The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740161590), received between Jul 6, 2022 and Feb 27, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results3
  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 27, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED GLUCOSE RESULT GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE ON A PATIENT. RESULTS PROVIDED: 13.14 / 5.87 / 5.90 MMOL/L (REFERENCE RANGE: 3.9-6.11 MMOL/L) NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21477206 ↗
  • MalfunctionOct 11, 2023Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DECREASED GLUCOSE RESULT GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE PATIENT. THE SAME SAMPLE WAS REPEATED AND HIGHER RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S NORMAL REFERENCE RANGE: 70 TO 105 MG/DL. (B)(6) 2023 SID (B)(6) INITIAL RESULT = 27.2 MG/DL. REPEATED RESULTS (ON THE SAME INSTRUMENT AND TWO OTHER INSTRUMENTS) = AROUND 160 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17911990 ↗
  • MalfunctionJan 27, 2023Low Test Results; Low Test Results

    AFTER FURTHER REVIEW, THE FOLLOWING ADDITIONAL SAMPLES WERE IMPACTED DUE TO THE GLUCOSE REAGENT BEING MISREAD BY THE ARCHITECT C16000 PROCESSING MODULE¿S BARCODE READER (RESULTS IN MG/DL): SAMPLE ID (B)(6) INITIAL RESULT WAS 49, REPEATED RESULT WAS 95. SAMPLE ID (B)(6) INITIAL RESULT WAS 61, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 50, REPEATED RESULT WAS 87. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 136. SAMPLE ID (B)(6) INITIAL RESULT WAS 62, REPEATED RESULT WAS 90. SAMPLE ID (B)(6) INITIAL RESULT WAS 70, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 69, REPEATED RESULT WAS 104. SAMPLE ID (B)(6) INITIAL RESULT WAS 52, REPEATED RESULT WAS 80. SAMPLE ID (B)(6) INITIAL RESULT WAS 57, REPEATED RESULT WAS 101. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 89. SAMPLE ID (B)(6) INITIAL RESULT WAS 54, REPEATED RESULT WAS 85. SAMPLE ID (B)(6) INITIAL RESULT WAS 63, REPEATED RESULT WAS 95. SAMPLE ID (B)(6) INITIAL RESULT WAS 53, REPEATED RESULT WAS 74. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED RESULT WAS 86. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED RESULT WAS 81. SAMPLE ID (B)(6) INITIAL RESULT WAS 68, REPEATED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16254151 ↗
  • MalfunctionJul 6, 2022Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A FALSELY REDUCED GLUCOSE RESULT GENERATED ON THE ARCHITECT C16000 ANALYZER ON A PATIENT. RESULTS PROVIDED : SID (B)(6): (B)(6) 2022 INITIAL RESULT 20 MG/DL, REPEAT ON ANOTHER ARCHITECT = 170 MG/DL. REPEATED ON THIS ANALYZER = 167.2 MG/DL AND 175.1 MG/DL.CUSTOMER'S REFERENCE RANGE: 70-135 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.