Skip to content

Alinity 08P4220Recalls & reported problems

Abbott GmbH · Alinity c Creatine Kinase Reagent Kit 1200 Tests · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740132590), received between Jul 27, 2020 and Feb 25, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Low Test Results

    THE CUSTOMER OBSERVED A FALSELY DECREASED ALINITY C CREATINE KINASE RESULT FOR ONE 13-YEAR-OLD PATIENT. THE LOW RESULT WAS NOT REPORTED. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 29-200 U/L): SID (B)(6) PROCESSED ON (B)(6) 2026 INITIAL RESULT WAS 576 WITH DILUTION. REPEAT RESULT = 5993 U/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443042 ↗
  • MalfunctionJul 27, 2020Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED CREATINE KINASE (CK) RESULT GENERATED ON THE ALINITY C ANALYZER ON ONE PATIENT. RESULTS PROVIDED: ON (B)(6) 2020 SID (B)(6) = 95.6 U/L, REPEATED AT 95 U/L, ANOTHER LABORATORY = > 1000. IT WAS NOTED THAT THE SAMPLE WAS LIPEMIC AND THE CK RESULT WAS OBTAINED POST CENTRIFUGATION FOR 5 MINUTES / 10,000 RPM. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10326205 ↗

Other NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices

All NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.