ARCHITECT 3L82-42Recalls & reported problems
Abbott GmbH · Glucose · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740161590), received between Jul 6, 2022 and Feb 27, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results3
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED GLUCOSE RESULT GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE ON A PATIENT. RESULTS PROVIDED: 13.14 / 5.87 / 5.90 MMOL/L (REFERENCE RANGE: 3.9-6.11 MMOL/L) NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21477206 ↗MalfunctionOct 11, 2023Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED GLUCOSE RESULT GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE PATIENT. THE SAME SAMPLE WAS REPEATED AND HIGHER RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S NORMAL REFERENCE RANGE: 70 TO 105 MG/DL. (B)(6) 2023 SID (B)(6) INITIAL RESULT = 27.2 MG/DL. REPEATED RESULTS (ON THE SAME INSTRUMENT AND TWO OTHER INSTRUMENTS) = AROUND 160 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17911990 ↗MalfunctionJan 27, 2023Low Test Results; Low Test Results
AFTER FURTHER REVIEW, THE FOLLOWING ADDITIONAL SAMPLES WERE IMPACTED DUE TO THE GLUCOSE REAGENT BEING MISREAD BY THE ARCHITECT C16000 PROCESSING MODULE¿S BARCODE READER (RESULTS IN MG/DL): SAMPLE ID (B)(6) INITIAL RESULT WAS 49, REPEATED RESULT WAS 95. SAMPLE ID (B)(6) INITIAL RESULT WAS 61, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 50, REPEATED RESULT WAS 87. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 136. SAMPLE ID (B)(6) INITIAL RESULT WAS 62, REPEATED RESULT WAS 90. SAMPLE ID (B)(6) INITIAL RESULT WAS 70, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 69, REPEATED RESULT WAS 104. SAMPLE ID (B)(6) INITIAL RESULT WAS 52, REPEATED RESULT WAS 80. SAMPLE ID (B)(6) INITIAL RESULT WAS 57, REPEATED RESULT WAS 101. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 89. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 85. SAMPLE ID (B)(6) INITIAL RESULT WAS 63, REPEATED RESULT WAS 95. SAMPLE ID (B)(6) INITIAL RESULT WAS 53, REPEATED RESULT WAS 74. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED RESULT WAS 81. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16254151 ↗MalfunctionJul 6, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTED A FALSELY REDUCED GLUCOSE RESULT GENERATED ON THE ARCHITECT C16000 ANALYZER ON A PATIENT. RESULTS PROVIDED : SID (B)(6): (B)(6) 2022 INITIAL RESULT 20 MG/DL, REPEAT ON ANOTHER ARCHITECT = 170 MG/DL. REPEATED ON THIS ANALYZER = 167.2 MG/DL AND 175.1 MG/DL.CUSTOMER'S REFERENCE RANGE: 70-135 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14935823 ↗
Other HEXOKINASE, GLUCOSE devices
- Glucose HK 4404483190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE30 reports
- Glucose HK 08057800190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE19 reports
- ARCHITECT 3L82-21Abbott Laboratories · HEXOKINASE, GLUCOSE14 reports
- Glucose HK Gen3 5168791190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE10 reports
- ARCHITECT 3L82-22Abbott GmbH · HEXOKINASE, GLUCOSE8 reports
- Glucose HK 4657527190Roche Diagnostics GmbH · HEXOKINASE, GLUCOSE7 reports
- Alinity 07P5520Abbott GmbH · HEXOKINASE, GLUCOSE3 reports
- Alinity 07P5530Abbott GmbH · HEXOKINASE, GLUCOSE3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page