Cytomics FC500 UPS (CXP)Recalls & reported problems

Beckman Coulter, Inc. · Cytomics FC 500 with CXP Software · Class 2

All models of Cytomics FC500 (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Beckman Coulter Cytomics FC500 flow cytometer has 1,196 adverse event reports, with no deaths and 3 injuries. Most reports describe imprecise or incorrect readings, and some narratives mention electronic board errors and signal loss during startup quality checks. Some reports note that no erroneous patient results were found. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,196
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,196 adverse event reports that list this exact device (DI 15099590572839), received between Oct 26, 2018 and Nov 20, 2024 — classified as 3 injuries, 1,193 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1,190 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1,190
  • Device Remains Activated1
  • Electrical /Electronic Property Problem1
  • Fire1
  • Inadequate or Insufficient Training1
  • No Apparent Adverse Event1
  • Output Problem1
  • Retraction Problem1

Patient problems

  • No Consequences Or Impact To Patient1,005
  • No Clinical Signs, Symptoms or Conditions190
  • Test Result1

Recent adverse event reports

The 25 most recent of 1,196 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED ABNORMAL HISTOGRAMS ON THE FC500 FLOW CYTOMETER. THE FIELD SERVICE ENGINEER (FSE) FOUND THERE WAS A TARPON AMP BOARD ERROR AT FORWARD SCATTER (FS). THE EVENT OCCURRED DURING START THE FLOW CHECK (QUALITY CONTROL) RUN. THERE WAS NO REPORT OF ERRONEOUS PATIENT RESULTS ASSOCIATED WITH THE EVENT. THERE WAS NO REPORTED DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20738111 ↗
  • MalfunctionJul 18, 2024No Apparent Adverse Event; No Apparent Adverse Event

    THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE FL5 FLUORESCENCE CHANNEL ON THEIR FC500 FLOW CYTOMETER WHILE RUNNING FLOWCHECK AT START-UP. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. PATIENT SAMPLES WERE NOT RUN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19778802 ↗
  • MalfunctionJul 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE FL5 FLUORESCENCE CHANNEL ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19776768 ↗
  • MalfunctionMar 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED FL3 AND FL5 AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. THERE WAS NO REPORT OF ERRONEOUS RESULTS GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18944227 ↗
  • MalfunctionJul 20, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED TARPON AMP BOARD ERRORS ON THE "FORWARD" SCATTER (FS), FL1 AND A FLOW CHECK SIGNAL DRIFT AT FL3 ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17364165 ↗
  • MalfunctionJul 12, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED NO SIGNAL AT THE FL4 POSITION ON THEIR FC 500 FLOW CYTOMETER WHILE RUNNING FLOW CHECK. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17311809 ↗
  • MalfunctionJun 26, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED FL4 PE TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17203053 ↗
  • MalfunctionJun 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED INTERMITTENT SIGNAL LOSS FOR FL4 AND UNSTABLE SIGNAL FOR FL3 WHEN RUNNING FLOWCHECK ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17091997 ↗
  • MalfunctionMay 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED FL1 AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER WHILE RUNNING FLOW CHECK. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16887916 ↗
  • MalfunctionApr 24, 2023Fire; Fire

