Skip to content

Dimension Vista® Flex® reagent cartridge CKI 10464356Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · Creatine Kinase Flex® reagent cartridge, 800 Tests · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768027925), received between Nov 8, 2018 and Feb 6, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K083465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2
  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025High Test Results; Low Test Results; High Test Results; Low Test Results

    THE CUSTOMER REPORTED TO SIEMENS THAT THEY OBTAINED SIX (6) ERRONEOUS CREATINE KINASE (CKI) RESULTS ON A PATIENT SAMPLE ON A DIMENSION VISTA 500 INTELLIGENT LAB SYSTEM. THE INITIAL ERRONEOUS RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED TWICE WITH AUTO DILUTION AND THRICE WITH MANUAL DILUTION ON THE ORIGINAL DIMENSION VISTA 500 INTELLIGENT LAB SYSTEM. THE REPROCESSED RESULTS OBTAINED WERE HIGHER AND CONSIDERED ERRONEOUS. THE SAME SAMPLE WAS REPROCESSED WITH 1:231 MANUAL DILUTION ON THE ORIGINAL DIMENSION VISTA 500 INTELLIGENT LAB SYSTEM AND A RESULT OBTAINED WAS HIGHER THAN THE INITIAL RESULT AND LOWER THAN THE OTHER REPROCESSED RESULTS. THE REPROCESSED RESULT WITH 1:231 MANUAL DILUTION WAS CONSIDERED CORRECT AND REPORTED TO THE PHYSICIAN. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE ERRONEOUS CREATINE KINASE (CKI) RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21320856 ↗
  • MalfunctionNov 8, 2018Low Test Results; Low Test Results

    A DISCORDANT, BELOW ASSAY RANGE, CREATINE KINASE (CKI) RESULT WAS REPORTED FOR A PATIENT SAMPLE FROM A DIMENSION VISTA 1500 SYSTEM. THE RESULT WAS PROVIDED TO THE PHYSICIAN. THE SAME SAMPLE WAS REPROCESSED THE SAME DAY WITH A DILUTION AND A HIGHER, CORRECT RESULT WAS OBTAINED. A CORRECTED REPORT WAS ISSUED. NO TREATMENT WAS PROVIDED TO THE PATIENT ON THE BASIS OF THE REPORTED DISCORDANT CKI RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT CKI RESULT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8053079 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083465. One submission often covers many models, so check the product and lot details in each record.

  • Z-1659-2018 · Jun 29, 2017 · Terminated
    Dimension Vista Creatine Kinase MB (MBI) assay; (Product Number: K3032/10464339) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)

Other NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices

All NAD REDUCTION/NADH OXIDATION, CPK OR ISOENZYMES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page