagnostics Inc. as the company responsible for it (brand Dimension Vista® Flex® reagent cartridge TBIL, model 10445146). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DIAZO COLORIMETRY, BILIRUBIN devices
- Dimension® Flex® reagent cartridge TBI 10444957Siemens Healthcare Diagnostics Inc. · DIAZO COLORIMETRY, BILIRUBIN2 recalls4 reports
- VITROS 8383051Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall5 reports
- VITROS 1612365Ortho-Clinical Diagnostics, Inc. · DIAZO COLORIMETRY, BILIRUBIN1 recall1 report
- ARCHITECT 6L45-21Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN46 reports
- ARCHITECT 6L45-22Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN30 reports
- ARCHITECT 6L45-41Abbott Laboratories · DIAZO COLORIMETRY, BILIRUBIN18 reports
- Bilirubin Total Gen3 5795397190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN16 reports
- ARCHITECT 6L45-42Abbott GmbH · DIAZO COLORIMETRY, BILIRUBIN15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page