Skip to content

Alinity 08P4220Recalls & reported problems

Abbott GmbH · Alinity c Creatine Kinase Reagent Kit 1200 Tests · Class 2

All models of Alinity (37)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740132590), received between Jul 27, 2020 and Feb 25, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Low Test Results

    THE CUSTOMER OBSERVED A FALSELY DECREASED ALINITY C CREATINE KINASE RESULT FOR ONE 13-YEAR-OLD PATIENT. THE LOW RESULT WAS NOT REPORTED. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 29-200 U/L): SID (B)(6) PROCESSED ON (B)(6) 2026 INITIAL RESULT WAS 576 WITH DILUTION. REPEAT RESULT = 5993 U/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24443042 ↗
  • MalfunctionJul 27, 2020Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY DECREASED CREATINE KINASE (CK) RESULT GENERATED ON THE ALINITY C ANALYZER ON ONE PATIENT. RESULTS PROVIDED: ON (B)(6) 2020 SID (B)(6) = 95.6 U/L, REPEATED AT 95 U/L, ANOTHER LABORATORY = > 1000. IT WAS NOTED THAT THE SAMPLE WAS LIPEMIC AND THE CK RESULT WAS OBTAINED POST CENTRIFUGATION FOR 5 MINUTES / 10,000 RPM. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10326205 ↗

FDA clearance and classification

The FDA regulates product code CGS (nad reduction/nadh oxidation, cpk or isoenzymes) under 21 CFR 862.1215 as a Class II device. 21 CFR 862.1215 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Alinity 08P4220

Has the Alinity 08P4220 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740132590), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Alinity 08P4220?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 27, 2020 and Feb 25, 2026. In those reports, the most common specific device problem is “Low Test Results” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alinity 08P4220?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alinity 08P4220, and is it still sold?

The FDA's device database lists Abbott GmbH as the company responsible for it (brand Alinity, model 08P4220). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices

All NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page