Elecsys Total PSA 09744860190Recalls & reported problems
Roche Diagnostics GmbH · Class 3
All models of Elecsys total PSA (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336217665), received between Jul 3, 2025 and Mar 30, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results3
- High Test Results2
- Incorrect Measurement1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026High Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH TOTAL PSA ELECSYS E2G ASSAY ON A COBAS E 801 ANALYTICAL UNIT, NOT MATCHING THE EXPECTED RESULTS. INITIAL RESULT: <0.014 NG/ML. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN AS 0.014 NG/ML (NOT <0.014 NG/ML), AND ACCOMPANIED BY "ERROR 27". THE RESULT OF 0.014 NG/ML WAS HIGHER THAN THE LABORATORY'S RULE OF 0.006 NG/ML AND WAS SENT TO THE PHYSICIAN. IT WAS CONSIDERED INACCURATE AS THE PHYSICIAN AND THE PATIENT EXPECTED A RESULT LOWER THAN THE REPORTABLE RANGE OF <0.006 NG/ML. THE SAMPLE WAS THEN REPEATED ON A DIFFERENT INSTRUMENT AT THE CUSTOMER'S SITE WITH A LABORATORY-DEVELOPED TEST WHOSE VERSION HAD NOT BEEN UPDATED, AND THE REPEAT RESULT WAS <0.006 NG/ML (ACCOMPANIED BY A DATA FLAG). THE RESULT OF <0.014 NG/ML USING THE UPDATED LABORATORY'S RULE WAS DEEMED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24715641 ↗MalfunctionFeb 24, 2026High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE HIGH TOTAL PSA ELECSYS E2G RESULTS FROM THE COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 100 G/L. THE REPEAT RESULT USING A DILUTION WAS 5000 G/L WITH A DATA FLAG. ON (B)(6) 2026, THE SAMPLE WAS REPEATED USING PEG, AND THE RESULT WAS 100 G/L. THE PHYSICIAN BELIEVED THAT THIS WAS NOT RELIABLE, ESPECIALLY AS THE PATIENT DIDN'T HAVE ANY SYMPTOMS AND WAS BEING TESTED AS A ROUTINE CHECK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24427941 ↗MalfunctionJul 25, 2025Incorrect Measurement; Non Reproducible Results; Incorrect Measurement; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT QUESTIONABLE RESULTS FOR 2 SAMPLES FROM THE SAME PATIENT TESTED WITH ELECSYS TOTAL PROSTATE SPECIFIC ANTIGEN (TOTAL PSA) ASSAY AND ELECSYS FREE PROSTATE SPECIFIC ANTIGEN (FREE PSA) ASSAY ON A COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH WILL APPLY TO THE TOTAL PSA ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FREE PSA ASSAY. SAMPLE 1: TOTAL PSA: INITIAL RESULT: 0.0821 NG/ML. FREE PSA: INITIAL RESULT: 1.99 NG/ML. THIS FREE PSA RESULT WAS NOT REPORTED OUTSIDE THE LABORATORY. THE FREE PSA RESULT WAS GREATER THAN THE TOTAL PSA RESULT, PROMPTING THE REPEAT OF THE SAMPLE THE NEXT DAY. THE REPEAT RESULTS WERE IDENTICAL TO THE INITIAL RESULTS. ON (B)(6) 2025, A NEW BLOOD SAMPLE (SAMPLE 2) WAS TAKEN FROM THE PATIENT AND TESTED: TOTAL PSA RESULT: 0.0807 NG/ML. FREE PSA RESULT: 1.99 NG/ML. OTHER TOTAL PSA AND FREE PSA TESTS WERE PERFORMED AT A DIFFERENT SERVICE LABORATORY USING SIEMENS ATELLICA SOLUTION, RESULTING IN THE FOLLOWING: TOTAL PSA RESULT: 1.220 UG/L. FREE PSA RESULT: 0.023 UG/L. THESE RESULTS WERE DEEMED TO BE CORRECT. IT IS NOT KNOWN IF SAMPLE 2 WAS REPEATED ON SIEMENS TESTING OR IF ANOTHER NEW S
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22612651 ↗MalfunctionJul 3, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS TOTAL PROSTATE SPECIFIC ANTIGEN (TOTAL PSA) ASSAY ON A COBAS E 801 ANALYTICAL UNIT. INITIAL RESULT: 0.6 NG/ML. ON (B)(6) 2025: REPEAT RESULT: 31.5 NG/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22393423 ↗
Other Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer devices
- IMMULITE® 2000 Systems PTS PSA 10706281Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall1 report
- IMMULITE® 2000 Systems PTS PSA 10706282Siemens Healthcare Diagnostics Products Ltd · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer1 recall0 reports
- Elecsys total PSA 08791732190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer48 reports
- Dimension Vista® System TPSA Flex® Reagent Cartridge 10445090Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer36 reports
- Dimension® clinical chemistry system TPSA Flex® reagent cartridge 10444889Siemens Healthcare Diagnostics Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer12 reports
- Elecsys total PSA 08791686160Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer3 reports
- Elecsys total PSA 07027966190Roche Diagnostics GmbH · Total, prostate specific antigen (noncomplexed & complexed) for detection of prostate cancer3 reports
- Access Hybritech PSA 37200Beckman Coulter, Inc. · Total,Prostate Specific Antigen(Noncomplexed&Complexed) For Detection Of Prostate Cancer3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page