Alinity 07P5131Recalls & reported problems
Abbott Ireland · Alinity i Total B-hCG Reagent Kit 1200 Tests · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380740160623), received between Jul 30, 2019 and Apr 17, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result5
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2025False Positive Result; False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY I TOTAL B-HCG RESULT FOR AN 18-YEAR-OLD FEMALE. THE DOCTOR QUESTIONED THE POSITIVE RESULT. THE ORIGINAL SAMPLE WAS REPEATED AND WAS NEGATIVE. THE PATIENT WAS RECOLLECTED, AND THE SECOND SAMPLE WAS ALSO NEGATIVE. THE FOLLOWING DATA WAS PROVIDED: ORIGINAL SAMPLE SID (B)(6): INITIAL RESULT = 103.8 MIU/ML. SID (B)(6) (SAME PATIENT/SAME SAMPLE): REPEAT RESULT = < 2.42 MIU/ML. SECOND SAMPLE COLLECTED: SID (B)(6): INITIAL RESULT = < 2.42 MIU/ML. NORMAL RANGE: <5 MIU/ML = NEGATIVE, 5 - 25 MIU/ML = INDETERMINATE (GRAYZONE), >25 MIU/ML = POSITIVE. ALL LEVELS OF QC WERE WITHIN RANGE. CUSTOMER VERIFIED THERE WAS NO FIBRIN IN THE SERUM SAMPLE. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21855801 ↗MalfunctionOct 12, 2023False Positive Result; False Positive Result
THE CUSTOMER REPORTED A FALSE POSITIVE ALINITY I TOTAL B-HCG RESULT FOR A 19-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE B-HCG: </= 5.00 MIU/ML IS NEGATIVE, > 5.00 AND < 25.00 MIU/ML IS GRAYZONE, >/= 25.00 MIU/ML IS POSITIVE): SID (B)(6): ON (B)(6) 2023 INITIAL B-HCG = 1204 MIU/ML (POSITIVE); ON (B)(6) 2023 POINT OF CARE URINE TEST = NEGATIVE; ON (B)(6) 2023 ORIGINAL SAMPLE REPEATED = 4 MIU/ML (NEGATIVE); PREVIOUS HISTORICAL RESULT GENERATED ON (B)(6) 2023 WAS 44 MIU/ML (POSITIVE). ON (B)(6) 2023, THE CUSTOMER PROVIDED ANOTHER CASE OF A FALSE POSITIVE ALINITY I TOTAL B-HCG RESULT FOR A PATIENT SAMPLE. THE SAMPLE ID (B)(6) GENERATED AN INITIAL HCG RESULT OF 339 MIU/ML; REPEAT RESULT GENERATED A VALUE OF < 1.20 MIU/ML. THE CUSTOMER STATED BOTH PATIENTS WERE ALMOST TREATED WITH METHOTREXATE BECAUSE DOCTORS INITIALLY SUSPECTED ECTOPIC PREGNANCY DUE TO INITIAL DISCREPANT RESULT. THE CUSTOMER CONFIRMED THE DOCTORS DID NOT TREAT THE PATIENTS SINCE REPEAT RESULTS WERE NEGATIVE. THERE WAS NO HARM TO THE PATIENTS. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17928006 ↗MalfunctionMar 7, 2023False Positive Result; False Positive Result
THE CUSTOMER OBSERVED A FALSE POSITIVE ALINITY I TOTAL -HCG RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2023, SID: (B)(6) INITIAL RESULT = 43.26 IU/L (POSITIVE) WHICH WAS QUESTIONED BY THE PHYSICIAN THE SAMPLE WAS REPEATED WITH A 1:15 DILUTION AND GENERATED A RESULT OF 7000 IU/L A NEW SAMPLE WAS DRAWN (SID: (B)(6) AND GENERATED A NEGATIVE RESULT. THE ORIGINAL SAMPLE WAS REPEATED ON ANOTHER ALINITY AND GENERATED A NEGATIVE RESULT OF 2.42 IU/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16499117 ↗MalfunctionJul 22, 2022False Positive Result; False Positive Result
THE ACCOUNT GENERATED FALSE POSITIVE ALINITY I TOTAL BHCG RESULTS ON A PRESURGICAL PATIENT SAMPLE FOR HYSTERECTOMY. ON (B)(6) 2022, SID (B)(6) (71 YEAR OLD FEMALE, NON-HISPANIC/LATINO, RACE NOT PROVIDED) GENERATED ALINITY I TOTAL BHCG POSITIVE OF 115.90, 114.31 MIU/ML AND ON ALTERNATE ALINITY ANALYZER (SERIAL (B)(4) POSITIVE OF 133.21 MIU/ML. THE PATIENT TESTED VITROS HCG NEGATIVE <3 MIU/ML AND CARDINAL QUAL HCG COMBO URINE NEGATIVE. THE ACCOUNT STATED THERE WAS NO ADVERSE IMPACT TO PATIENT CARE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15086727 ↗MalfunctionJul 30, 2019False Positive Result; False Positive Result
THE CUSTOMER STATED THAT FALSE POSITIVE ALINITY I TOTAL B-HCG RESULTS OF 228 IU/L AND 221 IU/L WERE GENERATED FOR TWO DIFFERENT PATIENT SAMPLES (SIDS (B)(6)) THAT RETESTED NEGATIVE AT <2.0 IU/L. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8841733 ↗
Other SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN devices
- IMMULITE® 2000 Systems HCG 10381206Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN1 recall173 reports
- ARCHITECT 7K78-25Abbott Ireland · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN176 reports
- ARCHITECT 7K78-30Abbott Ireland · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN110 reports
- i-STAT 05p58-25Abbott Point Of Care Inc. · System, Test, Human Chorionic Gonadotropin105 reports
- IMMULITE® 2000 Systems HCG 10381194Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN1 recall16 reports
- Alinity 07P5121Abbott Ireland · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN1 recall15 reports
- Elecsys HCG+beta 07251025190Roche Diagnostics GmbH · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN58 reports
- Elecsys HCG Test System 3271749160Roche Diagnostics GmbH · SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN39 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.