ARCHITECT 02K4529Recalls & reported problems
Abbott GmbH · ARCHITECT CA 125 II Reagent Kit · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380740145378), received between Jan 31, 2019 and Feb 15, 2021 — classified as 8 malfunctions.
The most frequently reported specific device problem is “High Test Results” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results8
Patient problems
- No Known Impact Or Consequence To Patient7
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 15, 2021High Test Results; High Test Results
THE CUSTOMER STATED THAT A FALSELY ELEVATED ARCHITECT CA 125 RESULT OF 40.72 U/ML WAS GENERATED FOR A PATIENT SAMPLE THAT RETESTED AT 25.21 U/ML. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11330008 ↗MalfunctionJul 14, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ARCHITECT CA 125 II RESULT ON AN (B)(6) FEMALE PATIENT DIAGNOSED WITH OVARIAN CANCER AND SQUAMOUS CELL CARCINOMA. RESULTS PROVIDED: (B)(6) 2020 = 17.0 U/ML; (B)(6) 2020 = 30.3 U/ML; (B)(6) 2020 = 99.2 U/ML; (B)(6) 2020 = 9.0 U/ML (REFERENCE RANGE: <35.0 U/ML). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10269889 ↗MalfunctionJan 2, 2020High Test Results; High Test Results
THE ACCOUNT GENERATED FALSE ELEVATED ARCHITECT CA125 OF 83.80 U/ML ON (B)(6) 2019 WITH A FEMALE, (B)(6) PATIENT WITH ENDOMETRIAL CANCER. HISTORICALLY, THE PATIENT HAD CA125 OF 16.5 U/ML AND HAD A TOTAL HYSTERECTOMY. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO ETHNICITY OR RACE PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9542045 ↗MalfunctionNov 19, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT CA125 RESULTS ON ONE (B)(6) FEMALE DIAGNOSED WITH OVARIAN SEROUS CYSTIC TUMOR. THE RESULTS PROVIDED WERE: (B)(6) 2019 CA125 = 84.2U/ML / (B)(6) 2019 OVARECTOMY AND SURGERY TO REMOVE TUMOR / (B)(6) 2019 CA125 = 295.5U/ML; ON (B)(6) 2019 CEA = 0.8 NG/ML AND CA19-9 = 7.1 U/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9342069 ↗MalfunctionAug 5, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT CA 125 RESULTS ON ONE PATIENT DIAGNOSED WITH OVARIAN CANCER. THE RESULTS PROVIDED WERE: PT#1: ON (B)(6) 2017 = 85.6U/ML; ON (B)(6) 2017 = 74.3U/ML; ON (B)(6) 2019 = 665.7 U/ML, SAMPLE WAS RE-SPUN AND RETESTED = 674.9 AND 665.7U/ML; ON (B)(6) 2019 = 81.9 U/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8860768 ↗MalfunctionJun 12, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT CA125 RESULTS ON ONE (B)(6) YEAR OLD FEMALE DIAGNOSED WITH OVARIAN SEROUS CYSTIC TUMOR. THE RESULTS PROVIDED WERE: (B)(6) 2019 CA125 = 84.2U/ML / (B)(6) 2019 OVARECTOMY AND SURGERY TO REMOVE TUMOR / (B)(6) 2019 CA125 = 295.5U/ML; ON (B)(6) 2019 CEA = 0.8 NG/ML AND CA19-9 = 7.1 U/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8693542 ↗MalfunctionFeb 1, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSE ELEVATED CA125 INITIAL RESULT OF 53.4 U/ML, WHEN PROCESSING ON THE ARCHITECT I2000SR. ANOTHER TUBE FROM THE SAME SAMPLE GENERATED A RESULT OF 14.3 U/ML. THE SAMPLE WAS RETESTED AGAIN AND YIELDED RESULTS OF 55.2 U/ML AND 51.6 U/ML. A SECOND SAMPLE WAS DRAWN AND YIELDED RESULTS OF 42.1 U/ML AND 25.9 U/ML. RESULTS FROM ANOTHER ARCHITECT GAVE A RESULT OF 28 U/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8296473 ↗MalfunctionJan 31, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT CA 125 RESULTS ON ONE PATIENT. THE RESULTS PROVIDED WERE: (B)(6) 2019 ON SERIAL NUMBER (B)(4) = 53.4U/ML / ON SN (B)(4) = 14.3U/ML; REPEATED ON (B)(6) 2019, SN (B)(4) = 55.2U/ML / ON (B)(6) 2019, SN (B)(4) = 51.6U/ML; SECOND SAMPLE DRAWN TESTED ON (B)(4) = 42.1U/ML / (B)(4) = 25.9U/ML; ON ANOTHER ARCHITECT ANALYZER AT A DIFFERENT SITE = 28U/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8293658 ↗
Other TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125) devices
- ADVIA Centaur® CA 125II 10310443Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)1 recall10 reports
- ADVIA Centaur® CA 125 II 10315696Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)1 recall0 reports
- Atellica IM CA 125II 10995482Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)3 reports
- ARCHITECT 02K4539Abbott GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)3 reports
- Elecsys CA 125 II Assay 11776223161Roche Diagnostics GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
- Alinity 08P4920Abbott GmbH · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
- IMMULITE® 2000 Systems OV OM-MA (CA125) 10380972Siemens Healthcare Diagnostics Products Ltd · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
- Atellica IM CA 125II 10995481Siemens Healthcare Diagnostics Inc. · TEST, EPITHELIAL OVARIAN TUMOR-ASSOCIATED ANTIGEN (CA125)2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.