Alinity 08P4220Recalls & reported problems
Abbott GmbH · Alinity c Creatine Kinase Reagent Kit 1200 Tests · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740132590), received between Jul 27, 2020 and Feb 25, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2026Low Test Results
THE CUSTOMER OBSERVED A FALSELY DECREASED ALINITY C CREATINE KINASE RESULT FOR ONE 13-YEAR-OLD PATIENT. THE LOW RESULT WAS NOT REPORTED. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 29-200 U/L): SID (B)(6) PROCESSED ON (B)(6) 2026 INITIAL RESULT WAS 576 WITH DILUTION. REPEAT RESULT = 5993 U/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24443042 ↗MalfunctionJul 27, 2020Low Test Results; Low Test Results
THE CUSTOMER REPORTED A FALSELY DECREASED CREATINE KINASE (CK) RESULT GENERATED ON THE ALINITY C ANALYZER ON ONE PATIENT. RESULTS PROVIDED: ON (B)(6) 2020 SID (B)(6) = 95.6 U/L, REPEATED AT 95 U/L, ANOTHER LABORATORY = > 1000. IT WAS NOTED THAT THE SAMPLE WAS LIPEMIC AND THE CK RESULT WAS OBTAINED POST CENTRIFUGATION FOR 5 MINUTES / 10,000 RPM. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10326205 ↗
Other NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices
- ARCHITECT 7D63-21Abbott Laboratories · NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES1 recall1,435 reports
- Dimension Vista® Flex® reagent cartridge CKI 10464356Siemens Healthcare Diagnostics Inc. · NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES2 reports
- ARCHITECT 7D63-22Abbott GmbH · NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES1 report
- ADVIA Chemistry Creatine Kinase 10729780Siemens Healthcare Diagnostics Inc. · NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.