Alinity 07P5530Recalls & reported problems

Abbott GmbH · Alinity c Glucose Reagent Kit · Class 2

All models of Alinity (37)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740135454), received between May 27, 2021 and Feb 21, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2
  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C GLUCOSE RESULT GENERATED BY THE ALINITY C PROCESSING MODULE FOR A PATIENT WITH ACUTE KIDNEY FAILURE. THE SAMPLE GENERATED ERROR 1039 (UNABLE TO CALCULATE RESULT. ABSORBANCE EXCEEDED OPTICAL LIMITS). THE LAB TECH THEN PERFORMED AN AUTOMATIC DILUTION WHICH GENERATED A RESULT OF 1,867 MG/DL. THIS RESULT WAS RELEASED OUT OF THE LAB AND QUESTIONED BY THE PHYSICIAN AS IT WAS INCONSISTENT WITH THE RESULTS FROM THE POINT-OF-CARE DEVICE. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 70 TO 109 MG/DL. SID (B)(6): ON (B)(6)2025 @20:53 PM INITIAL RESULT = NO RESULT (EXCEPTION ERROR 1039), REPEAT #1 (UNDILUTED) = NO RESULT (EXCEPTION ERROR 1039), REPEAT #2 (UNDILUTED) = NO RESULT (EXCEPTION ERROR 1039), REPEAT #3 AUTOMATIC DILUTION (1:5) RESULT = 1,867 MG/DL, REPEAT #5 (UNDILUTED) = NO RESULT (EXCEPTION ERROR 1039), REPEAT #6 (UNDILUTED) = NO RESULT (EXCEPTION ERROR 1039), REPEAT #7 MANUAL DILUTION = NO RESULT (EXCEPTION ERROR 1039). GLUCOSE RESULTS BEFORE AND AFTER: ON (B)(6) 2025 @16:40 PM RESULT = 96 MG/DL, ON (B)(6) 2025 @03:13 AM RESULT = 93 MG/DL. GLUCOSE RESULTS FROM POINT-OF-CARE: 97 MG/DL, 101 MG/DL, 100 MG/DL. THE CUSTOMER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21431204 ↗
  • MalfunctionAug 22, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSELY DECREASED ALINITY C GLUCOSE RESULTS GENERATED ON THE ALINITY C PROCESSING MODULE FOR ONE PATIENT WITH DIABETES. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): INITIAL RESULT = < 5 MG/DL A RAPID BLOOD GLUCOSE TEST WAS PERFORMED ON THE PATIENT AND THE GLUCOSE RESULT OBTAINED WAS 240 MG/DL THE SAMPLE WAS REPEATED ON TWO OTHER INSTRUMENTS PRESENT IN THE LABORATORY: INSTRUMENT AC07153 REPEATED IN DUPLICATE AND GENERATED A RESULT OF < 5 MG/DL BOTH TIMES. INSTRUMENT (B)(6) REPEATED IN DUPLICATE ON CONCENTRATED AND DILUTED SAMPLE: RESULT OF UNDILUTED SAMPLE = < 5 MG/DL. RESULT OF DILUTED SAMPLE = < 25 MG/DL. THE CUSTOMER PERFORMED TESTING ON A SECOND SAMPLE FROM THE DIABETIC PATIENT: SAMPLE TUBE WITHOUT GEL SID (B)(6) RESULT = 50 MG/DL (RESULT BEING QUESTIONED) SAMPLE TUBE WITH GEL SID (B)(6) RESULT = 150 MG/DL (RESULT CONSIDERED IN LINE WITH PATIENT HISTORY) THE MEASURING INTERVAL OF THE ALINITY C GLUCOSE ASSAY SERUM/PLASMA APPLICATION IS 5 TO 800 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20042293 ↗
  • MalfunctionMay 27, 2021Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED ALINICY C GLUCOSE RESULTS FOR ONE PATIENT. THE INITIAL RESULT WAS >800 MG/DL, REPEATED WITH 1:5 DILUTION AND THE RESULT WAS 315 MG/DL; THE INITIAL SAMPLE WAS REPEATED AND THE RESULT WAS >800 MG/DL, REPEATED WITH 1:5 DILUTION AND THE RESULT WAS 845 MG/DL. THE CUSTOMER IS QUESTIONING THE 1:5 DILUTION RESULT OF 315 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11895114 ↗

FDA clearance and classification

The FDA regulates product code CFR (hexokinase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the Alinity 07P5530

Has the Alinity 07P5530 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740135454), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Alinity 07P5530?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 27, 2021 and Feb 21, 2025. In those reports, the most common specific device problems are “Low Test Results” (2) and “High Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alinity 07P5530?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alinity 07P5530, and is it still sold?

The FDA's device database lists Abbott GmbH as the company responsible for it (brand Alinity, model 07P5530). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.