Skip to content

ARCHITECT 7D63-22Recalls & reported problems

Abbott GmbH · Creatine Kinase · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740161699), received between Apr 22, 2024 and Apr 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2024Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED CREATINE KINASE RESULTS FOR ONE PATIENT. THE FOLLOWING RESULTS FROM ONE PATIENT WERE PROVIDED: SID (B)(6). INITIAL RUN (ERROR CODE 1350 ¿UNABLE TO CALCULATE RESULT NO ABSORBANCE READS WITHIN ABSORBANCE RANGE¿). DILUTION: 1:2, NO RESULT EC1350. RESULTS WERE GENERATED WITH DILUTED SAMPLE (1:10): <70 U/L. REPEAT RUN ON SN (B)(6). (ERROR CODE 1350 ¿UNABLE TO CALCULATE RESULT NO ABSORBANCE READS WITHIN ABSORBANCE RANGE¿), DILUTION:1:2, NO RESULT, EC 1350. DILUTION:1:10, RESULT >42670 U/L. SID (B)(6). INITIAL RUN (ERROR CODE 1350 ¿UNABLE TO CALCULATE RESULT NO ABSORBANCE READS WITHIN ABSORBANCE RANGE¿) DILUTED SPECIMEN 1:2, NO RESULT EC 1350. DILUTED SPECIMEN 1:10; RESULT:<70 U/L. REPEAT RUN ON SN (B)(6). (ERROR CODE 1350 ¿UNABLE TO CALCULATE RESULT NO ABSORBANCE READS WITHIN ABSORBANCE RANGE¿), DILUTION:1:2, NO RESULT EC 1350, DILUTION:1:10, RESULT >42670. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19156592 ↗

Other NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices

All NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.