Alinity 08P4920Recalls & reported problems
Abbott GmbH · Alinity i CA 125 II Reagent Kit 200 Tests · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740131043), received between Sep 10, 2019 and Mar 31, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2026Low Test Results
THE CUSTOMER REPORTED A FALSELY DECREASED ALINITY I CA 125 II RESULT FOR A 75 YEAR OLD FEMALE PATIENT, WHICH WAS INCONSISTENT WITH THE PATIENT¿S HISTORICAL RESULTS. IT IS UNKNOWN WHETHER THE PATIENT HAS EPITHELIAL OVARIAN CANCER. THE PATIENT QUESTIONED THE RESULT, AND THE SAMPLE WAS REPEATED, YIELDING HIGHER RESULTS. THE FOLLOWING DATA WERE PROVIDED: SID (B)(6): (B)(6) 2026 INITIAL RESULT = 16.8 U/ML (NEGATIVE). REPEAT RESULT = >1000 U/ML. REPEAT RESULT (1:10 DILUTION) = >10,000 U/ML. HISTORICAL RESULTS: (B)(6) 2025 RESULT = 8998.0 U/ML (POSITIVE). (B)(6) 2025 RESULT = 4001.0 U/ML (POSITIVE). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746116 ↗MalfunctionSep 10, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I CA125 RESULT ON ONE PATIENT. RESULTS PROVIDED: (B)(6) 2019 = 136 U/ML; NEW SAMPLE TESTED AT ANOTHER LABORATORY (METHOD UNKNOWN) = 16; NEW SAMPLE FROM OTHER LABORATORY PROCESSED ON THIS ALINITY = 16 U/ML. IT IS UNKNOWN IF THE PATIENT IS DIAGNOSED WITH OVARIAN CANCER. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8983835 ↗
Other TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125) devices
- ADVIA Centaur® CA 125II 10310443Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)1 recall10 reports
- ADVIA Centaur® CA 125 II 10315696Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)1 recall0 reports
- ARCHITECT 02K4529Abbott GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)8 reports
- Atellica IM CA 125II 10995482Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)3 reports
- ARCHITECT 02K4539Abbott GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)3 reports
- Elecsys CA 125 II Assay 11776223161Roche Diagnostics GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
- IMMULITE® 2000 Systems OV OM-MA (CA125) 10380972Siemens Healthcare Diagnostics Products Ltd · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
- Atellica IM CA 125II 10995481Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.