Skip to content

Elecsys CA 125 II Assay 11776223161Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04015630945795), received between Sep 25, 2024 and Nov 21, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS CA 125 II (CA 125 II) ON A COBAS 6000 E601 MODULE. THE INITIAL RESULT WAS 5.41 U/ML. THIS RESULT WAS REJECTED BY THE PHYSICIAN. A NEW SAMPLE WAS OBTAINED, AND THE RESULT WAS 507.6 U/ML. THIS RESULT WAS BELIEVED TO BE CORRECT. THE ORIGINAL SAMPLE WAS REPEATED WITH A RESULT OF 487.8 U/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23621615 ↗
  • MalfunctionSep 25, 2024High Test Results; High Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH THE ELECSYS CA 125 II ASSAY ON A COBAS E411 RACK. THE INITIAL VALUES WERE REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED BY A PHYSICIAN AS THEY WERE HIGHER THAN EXPECTED. THE SAMPLES WERE REPEATED AND THESE VALUES WERE DEEMED CORRECT. THE FIRST SAMPLE INITIALLY RESULTED IN A CA 125 VALUE OF 974.9 U/ML AND IT REPEATED AS 676.2 U/ML. THE SECOND SAMPLE INITIALLY RESULTED IN A CA 125 VALUE OF 22259 U/ML AND IT REPEATED AS 12714 U/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305700 ↗

Other TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125) devices

All TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125) devices →

Sources and updates