Elecsys total PSA 07027966190Recalls & reported problems
Roche Diagnostics GmbH · Class 3
All models of Elecsys total PSA (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04015630939824), received between Aug 17, 2018 and May 29, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect Measurement1
- Insufficient Device Problem Information1
- Low Test Results1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2025Signal Artifact/Noise; Incorrect Measurement; Signal Artifact/Noise; Incorrect Measurement
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS TOTAL PSA RESULTS FROM THE COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH WILL APPLY TO THE ELECSYS TOTAL PSA ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR THE INFORMATION RELATED TO THE ELECSYS FREE PSA ASSAY. PATIENT 1'S FREE PSA RESULT ON (B)(6) 2025, WAS 3.33 NG/ML, AND THE TOTAL PSA RESULT WAS 2.46 NG/ML. A NEW SAMPLE WAS REQUESTED ON (B)(6) 2025, AND THE FREE PSA RESULT WAS 2.48 NG/ML, AND THE TOTAL PSA RESULT WAS 1.78 NG/ML. PATIENT 2'S FREE PSA RESULT ON (B)(6) 2025, WAS 0.798 NG/ML, AND THE TOTAL PSA RESULT WAS 0.220 NG/ML. PATIENT 3'S FREE PSA RESULT ON (B)(6) 2025, WAS 0.332 NG/ML, AND THE TOTAL PSA RESULT WAS 0.903 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22106088 ↗MalfunctionJul 14, 2021High Test Results; Insufficient Device Problem Information; High Test Results; Insufficient Device Problem Information
THE INITIAL REPORTER RECEIVED QUESTIONABLE ELECSYS FREE PSA IMMUNOASSAY AND ELECSYS TOTAL PSA IMMUNOASSAY RESULTS FOR ONE PATIENT TESTED ON A COBAS 8000 E 801 MODULE. PRIOR TO PATIENT TESTING, THE CUSTOMER CONFIRMED QC WAS ACCEPTABLE. THE PATIENT HAD A SAMPLE COLLECTED ON (B)(6) 2021 FOR FREE PSA AND TOTAL PSA TESTING. ON EACH DATE OF TESTING, THE INITIAL RESULTS WERE REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER QUESTIONED THE REPORTED FREE PSA AND TOTAL PSA RESULTS. THE PATIENT'S SAMPLES WERE NOT REPEATED DUE TO THE SAMPLES BEING DISCARDED. ON (B)(6) 2021, THE PATIENT'S RESULTS WERE FREE PSA 4.240 NG/ML AND TOTAL PSA 1.34 NG/ML. ON (B)(6) 2021, THE PATIENT'S RESULTS WERE FREE PSA 4.810 NG/ML AND TOTAL PSA 1.630 NG/ML. THE CUSTOMER WAS UNSURE WHICH E 801 MODULE WAS USED FOR TESTING. THE E 801 MODULE SERIAL NUMBERS ARE (B)(4). THIS MEDWATCH IS FOR TOTAL PSA. REFER TO THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR THE FREE PSA ASSAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12167891 ↗MalfunctionAug 17, 2018Low Test Results; Low Test Results
THE CUSTOMER COMPLAINED OF QUESTIONABLE ELECSYS FREE PSA IMMUNOASSAY AND ELECSYS TOTAL PSA IMMUNOASSAY RESULTS FOR 1 PATIENT SAMPLE TESTED ON A COBAS 8000 E801 MODULE. THE CUSTOMER SUSPECTS THAT THERE IS AN INTERFERENCE PRESENT IN THE PATIENT SAMPLE. THIS MEDWATCH WILL COVER THE TOTAL PSA DATA. PLEASE REFER TO THE MEDWATCH WITH PATIENT IDENTIFIER (B)(6) FOR DATA ON THE FREE PSA. THE INITIAL FREE PSA RESULT WAS 3.87 NG/ML WITH REPEAT RESULTS OF 3.91 NG/ML AND 3.65 NG/ML. THE INITIAL TOTAL PSA RESULT WAS 3.00 NG/ML WITH REPEAT RESULTS OF 3.05 NG/ML AND 2.99 NG/ML. NO ERRONEOUS RESULTS WERE RELEASED OUTSIDE OF THE LABORATORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE COBAS E801 SERIAL NUMBER WAS (B)(4). THE INVESTIGATION IS CURRENTLY ONGOING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7792763 ↗
Other Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer devices
- IMMULITE® 2000 Systems PTS PSA 10706281Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall1 report
- IMMULITE® 2000 Systems PTS PSA 10706282Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall0 reports
- Elecsys total PSA 08791732190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer48 reports
- Dimension Vista® System TPSA Flex® Reagent Cartridge 10445090Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer36 reports
- Dimension® clinical chemistry system TPSA Flex® reagent cartridge 10444889Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer12 reports
- Elecsys Total PSA 09744860190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer4 reports
- Elecsys total PSA 08791686160Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer3 reports
- Access Hybritech PSA 37200Beckman Coulter, Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page