CELL-DYN 08H00-03Recalls & reported problems
Abbott Laboratories · CELL-DYN Sapphire · Class 2
All models of CELL-DYN (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00380740016623), received between Nov 8, 2017 and Jul 15, 2022 — classified as 7 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- High Test Results3
- Thermal Decomposition of Device2
- Leak/Splash1
- Low Test Results1
- Smoking1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions1
- No Patient Involvement1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2022Smoking; Smoking
THE ACCOUNT OBSERVED SMOKE COMING FROM THE BACK OF THE CELL-DYN SAPPHIRE. THE ACCOUNT STATED AFTER A STRANGE SOUND FROM THE LEFT BACKSIDE OF THE ANALYZER, SMOKE WAS SEEN AND A STRANGE ODOR. THE ACCOUNT UNPLUGGED THE POWER CORD. THE ACCOUNT TECHNICAL SERVICE CONFIRMED NOTHING WAS WRONG WITH THE WALL SOCKET OR POWER AT THE ACCOUNT. NO SPECIFIC PATIENT INFORMATION PROVIDED AS THERE WAS NO PATIENT INVOLVEMENT. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15029408 ↗MalfunctionJun 18, 2020Leak/Splash; Leak/Splash
THE CUSTOMER WAS OBSERVING MULTIPLE ISSUES WITH THE CELLDYN SAPPHIRE WITH LIQUID LEAKING OUT OF TUBING BEHIND THE VALVE OF THE ANALYZER. THE CUSTOMER WAS WORKING WITH A FIELD SERVICE ENGINEER (FSE) OVER THE PHONE NEAR THE LEAK AND IT SPRAYED OUT OF THE TUBING. THE CUSTOMER FELT A DROP ON HIS FOREHEAD. THE CUSTOMER WAS NOT WEARING GOGGLES OR A FACE SHIELD. THERE WAS NO INJURY/INTERVENTION TO THE CUSTOMER REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10170382 ↗MalfunctionJun 17, 2020High Test Results; High Test Results
THE ACCOUNT GENERATED FALSE ELEVATED HEMOGLOBIN OF 19 G/DL WITH THE CELL-DYN SAPPHIRE ON PATIENT SAMPLE THAT GENERATED HEMOGLOBIN OF 6 G/DL ON CELL-DYN RUBY. THE ACCOUNT ACCEPTED THE CELL-DYN RUBY RESULTS TO BE CORRECT. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10160999 ↗MalfunctionFeb 10, 2020Thermal Decomposition of Device; Thermal Decomposition of Device
THE CUSTOMER REPORTED THE CELL-DYN SAPPHIRE POWER CORD FROM THE OUTLET IS MELTED AT THE CONNECTION TO THE BACK OF THE UPS. THE CUSTOMER REPORTED THE INSTRUMENT MADE A BEEPING SOUND AND THE TECHNICIAN REPORTED A PLASTIC ODOR, AFTER WHICH THEY UNPLUGGED THE CORD. NO INJURY WAS REPORTED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9688923 ↗MalfunctionMar 3, 2019High Test Results; Low Test Results; High Test Results; Low Test Results
THE CELL-DYN SAPPHIRE ANALYZER GENERATED A FALSELY ELEVATED PLATELET RESULTS AND LOW %NEUTROPHIL RESULTS. SAMPLE ID (B)(6) GENERATED PLATELET RESULTS OF 21.8, 16.5, 14.9 AND 16.62. ADDITIONALLY, MULTIPLE SAMPLES GENERATED %NEUTROPHIL RESULTS LESS THAN 2.5%. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8384856 ↗MalfunctionJan 2, 2019Thermal Decomposition of Device; Thermal Decomposition of Device
DURING INSTALLATION AFTER DOWNLOADING THE AOS FIRMWARE ON THE CELL-DYN SAPPHIRE ANALYZER, UPON RESTARTING THE INSTRUMENT THERE WAS A BURNING SMELL FROM THE MACHINE AND VISIBLE SMOKE. THE INSTRUMENT WAS SHUT DOWN. ON THE DATA STATION, ERROR 1421 LASER CURRENT TOO HIGH. THE ANALYZER WAS NOT INSTALLED AT THE CUSTOMER SITE. NO INJURIES OCCURRED.
Patient problems: No Patient Involvement
MAUDE report 8209326 ↗MalfunctionNov 8, 2017High Test Results; High Test Results
THE CUSTOMER STATED THAT A FALSELY ELEVATED WBC RESULT OF 1630/UL WAS GENERATED IN THE CLOSED MODE ON THE CELL-DYN SAPPHIRE ANALYZER. THE PATIENT WAS KNOWN TO HAVE APLASIA. THE SAMPLE HAD FOLLOWED A VERY HIGH SAMPLE WHICH HAD GENERATED A RESULT OF 258,000/UL. THE CUSTOMER RERAN THE SAMPLE IN THE OPEN MODE AND A WBC RESULT OF 283/UL WAS GENERATED WHICH WAS THE EXPECTED RESULT. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7011358 ↗
Other Counter, differential cell devices
- Cytomics FC500 UPS (CXP)Beckman Coulter, Inc. · Counter, Differential Cell1,196 reports
- CELL-DYN 08H67-01Abbott Laboratories · Counter, differential cell6 recalls269 reports
- Alinity 09P6801Abbott Laboratories · Counter, differential cell2 recalls96 reports
- FACSCanto 338960Becton, Dickinson And Company · Counter, differential cell1 recall117 reports
- Cytomics FC500 UPS (MPL)Beckman Coulter, Inc. · Counter, Differential Cell141 reports
- CELL-DYN 08H00-01Abbott Laboratories · Counter, differential cell1 recall90 reports
- CELL-DYN 09H39-01Abbott Laboratories · Counter, differential cell123 reports
- BD Multitest 662995Becton, Dickinson And Company · Counter, differential cell2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026