Skip to content

Alinity 07P9221Recalls & reported problems

Abbott Ireland Diagnostics · Alinity i Total PSA Reagent Kit 200 Tests · Class 3

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740158316), received between Feb 5, 2023 and Dec 21, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 21, 2023Low Test Results; Low Test Results

    THE CUSTOMER STATED THAT ERROR 1044, UNABLE TO CALCULATE RESULT. FINAL RLU READ IS OUTSIDE THE SPECIFICATION OF THE LOWEST CALIBRATOR FOR TOTAL PSA, WAS GENERATED FOR THREE PATIENT SAMPLES WHILE USING THE ALINITY I TOTAL PSA ASSAY. TWO OF THE THREE SAMPLES WERE TESTED ON A DIFFERENT ALINITY I PROCESSING MODULE AND THE SAME ERROR OCCURRED. THE CUSTOMER STATED THAT TWO OF THE PATIENTS DID HAVE VERY LOW TOTAL PSA RESULTS HISTORICALLY. THE CUSTOMER USED NEW TOTAL PSA CALIBRATOR LOT 51060FN00 TO RESOLVE THE ISSUE AND THE THREE PATIENT SAMPLES WERE RETESTED. PATIENT (B)(6) AND (B)(6): ALINITY I TOTAL PSA RESULT: <0.1 NG/ML. THE RESULTS WERE EXPECTED TO BE LOW FOR BOTH PATIENTS. NO HISTORICAL VALUES WERE PROVIDED. PATIENT (B)(6): ALINITY I TOTAL PSA RESULT: <0.1 NG/ML; ALINITY I TOTAL PSA RESULT (DIFFERENT SAMPLE TUBE): 3.86 AND 4.12 NG/ML; THE PATIENT'S LAST TOTAL PSA RESULT FROM (B)(6) 2023 WAS 7.7 NG/ML. THE CUSTOMER HAD THE PATIENT REDRAWN BECAUSE THE MEDICAL PROVIDER QUESTIONED THE LOW RESULT OF <0.1 NG/ML. THE REDRAW RESULT WAS 3.3 NG/ML. THE CUSTOMER'S REFERENCE RANGE IS 0 - 4 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18376168 ↗
  • MalfunctionFeb 5, 2023Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED ALINITY I TOTAL PSA RESULTS FOR ONE PATIENT THAT WAS QUESTIONED BY THE PHYSICIAN. THE PATIENT RESULTS WERE ELEVATED FROM (B)(6) 2021 THROUGH (B)(6) 2022 AND THEN SUDDENLY DROPPED (B)(6) 2022. THE RESULTS WERE STILL LOW ON (B)(6) 2023. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE WAS <4.0 NG/ML): (B)(6) 2021 RESULT WAS <3.8 NG/ML (FROM SIEMENS INSTRUMENT), (B)(6) 2021 RESULT = 6.01 NG/ML (FROM SIEMENS INSTRUMENT), (B)(6) 2022 RESULT = 5.03 NG/ML (FROM SIEMENS INSTRUMENT), (B)(6) 2022 RESULT = 5.9 NG/ML (FROM ALINITY OFF TRACK), (B)(6) 2022 RESULT = 7.8 NG/ML (FROM ALINITY OFF TRACK), (B)(6) 2022 RESULT = 2.0 NG/ML (FROM ALINITY ON TRACK), (B)(6) 2023 RESULT = 2.5 NG/ML (FROM ALINITY ON TRACK), THE CUSTOMER TRANSITIONED FROM SIEMENS INSTRUMENT TO ABBOTT¿S (ALINITY I PROCESSING MODULE) AFTER (B)(6) 2022. A NEW SAMPLE WAS COLLECTED FROM THE PATIENT (UNKNOWN COLLECTION DATE): SID (B)(6): (GOLD TOP, MOTHER TUBE) RESULTS = 5.98 NG/ML, 6.00 NG/ML, AND 6.37 NG/ML. SID (B)(6): (ALIQUOT, FROZEN TUBE) RESULTS = 6.05 NG/ML, 6.30 NG/ML, AND 6.53 NG/ML. ADDITIONAL DATA WAS PROVIDED FOR TROUBLESHOOTING PURPOSES ONLY. THE DATA SHOWS ANOT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16301957 ↗

Other Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer devices

All Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.