    THE CUSTOMER REPORTED HEARING A POPPING ELECTRICAL NOISE ON THEIR FC 500 FLOW CYTOMETER. THE UNIT (POWER SUPPLY CART) CAUGHT FIRE AS IT SAT IDLE. THE CUSTOMER REPORTED SEEING FLAMES THAT WERE PUT OUT WITH A FIRE EXTINGUISHER. THERE WAS NO DIRECT HARM TO LABORATORY OR LABORATORY PERSONNEL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16799586 ↗
  • MalfunctionApr 10, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED FL4 TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER, WHILE PROCESSING BOTH FLOW CHECK AND QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16709296 ↗
  • MalfunctionMar 29, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED INTERMITTENT FL4 X-MEAN SIGNAL SHIFT AND FL4 TARPON AMP BOARD ERRORS ON THEIR FC 500 FLOW CYTOMETER, WHILE PROCESSING BOTH FLOW CHECK AND QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16638285 ↗
  • MalfunctionMar 9, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED TARPON AMP BOARD ERRORS AT THE FL3 AND SS (SIDE SCATTER) POSITION ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16512059 ↗
  • MalfunctionJan 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED AMP BOARD ERROR AT THE FLUORESCENCE CHANNELS (FL1,FL2,FL3, FL4 AND FL5) WHEN PROCESSING FLOW CHECK, QUALITY CONTROL. AND PATIENT SAMPLES ON THEIR FC500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16099811 ↗
  • MalfunctionJan 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT FLOWSET FLUORESCENCE CHANNEL ( FL1 AT 2%) AND FL1 HAS SHIFTED TO THE LEFT OF THE TARGETS WHEN RUNNING QUALITY CONTROL ON THEIR FC500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16096901 ↗
  • MalfunctionDec 14, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE FIELD SERVICE ENGINEER (FSE) REPORTED SIGNAL SHIFT AT FL2 AND FL3 POSITIONS ON THE FC500 FLOW CYTOMETER WHILE RUNNING FLOW CHECK CONTROL WHILE SERVICING THE INSTRUMENT. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT. THERE WERE NO ERRONEOUS PATIENT RESULTS GENERATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15982324 ↗
  • MalfunctionDec 5, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A FORWARD SCATTER (FS) SIGNAL SHIFT ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15918797 ↗
  • MalfunctionNov 22, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED AMP BOARD WARNINGS AT THE FORWARD SCATTER (FS) POSITION OF THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF ERRONEOUS PATIENT RESULTS ASSOCIATED WITH THE EVENT. THERE WAS NO REPORTED DEATH, INJURY OR CHANGE TO PATIENT TREATMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15849488 ↗
  • MalfunctionOct 21, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED INTERMITTENT SS AMP BOARD WARNING ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15654108 ↗
  • MalfunctionOct 13, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED SIGNAL DRIFT AT THE FORWARD SCATTER (FS), AND FL5 POSITIONS ON THEIR FC 500 FLOW CYTOMETER WHILE RUNNING QUALITY CONTROL. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15594938 ↗
  • MalfunctionSep 22, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE FSE REPORTED SIGNAL/PEAK SHIFTING AT THE FL2 AND FL3 WHEN RUNNING FLOW CHECK DURING A PM (PREVENTATIVE MAINTENANCE) ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15464578 ↗
  • MalfunctionSep 16, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER THAT THE PLOTS ARE COMING OFF THE BASELINE AT THE FL2 POSITION OF THEIR FC 500 FLOW CYTOMETER. THERE WAS AN INTERMITTENT UPWARD SIGNAL SHIFT AT THE FL2 POSITION CAUSING ERRONEOUS RESULTS FOR THEIR LEUKEMIA/LYMPHOMA LDT. THE CUSTOMER DISCARDED THE RESULTS AND RE-RAN THE SAMPLES ON A DIFFERENT INSTRUMENT TO OBTAIN CORRECT RESULTS. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15442246 ↗
  • MalfunctionSep 14, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED AMP BOARD ERRORS AT THE SS (SIDE SCATTER) POSITION ON THEIR FC 500 FLOW CYTOMETER WHEN RUNNING FLOW CHECK (CONTROL). THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15420389 ↗
  • MalfunctionSep 13, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A PROBLEM AT THE FL4 POSITION ON THE FC500 ON THEIR FC 500 FLOW CYTOMETER. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15414650 ↗
  • MalfunctionAug 26, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THE POPULATIONS AT THE FL4 POSITION ON THEIR FC 500 FLOW CYTOMETER WERE NOT BEING GATED CORRECTLY CAUSING ERRONEOUS CD34+ RESULTS. ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LAB. THERE WAS NO REPORT OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15306968 ↗

FDA clearance and classification

The FDA regulates product code GKZ (counter, differential cell) under 21 CFR 864.5220 as a Class II device. 21 CFR 864.5220 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Cytomics FC500 UPS (CXP)

Has the Cytomics FC500 UPS (CXP) been recalled?

No FDA recall record in our data names this exact device identifier (DI 15099590572839), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Cytomics FC500 UPS (CXP)?

The FDA's MAUDE database lists 1,196 adverse event reports for this device, received between Oct 26, 2018 and Nov 20, 2024. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (1,190), “Device Remains Activated” (1) and “Electrical /Electronic Property Problem” (1); the most common patient problem is “Test Result” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cytomics FC500 UPS (CXP)?

None of the reports are classified as deaths, 3 as injuries and 1,193 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Cytomics FC500 UPS (CXP), and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand Cytomics FC500, model UPS (CXP)). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Counter, Differential Cell devices

All Counter, Differential Cell devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page