Alinity 08P4920Recalls & reported problems

Abbott GmbH · Alinity i CA 125 II Reagent Kit 200 Tests · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740131043), received between Sep 10, 2019 and Mar 31, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED ALINITY I CA 125 II RESULT FOR A 75 YEAR OLD FEMALE PATIENT, WHICH WAS INCONSISTENT WITH THE PATIENT¿S HISTORICAL RESULTS. IT IS UNKNOWN WHETHER THE PATIENT HAS EPITHELIAL OVARIAN CANCER. THE PATIENT QUESTIONED THE RESULT, AND THE SAMPLE WAS REPEATED, YIELDING HIGHER RESULTS. THE FOLLOWING DATA WERE PROVIDED: SID (B)(6): (B)(6) 2026 INITIAL RESULT = 16.8 U/ML (NEGATIVE). REPEAT RESULT = >1000 U/ML. REPEAT RESULT (1:10 DILUTION) = >10,000 U/ML. HISTORICAL RESULTS: (B)(6) 2025 RESULT = 8998.0 U/ML (POSITIVE). (B)(6) 2025 RESULT = 4001.0 U/ML (POSITIVE). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24746116 ↗
  • MalfunctionSep 10, 2019High Test Results; High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I CA125 RESULT ON ONE PATIENT. RESULTS PROVIDED: (B)(6) 2019 = 136 U/ML; NEW SAMPLE TESTED AT ANOTHER LABORATORY (METHOD UNKNOWN) = 16; NEW SAMPLE FROM OTHER LABORATORY PROCESSED ON THIS ALINITY = 16 U/ML. IT IS UNKNOWN IF THE PATIENT IS DIAGNOSED WITH OVARIAN CANCER. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient